1 / 9

Biodecontamination Validation Experiments: Method Comparison and Next Steps

Two methods of biodecontamination validation experiments using fixed mass and concentration of CD, conditions, results, and next steps for Method 1 and Method 2, including successful trials, treatments, and dosage considerations.

Télécharger la présentation

Biodecontamination Validation Experiments: Method Comparison and Next Steps

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Validation Experiments Method 1 (fixed mass of CD) • Experimenters – Micro-Clean • BSC Manufacturer – NuAire • Location – NuAire (Plymouth, MN) • Conditions: • 60 – 75% RH • CD amount - 0.1 g/ft3 • Exposure time – 80 min

  2. Results – Method 1 A2 -6' Console A2 – 4' Bench Top 0, 1, 2 = # positive BIs out of 2 FAIL PASS* Dose – mg-hr/L

  3. Results – Method 1 B1 -4′ Console B2 – 6' Console 0, 1, 2 = # positive BIs out of 2 FAIL PASS

  4. Method 1 – Next Steps • A2-4' and B2 cabinet trials successful • Repeat A2-6' experiment with proper BI treatment • Repeat B1 experiment with proper BI treatment and increase CD • CD requirement for Method 1 will be based on B1 requirement

  5. Validation Experiments Method 2 (fixed concentration of CD) • Experimenters – ClorDiSys Solutions • BSC Manufacturer – Baker • Location – Micro-Clean (PA) • Conditions: • 70% RH • Two concentrations (24+ trials) • 4.8 mg/L for 40 min • 2.8 gm/L for 55 min

  6. Results – Method 2 4.8 mg/L for 40 min A1 -6' Console A2 – 4' Console 0, 1, 2 = # positive BIs out of 2 PASS PASS Dose – mg-hr/L

  7. Results – Method 2 4.8 mg/L for 40 min B1 -6' Console B2 – 4' Console 0, 1, 2 = # positive BIs out of 2 PASS PASS

  8. Results – Method 2 2.8 mg/L for 55 min A2 B1 B2 A1 PASS PASS PASS

  9. Method 2 - Issues • Need to define appropriate recirculation blower (1 failed trial) Next Steps • A1, A2, B1, and B2 trials successful with 4.8 mg/L for 40 min. • Complete trials with 2.8 mg/L for 55 min.