HAPPEX/ Lumi PMT Studies

HAPPEX/ Lumi PMT Studies. Luis Mercado UMASS 7/10/09. PMT Linearity. Past tests have shown nonlinearities in PMT output dependant on input and gain. Want to study and document good linearity regions for Lumi and HAPPEX PMTs. PMT Linearity (Formalism).

anahid
Télécharger la présentation

HAPPEX/ Lumi PMT Studies

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. HAPPEX/Lumi PMT Studies Luis Mercado UMASS 7/10/09

  2. PMT Linearity • Past tests have shown nonlinearities in PMT output dependant on input and gain. • Want to study and document good linearity regions for Lumi and HAPPEX PMTs.

  3. PMT Linearity (Formalism) • A PMT’s non-linear response can be expressed as follows: • Experimental asymmetry can be approximated:

  4. Experimental Test Setup

  5. Lumi PMT Linearity Input = 12nA HV = 370 Output = uA B*N_0 = -5.14%

  6. Good Lumi PMT Linearity Input =~1nA HV = 690 Output ~ 7uA B*N_0 = 1.19%

  7. HAPPEX PMTs • Received one HAPPEX PMT last week. • Re-built experimental setup with bigger box. • Have done linearity tests at input photocathode currents of [0.06-6.0] nA. • Only good linearity region found was for an input of ~1nA at HV=1250.

  8. Linearity of output signal breaks down with high enough HV. • Signal becomes unpredictable in these regions.

  9. HAPPEX PMT Linearity Input ~1nA HV = 1100 Output = 7uA B*N_0 = -15.69%

  10. Good HAPPEX PMT Linearity Input ~ 1nA HV = 1250 Output~ 37 uA B*N_0 = 0.35%

  11. Odd Pedestal Behavior • The pedestal signal is huge for HV=1250. • Huge signal with no light.

  12. Odd Pedestal Behavior • Also happens with Lumi PMT.

  13. Ongoing Tests • Need to track down the cause of the odd pedestal behavior. • Study other available HAPPEX PMT. • Finalize linearity mapping for all PMTs.

  14. Acknowledgements • Krishna Kumar • Kent Pashke • Dustin McNulty • Robert Michaels • Aimee Shore • HAPPEX/PREX Collaboration