Nathaniel Knox Cartwright & Derek Tole Bristol Eye Hospital, Bristol, United Kingdom

Effect of Axial Length Measurement Method on Refractive Outcomes of Cataract Surgery: Real World Comparison of Partial Coherence Interferometry and Immersion Ultrasound. Nathaniel Knox Cartwright & Derek Tole Bristol Eye Hospital, Bristol, United Kingdom Financial Disclosure: none. Background.

boone
Télécharger la présentation

Nathaniel Knox Cartwright & Derek Tole Bristol Eye Hospital, Bristol, United Kingdom

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Effect of Axial Length Measurement Method on Refractive Outcomes of Cataract Surgery:Real World Comparison of Partial Coherence Interferometry and Immersion Ultrasound. Nathaniel Knox Cartwright & Derek Tole Bristol Eye Hospital, Bristol, United Kingdom Financial Disclosure: none

  2. Background • Partial coherence interferometry (PCI) has become the clinical gold standard for axial length (AL) measurement but cannot be used in all eyes • Intraindividual comparisons of PCI and immersion ultrasound (IUS) have found they result in postoperative refractive equivalence after cataract surgery • This contrasts with the clinical situation for PCI users who only use IUS as a second line measurement method • The purpose of this study was to compare refractive outcomes after cataract surgery when PCI is used for AL measurement with when IUS is used due to PCI failure

  3. Cohort • 2544 phacoemulsification and Akreos Adapt AO (Bausch & Lomb) IOL implantations • No copathology • Uncombined and uncomplicated surgery • Postoperative subjective refraction • Corrected distance visual acuity ≥6/12 (20/40) • IOLMaster (software version 3, Carl Zeiss Meditec) keratometry performed in all • PCI AL measurement attempted in all but IUS used if this was not possible

  4. Optimisation Process • Hoffer Q, Holladay 1 and SRK/T formulae entered in code into Freemat (http://freemat.sourceforge.net/) • Optimised biometry formula constants calculated by 2 methods • To give an overall mean deviation from predicted postoperative refraction of zero (zero method) • As recommended in the IOLMaster manual, that is as the mean of the all personalised constants excluding those > 2 SD from overall mean (Zeiss method)1 1. IOL Master Software Version 3.xx User Manual.http://www.doctor-hill.com/iol-master/docs/iolmaster_3.pdf (downloaded 17/0/2008).

  5. Comparison of Groups

  6. Results: Zero method Unpaired t-test: PCI V IUS all p>0.05

  7. Results: Zeiss method Unpaired t-test: PCI V IUS all p<0.05

  8. Zeiss method Zero method

  9. Conclusions • Optimised biometry formula constants (BFC) varied according the method by which they were calculated • When the Zeiss optimisation method was used, AL measurement by IUS following PCI measurement failure did not require use of different BFC and did not worsen refractive outcomes • Additionally, the Zeiss optimisation method resulted in a lower mean absolute error than when the zero method is used