An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Shielding Workshop for Diagnostic Imaging Professor Jim Malone Trinity College, Dubin Robert Boyle Foundation

  2. Overview of posible topics • A Few CONTEXT Issues • Five Step Shielding Calculations • Leading questions, Socratic Method

  3. Context: Equipment

  4. Context: Makes Problems for Facility Design • Changes in Legislation, Technology, Clinical Practice, Building Style, Building Materials • On whose behalf are you working? Hospital management or the public interest? • Changes in Dose Limits and Constraints • Practical Tips and Solutions

  5. Context: New Problems • Walls not always solid • Use of Upper Floors • Ceilings, Floors, Higher levels of walls • Theatre workloads not consistent with modern practice. • Other – recovery rooms, endo suites, lithotripsy, cardiac pacing • Radionuclides in Theatres

  6. Context: Medical Physics Medical Physics Support • Deficits in numbers, leadership and academic involvement/connectivity • Often poor, or inadequate training, and narrow or inadequate experience • Research Output: Poor to None • Risk of litigation and difficulties with public accountability

  7. Published, 1988. www.rpii.ie Revised, 2009. RPII, 3 Clonskeagh Square, Dublin 14, Ireland

  8. Five Step Shielding Calculations • Identify boundary to be shielded • Determine design dose to be achieved beyond the boundary • Estimate the dose/dose rate before the boundary • Read the shielding required to attenuate dose in 3 to dose in 2 from graphs or tables • Practical Implementation issues

  9. 1. Identify boundary to be shielded

  10. *

  11. Mobile X-Ray Equipment

  12. 2. Determine design dose to be achieved

  13. Dose Limits and Constraints Design Limit is generally not a matter of Discretion, it is usually prescribed by law of formal advice.

  14. DOSE CONSTRAINTS • Prospective; Used in Planning; Verifiable retrospectively. • (EU BSS, 96/29) Definition: -- A restriction on the prospective dose to individuals which may result from a defined source, for use in the planning stage in radiation protection whenever optimization is involved. • Article 7: • Should be used when appropriate, within optimization • May be in Guidance established by member states. • In many regulatory documents (BSS/EU)

  15. Some EU DOSE Constraints (Public)

  16. Occupancy

  17. Occupancy • Occupancy of adjoining areas to be assessed • Try and get real information • Consider rooms on other side of corridor • Extremes: Office, 100%; Unattended car-park, 2.5 to 5 % • Reservation about NCRP door value in new builds, and Remember: • Transparency and accountability to the public • Defendable legally (reasonable patient, not doctor or physicist)

  18. Clear Mind and Objectives • We have to be clear on what the design objective is; • Everything else flows from it • Public demands transparency and accountability • The rest is science

  19. Two Types of Radiation • Primary Beam: • Collimated from the x-ray tube. Intercepted by the patient, image receptor and some beam-stopping heavy shielding. • In a well designed facility, now seldom a problem. • Scatered or Secondary Radiation: • Main source is the patient, and is normally main source of radiation in a room.

  20. Dose/dose rate before boundary • Primary often not an issue outside of plain radiography and mobiles

  21. Scattered Radiation • When X-Ray beam interacts with the patient some is scattered (Compton). • Room shielded to protect those outside it, ie, workers, the public, other patients et al. • Level of shielding required depends on many factors, including the type of equipment. Also the distance (i.e. room size), workload in room, and adjacent occupancy.

  22. Workload and Shielding Calculations • BIR: Workload is mainly based (ESD) and (DAP) • NCRP: Workload based on “beam-on” time. in mA min per week X-Ray • Two widely used Methodologies • BIR, 2000 • NCRP, 2004 • Variables • Distance from Barrier • Workload • Occupancy Issue of New Build versus Conversion/Refit

  23. Workload • Either DAP (BIR) or mA-min/week (NCRP) • Try and base figures on real information, audit/projection • Historical or published data can mislead • If no other option, use published data • Transparency/accountability to public • Defendable legally (reasonable patient, not reasonable doctor)

  24. Workload

  25. Workload

  26. Room Design and Layout • Radiology Room Types • General Comments on Shielding • Radiography Rooms • General • Chest Room • Mammography • DXA • Dental • Fluoroscopy Rooms • General Fluoroscopy • Interventional Radiology and Cardiology • CT • Shared Function Rooms (A&E, Theatres, wards) • Equipment in Trailers

  27. General Room • Two-corridor Design • Large enough for trolleys, table and chest radiology • Typical room sizes given • Generally 2mm Lead – assess on individual basis • Primary Beam absorber • Staff entrance behind protective screen • Typical screen lengths presented • Chest stand positioned to minimise scatter entering protective console • Changing cubicles Specific Requirements for each type of facility Examples of Good Layouts

  28. Dental Surgery and CT Room Dental Surgery No shielding required if: < 20 exps/wk and 2m between patient and all boundaries CT Separate staff area – Other staff present Need good view of door and patient Scanner angled for access and visibility MSCT: 3-4mm Pb

  29. DXA and CT

  30. Equipment not in Rooms

  31. Equipment not in Radiology Rooms

  32. Equipment not even in Building • Trailer • Re-shielding often means complete refit • Hospital can’t do without equipment for 6 months • Trailer arrives -----

  33. CT Room Shielding • Early understanding – no shielding (1970s) • BIR/IPEM Report (2000) example 1.7mm Pb • NCRP Report 147 (2004) example 1.4mm Pb • NHS Estates (HBN 6, 2001), 3 - 4mm Pb

  34. 5.2 2.6 1.3 0.6 0.6 1.3 2.6 5.2 CT Isodose Profiles BIR / IPEM: “Radiation Shielding for Diagnostic X-rays”, Sutton D.G. and Williams J.R. (2000) Scatter Profile (μSv/100mAs/20mm Slice)

  35. 1. Superimpose Scatter profile on Plans. Find dose at boundaries 2.6μSv

  36. Ceiling 5.2 Floor Similarly for Ceiling, Floor & Windows

  37. 5.2 Gantry Tilt may increase exposure of ceiling floor

  38. 2. Multiply Boundary Dose • by CT Workload. • Typical Exposure Factors • Head: 120kVp, 350mAs, 15cm • Body: 120kVp, 250mAs, 40cm 130μSv / body scan

  39. 2. Multiply Boundary Dose • Workload. • Typical Exposure Factors • Head: 120kVp, 350mAs, 15cm • Body: 120kVp, 250mAs, 40cm • (b) Workload, eg.100 per week 650mSv / year

  40. 3. Find Target Dose in Adjacent areas. • Target Dose – 0.3mSv (BIR / IPEM) • Occupancy (BIR / IPEM) Corridor, 20%, 1.5mSv Control Room, 100% 0.3mSv Office 100% 0.3mSv Ward 20% 1.5mSv

  41. 4. Find Maximum Transmission for each Boundary 0.3mSv / 650mSv =5x10-4

  42. 4. Read the shielding required from graphs or tables

  43. 5. Find thickness of lead / concrete (not Barium) 1x100 BIR/IPEM Report Secondary Transmission 150kVp 1x10-1 Attenuation 1x10-2 1x10-3 1x10-4 1x10-5 2 1 3 4 0 Lead thickness (mm)

  44. Some Data 2:

  45. General Radiographic Room Ceiling (BIR method)

  46. General Radiographic Room Ceiling (NCRP method) General Rm Window, scatter only, at 10 m

  47. 5. Practical Implementation issues

  48. 6. Practical Considerations • Walls • Floors and Ceilings • Doors • Windows • Staff Areas • Joints, Services, Openings and Perforations • Assessment of Shielding • Nuclear Medicine • Building Materials • Lead sheet and lead products • Concrete and concrete Blocks • Barium Plaster • Brick • Gypsum Wallboard • Lead Glass • Lead Acrylic ALWAYS NEED TO VISIT AND SEE IMPLEMENTATION