1 / 41

TURKEY REPUBLIC MINISTRY OF ENVIRONMENT AND FORESTRY TURKISH STATE METEOROLOGICAL SERVICE

TURKEY REPUBLIC MINISTRY OF ENVIRONMENT AND FORESTRY TURKISH STATE METEOROLOGICAL SERVICE DIRECTORY OF METEOROLOGICAL INSTRUMENTS AND CALIBRATION. CALIBRATION CENTER. KALMER. Calibration Center.

abner
Télécharger la présentation

TURKEY REPUBLIC MINISTRY OF ENVIRONMENT AND FORESTRY TURKISH STATE METEOROLOGICAL SERVICE

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. TURKEY REPUBLIC MINISTRY OF ENVIRONMENT AND FORESTRY TURKISH STATE METEOROLOGICAL SERVICE DIRECTORY OF METEOROLOGICAL INSTRUMENTS AND CALIBRATION CALIBRATION CENTER KALMER Calibration Center - Ankara 2010

  2. Calibration Center At the Calibration Center: Automated Weather Observing System's sensors that is installed to all around the Türkiye are calibrated by applying the 17025 standards. CalibrationCenter is accreditedforthecalibrations of tempetature, relativehumuidity, airpressureandwindspeedsensors. Calibration Center - Ankara 2010

  3. CALIBRASTION CENTER ORGANIZATIONAL CHART Calibration Center - Ankara 2010

  4. DIRECTORY OF METEOROLOGICAL INSTRUMENTS and CALIBRATION PRECIPITATION WIND CalibrationCenteroperatesunder MeteorologicalInstrumentsandCalibrationDivision. TEMP&HUM SOLAR RADIATION Mr. Nur DOĞAN Director of MeteorologicalInstruments andCalibrationDivision PRESSURE Calibration Center - Ankara 2010

  5. QUALITY MANAGER On behalf of the management: Responsible for the operation of the Calibration Center in accordance with the rules of the 17025 standards. Mr. Yusuf Ö. CANDEMİR QUALITY MANAGEMENT PERSONNEL Calibration Center - Ankara 2010

  6. CALIBRATION CENTER UNITS • InstrumentAcceptance, Registration and Archive, • Temperature & Humidity Calibration Laboratory, • Pressure Calibration Laboratory, • Wind Speed Calibration Laboratory, • Radiation Calibration Laboratory, • Precipitation Calibration Laboratory. • Mr. Tekin SOYLU • Seniorchief of theLaboratories CalibrationCenter - Ankara 2010

  7. CALIBRATION CENTER PERSONNEL CHART Calibration Center - Ankara 2010

  8. INSTRUMENT ACCEPTANCE & REGISTRATION AND ARCHİVE Instrumentsacceptedtothe Calibration Center are registeredandrelateddocumentsarearchivedin accordance with the rules of 17025 standards. Calibration Center - Ankara 2010

  9. CALIBRATION CENTER ENVIRONMENTAL CONDITIONS OF THE LABRATORIES Environmental conditions of theLaboratories are controlled by using precisionair-conditioner. The measured data of temperatureandhumuidity are automatically displayed and stored. Calibration Center - Ankara 2010

  10. CALIBRATION CENTER ENVIRONMENTAL CONDITIONS OF THE LABRATORIES Devices to control the laboratory environmental conditions: Fluke Hart Scientific 1620A Measuring, displaying and recording. LiebertHiross HPM PrecisionAir conditioner Calibration Center - Ankara 2010

  11. TEMPERATURE & HUMIDITY LABORATORY The thermometer typesto be calibrated: Liquid glass thermometers, Resistance thermometers, Analog output thermometers, Thermometerswithdisplay. All typesof relative humidity sensors are calibrated. Calibration Center - Ankara 2010

  12. TEMPERATURE & HUMIDITY LABORATORY Scope of Accreditation Calibration Center - Ankara 2010

  13. TEMPERATURE & HUMIDITY LABORATORY Reference devices for relative humidity calibration: Thunder Scientific 2500 Humidity Generator. Operation range: -10˚C / +70˚C 10%RH / 90%RH Vaisala HMP76 Calibration Center - Ankara 2010

  14. TEMPERATURE & HUMIDITY LABORATORY Climate provider forRelative Humidity Calibration: Weiss SB22 500 Climate Chamber. Operation Range: -50˚C / +50˚C 10%RH / 90%RH Calibration Center - Ankara 2010

  15. TEMPERATURE & HUMIDITY LABORATORY Reference devices forTemperature Calibration: Isotech 909Q Standard Platinum Resistance Thermometer Hart Scientific 5615 Platinum Resistance Thermometer Calibration Center - Ankara 2010

  16. TEMPERATURE & HUMIDITY LABORATORY Climate provider forTemperature Calibration: Weiss SB22 500 Climate Chamber. Operation Range: -50˚C / +50˚C 10%RH / 90%RH Fluke 7381 Liquid bath. Operation range: -80˚C /+110˚C Calibration Center - Ankara 2010

  17. TEMPERATURE & HUMIDITY LABORATORY Secondarydevices forTemperature & Humidity Laboratory: Fluke PRT measuring unit, Zeno 3200 Data Logger and Fluke 8846A 6 ½ digit multimeter are used to measure the data. Calibration Center - Ankara 2010

  18. TEMPERATURE & HUMIDITY LABORATORY Calibration at The Triple Point of Water (TPW) and Ice Point Calibration : Calibration Center - Ankara 2010

  19. PRESSURE CALIBRATION LABORATORY Pressure sensors typesto be calibrated: Frequency output barometers, Digital output barometers, Barometerswith digital display, Mechanical barometers. Calibration Center - Ankara 2010

  20. PRESSURE CALIBRATION LABORATORY Scope of Accreditation Calibration Center - Ankara 2010

  21. PRESSURE CALIBRATION LABORATORY Reference device for Pressure Calibration: Parosceintific 785 digiquartz barometer Calibration Center - Ankara 2010

  22. PRESSURE CALIBRATION LABORATORY Climate providers forPressure Calibration: Fluke PPC4-ui A100K Calibrator. Dry air are used (Azoth gas). Pressure Chamber Calibration Center - Ankara 2010

  23. PRESSURE CALIBRATION LABORATORY Secondary devicesof Pressure Laboratory: Druck DPI 145 Measurement device for frequency output barometer. Measurementdevice for digital output barometer. Calibration Center - Ankara 2010

  24. PRESSURE CALIBRATION LABORATORY Secondarydevices of the Pressure Laboratory: Paroscientific 765 digiquartz barometer. Ruska 6200 barometer. Calibration Center - Ankara 2010

  25. WIND SPEED CALIBRATION LABORATORY WindSpeed sensor typesto be calibrated (Cup, hotwire, ultrasonic, pitot, propeller): Electronics wind speed sensors, Mechanical wind speed sensors. Calibration Center - Ankara 2010

  26. WIND SPEED CALIBRATION LABORATORY Scope of Accreditation Calibration Center - Ankara 2010

  27. WIND SPEED CALIBRATION LABORATORY Reference devicesfor Wind Speed Calibration: Furness Control FCO510 Micromanometer and Pitot tube. Thies 2D 4.3810.00.340 Ultrasonic wind speed sensor Calibration Center - Ankara 2010

  28. WIND SPEED CALIBRATION LABORATORY Medium provider for Wind Speed Calibration: Wind Speed Tunnel is the Göttinger type, closed loop and open test area ( 0 – 40 m/s). Calibration Center - Ankara 2010

  29. WIND SPEED CALIBRATION LABORATORY Devicesof the Wind Speed Calibration Laboratory: Hotwire Testo 435 Wind Speed Sensor. Druck PTX7517-1 Pressure (600–1100mbar) PR 5333A Temperature (0-40ºC) sensors. Calibration Center - Ankara 2010

  30. WIND SPEED CALIBRATION LABORATORY Devicesof the Wind Speed Calibration Laboratory: Vaisala WMS302M and WAA151, ThiesClima 4.3303.22.007, Qualimetrics ECN4738, Siaphandel type, Vector İnst. A100M/HE-1 and Lastem DN002 Wind Speed sensors are used. Calibration Center - Ankara 2010

  31. RADIATION CALIBRATION LABORATORY Pyranometers are calibrated at the Radiation Calibration Laboratory under the constant irradiance and new sensitivity coefficients forthecalibratedsensorsare calculated. Calibration Center - Ankara 2010

  32. RADIATION CALIBRATION LABORATORY Scope of theCalibration Calibration Center - Ankara 2010

  33. RADIATION CALIBRATION LABORATORY Reference device for Pyranometer Calibration: Kipp&zonen CMP22 Pyranometer Calibration Center - Ankara 2010

  34. RADIATION CALIBRATION LABORATORY Medium provider forPyranometers Calibration: Kipp&Zonen Calibration Facility for Pyranometers. Indoor Pyranometer Calibration System under constant irradiance light (500W/m² - 700W/m²). Calibration Center - Ankara 2010

  35. RADIATION CALIBRATION LABORATORY Devicesof the Pyranometers Calibration Laboratory: Kipp&Zonen CMP21 Pyranometer Keithley 2000 6 ½ digit multimeter Calibration Center - Ankara 2010

  36. PRECIPITATION CALIBRATION LABORATORY Precipitation (Rain Gauge) sensors typesto be calibrated: Tipping bucket rain gauges, Weighing type rain gauges. Calibration Center - Ankara 2010

  37. PRECIPITATION CALIBRATION LABORATORY Scope of calibration CalibrationCenter - Ankara 2010

  38. PRECIPITATION CALIBRATION LABORATORY Reference devices forRain Gauges Calibration: Mettler Toledo XP4002S Precision ElectronicBalances 800 gr. / 4100 gr. Baykon F1 Class Weigths 1 gr / 2 kg Calibration Center - Ankara 2010

  39. PRECIPITATION CALIBRATION LABORATORY Medium provider for Rain Gauges Calibration: Precipitation simulation is prepared by using the Watson Marlow 520Du peristaltic pump which calculates the necessary amount of water for the desired amount of Rain İntensity. Calibration Center - Ankara 2010

  40. PRECIPITATION CALIBRATION LABORATORY Secondarydevices of the Precipitation Laboratory: Zeno 3200 Data Logger Caps and tubeswithdifferent type and dimensions. Calibration Center - Ankara 2010

  41. THANK YOU FOR YOUR INTEREST Calibration Center - Ankara 2010

More Related