1 / 32

Nano-chemical mapping performed by an AFM-based (“AFMIR”) acousto-optic technique and applications

Nano-chemical mapping performed by an AFM-based (“AFMIR”) acousto-optic technique and applications. A. Dazzi, F. Glotin, C. Mayet, J.-M. Ortega , R. Prazeres CLIO/LCP - Université Paris Sud - France. Why infrared ?. Many species have a characteristic absorption spectrum

channer
Télécharger la présentation

Nano-chemical mapping performed by an AFM-based (“AFMIR”) acousto-optic technique and applications

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Nano-chemical mapping performed by an • AFM-based (“AFMIR”) acousto-optic technique and applications A. Dazzi, F. Glotin, C. Mayet, J.-M. Ortega, R. Prazeres CLIO/LCP - Université Paris Sud - France

  2. Why infrared ? • Many species have a characteristic absorption spectrum • (“chemical signature”) • Molecules (vibrations) • Pseudo-atoms and molecules (defects in solids, nanostructures) • Elementary excitations in solids (plasmons, Cooper pairs...) Linear spectroscopy : Performed with blackbody sources and synchrotron radiation Non-Linear spectroscopy : Performed with lasers, particularly pulsed lasers Microspectroscopy : Resolution > l/2 : performed with blackbody sources and synchrotron radiation Resolution << l/2 : performed with lasers  tunable IR FEL

  3. Infrared microspectroscopy The infrared microspectroscopy associates lateral resolution (“microscopy”) and IR spectral analysis (“chemical identification”) Classical sources provide lateral resolution > l/2 Laser sources + near-field techniques should provide resolution << l (illumination of small objects requires beam of sufficient brightness) • Near-field techniques : • Fully optical methods (SNOM, PSTM, apertureless...) • New photo-acoustic method (“AFMIR”) to overcome • the problems encountered with usual means

  4. Example of a near-field method : the PSTM (photon scanning microscope) Optical fiber The electric field above the sample is sampled by the tip (< l) of an infrared transmitting optical fiber (1st method used at CLIO in1995)

  5. Disadvantages of all-optical near-field High spatial resolution (<< l) requires a very small thickness of the sample (<< l) --> Therefore the optical contrast becomes very small (< 1 %) (necessarily pulsed) tunable lasers do not allow observation of such low contrasts Moreover, there is no general method to separate the contributions from : - The real part of n = index of refraction (topography, inhomogeneities) - The imaginary part of n = optical absorption :desired quantity Differential methods are needed : photoacoustics

  6. The photothermal effect CLIO IR light

  7. deflection F Detection of dilatation Diode laser 4 quadrant detector AFM cantilever dilatation

  8. Experimental setup

  9. A new method : « AFMIR » • A commercial AFM is used • The sample is irradiated in PSTM configuration • An IR absorption results in a local dilatation • This dilatation is recorded by optical detection of the tip motion • The transient character of the dilatation (due to the pulsed nature of the laser) ==> spatial resolution • The topography is recorded at the same time

  10. Test experiment Amide I Amide II Amide III ADN Bacteria Escherichia coli Typical size : 2-6 µm long 1-2 µm large and 500 nm height FT-IR spectrum AFM microscopy

  11. Temporal response Tip Tip s=1650cm-1 (amide I) E.coli

  12. Tip Temporal response s=1800cm-1 (out of bands) E.coli

  13. Tip Tip Temporal response s=1650cm-1 (amide I) E.coli Time (µs)

  14. Fourier analysis s=1650cm-1 Resonances on E.coli Resonances on ZnSe

  15. Localised spectroscopy 66 kHz Tip FT-IR AFMIR FFT amplitude (nm) Absorbance

  16. Chemical mapping s=1800 cm-1 AFM topography

  17. Chemical mapping s=1650 cm-1 AFM topography

  18. Example of light distribution in tree different bacteria : a) height=100 nm width = 1.5µm b) height=210 nm width = 1.5µm c) height=450 nm width = 2µm

  19. Mapping of amide I on a thin spectrally homogeneous sample(nearly homogeneous lighting) Topography AFMIR

  20. Virus T5 infecting E. Coli bacteria • Most of its weight is constituted by DNA • It has a protein envelope : capsid • The virus injects its DNA in the bacterium through its tail • The DNA is replicated inside the infected cell • New capsids are synthesized • The process can be stopped at any stage • Cells were dried in this experiment

  21. Observation of an isolated virus T5 (Amide I band at 1650 cm-1 : capsid) AFM AFMIR

  22. Observation of isolated virus T5 (DNA band at 1050 cm-1 : the dried virus has expelled its DNA and looks diluted) AFM AFMIR

  23. Observation of an heavily infected bacterium Topography (AFM) AFMIR (bande de l’ADN)

  24. Observation of a single virus infecting a bacterium AFMIR (ADN bande) Topographie (AFM)

  25. Zooming on the single virus

  26. First experiments living cells (in water)(on candida albicans cells) Frequency response of the cantilever : On the cell in air In water : on the cell (red) on substrate (blue) ==> selecting the peak frequency (31,6 MHz) allows to image the cell

  27. AFM topography Mapping in water of living cells (candida albicans glued on the substrate) Wavelength : 1080 cm-1 AFMIR Topography at 80 kHz AFMIR Topography at 31.6 kHz

  28. Why is the observed spatial resolution so good ? • AFMIR spatial resolution is < 100 nm • limited by tip size, sample characteristics, S/N ratio • One expect heat diffusion of @1 µm/µs (speed of sound) • Measurement is done in more than tens of µs • ===> surprising that resolution << 1 µm • Simulations show that the heat is diluted in • the substrate and in air • ===> Almost no signal outside the absorbing zone

  29. COMSOL simulation of cooling of an hemispherical object ==> After excitation of the cantilever vibration, the heat is diluted extremely rapidly in the substrate (but too many excited samples on the substrate can “pollute” the signal”)

  30. COMSOL simulation of transient dilatation around the virus The spatial resolution is limited at @ 200 nm in the case of the « buried » virus

  31. Application : absorption nanospectroscopy of a single quantum dot Excitation 800 nm InAs/GaAs self-assembled quantum dot Quantum dots are “buried” and cannot be observed by AFM Density : 400 dots per µm2 20 nm diameter n-type doping Resonant absorption in the mid-infrared Single quantum dot absorption @ room temperature lexc = 9.6 µm Measurement performed with CLIO + AFMIR Spatial resolution ~ 80 nm J. Houel, S. Sauvage, P. Boucaud

  32. Conclusion:AFMIR + CLIO/OPO capabilities • Powerful tool to measure local relative absorption • Good spectral resolution (fit well FT-IR spectrum) • High spatial resolution (< 100 nm << linfrared) • Various applications underway