An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Generating time domain strain data (h(t)) for the ALLEGRO resonant detectororcalibration of ALLEGRO data Martin McHugh Loyola University New Orleans on behalf of the ALLEGRO group http://sam.phys.lsu.edu/ GWDAW-9, Annecy 16 December , 2004

  2. Outline • Motivation • Signal flow diagram, transfer function equations • Discussion of calibration measurements • Noise curves, stability • future GWDAW-9, Annecy 16 December , 2004

  3. Motivation • LSC stochastic background analysis using S2 data from ALLEGRO and LIGO Livingston (see John Whelan’s talk later today) • Unlike an event list based search, a coherent search such as this requires a phase consistent response function for the detector signal path GWDAW-9, Annecy 16 December , 2004

  4. ALLEGRO schematic GWDAW-9, Annecy 16 December , 2004

  5. Signal path GWDAW-9, Annecy 16 December , 2004

  6. Form of each Transfer Function GWDAW-9, Annecy 16 December , 2004

  7. Need to determine -- So in practice the calibration amounts to -- • Mode frequencies and Q’s -- fm , fp , Qm , Qp • overall scale -- in practice we measure •  is mechanical gain -- includes ‘tuning factor’ • Lock-in amplifier parameters -- gain, filter delay and phase shift • Also need to know the phase of the lock-in reference oscillator Inverse fft then gives complex heterodyned strain time series GWDAW-9, Annecy 16 December , 2004

  8. transducer current changed temperature fluctuations S2 data -- 2003 GWDAW-9, Annecy 16 December , 2004

  9. New calibrator • One plate of capacitor is tightly coupled to bar. • Other plate is weakly coupled to the bar, so acts like a free mass. • Both plates electrically isolated. GWDAW-9, Annecy 16 December , 2004

  10. The calibrator mounted on the bar GWDAW-9, Annecy 16 December , 2004

  11. Transfer function - white noise excitation to measured output measurements from 20 March 2004 -- excitation measured through lock-in and A/D plotted here we have gives us the overall scale -  GWDAW-9, Annecy 16 December , 2004

  12. GWDAW-9, Annecy 16 December , 2004

  13. Lock-in measurements Band-limited white noise injection -- recorded directly and through lock-in/anti-aliasing filters GWDAW-9, Annecy 16 December , 2004

  14. Compare fourier coefficients Lock-in/filter introduces an 11ms delay and 18 degree phase shift GWDAW-9, Annecy 16 December , 2004

  15. GWDAW-9, Annecy 16 December , 2004

  16. Calibration line Detector resonances Extra mechanical resonances GWDAW-9, Annecy 16 December , 2004

  17. Calibrated strain spectrum from S2 GWDAW-9, Annecy 16 December , 2004

  18. Stability of calibration GWDAW-9, Annecy 16 December , 2004

  19. Summary, future • We have calibrated h(t) for S2 data set at level • Currently stored in Matlab files, plan to put these data into frames • detector much more stable now, should continue through S4 • will determine overall sign • hardware injections are planned GWDAW-9, Annecy 16 December , 2004