Ionization Chamber Array for External Beam Radiotherapy. Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - Septem ber, 30 th 2014. Project goal. A short summary. MV X-Rays. ~100÷230MeV protons. Absolute dose measurement Machine quality assurance Patient quality assurance.

Ionization Chamber Array for External Beam Radiotherapy - slide 1
01 / 23Ionization Chamber Array for External Beam Radiotherapy
1 / 17

Télécharger la présentation

Ionization Chamber Array for External Beam Radiotherapy
Une image / un lien ci-dessous est fourni (tel quel) pour télécharger la présentation

Politique de téléchargement : le contenu du site vous est fourni TEL QUEL pour votre information et votre usage personnel ; il ne peut être vendu, concédé sous licence ni partagé sur d’autres sites sans l’accord de son auteur.Le contenu vous est fourni TEL QUEL, pour votre information et votre usage personnel uniquement. Téléchargez la présentation en cliquant sur ce lien. Lors du téléchargement, si vous ne parvenez pas à télécharger une présentation, l’éditeur a peut-être supprimé le fichier de son serveur.Pendant le téléchargement, si vous ne parvenez pas à obtenir une présentation, le fichier a peut-être été supprimé par l’éditeur.

F I N

Plus de contenu associé

Transcription de la présentation
  1. Ionization Chamber Array for External Beam Radiotherapy Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014

  2. Project goal A short summary MV X-Rays ~100÷230MeV protons • Absolute dose measurement • Machine quality assurance • Patient quality assurance

  3. Outline • Experimental Activity • Characterization at IBA DosLab • Characterization at Klinikum rechts der Isar & University of California (San Francisco)-Radiation Oncology dep. • Characterization at Proton Therapy Center Czech s.r.o. • Ongoing activities and future work • ARDENT project: trainings, conferences, secondments

  4. Experimental Activity Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014

  5. Detector evolution First Prototype Measurement setup • 80 ionization chambers, 3.5mm pitch Second Prototype

  6. Characterization at IBA DosLab Prototype optimization Main debug issue: time stability of the chamber current. Example: two 60Co irradiations (detector biased at 100V). Current ~ 10pA Signal (normalized) 0.92 0.96 1.00 1.04 Signal about 8pA (@2.2Gy/min) 0 50 100 150 200 250 300 Time (s) Signal (normalized) -0.2 0 0.2 0.4 0.6 0.8 1 1.2 • Optimization of signal routing led to a better time stability of the sensor: • Pretty good stability in time of the signal • Prompt rise and fall time 60Co, 150V bias, Tint=0.8s Sensitivity  200pC/Gy Fluctuation  0.3% (st.dev.) 0 20 40 60 80 90 100 120 40 0 -20 60 Time (s)

  7. Characterization at IBA DosLab Charge collection efficiency • Charge collection efficiency higher than 99% already at 150V (99.5% ± 0.3% at 250V)

  8. Characterization at IBA DosLab Sensitivity dependence on dose per pulse • ± 0.8% sensitivity change on dose per pulse in the range 0.09÷2.67mGy/pulse

  9. Characterization at IBA DosLab Tissue to phantom ratio measurement • Very good agreement with PPC05 (better than 0.5%) at every depth of measurement

  10. Characterization at IBA DosLab Elekta Agility Output Factors • Good agreement with reference for small fields, reduced sensitivity with large fields -3.7% -2.4% a:Si flat panel, a prototype optimized for dosimetry

  11. Characterization at IBA DosLab Beam profiles (photons)

  12. Characterization at MRI Machine QA: virtual wedges, MLC QA, dose in depth (see pics) MapCheck2 (SunNuclear) diode array Varian Clinac DHX • Secondment, 22nd April - 3rd May, Klinikum rechts der Isar (Munich)

  13. Characterization at MRI Example of patient QA Varian Trilogy, Millennium MLC, VMAT delivery technique EBT3

  14. Characterization at UCSF Clinical tests & clinical training • Secondment, 28th July - 8th August, UCSF department of Radiation Oncology; • Detector characterization through clinical test; • Clinical training: • Machine QA; • Introductory training to Pinnacle TPS; • Use of gafchromic films; • Joined the daily activity of medical physicists.

  15. Characterization at UCSF Example of small field dosimetry and linac commissioning 1x1cm2 Varian TrueBeam Virtual wedge 45°

  16. Characterization at UCSF Example of linac commissioning measurement Water phantom as reference

  17. Characterization at PTC czechs.r.o. Dose linearity PBS mode only 1ms pulse duration, 10ms pulse period, 3.5mm spot  at 226MeV 0.02-2MU/spot 33x17 spot map Output norm. to 1MU/spot 0.01% -0.04% -0.05% -0.1% Ref detector: PPC05 0.8% -0.9% IC array signal (pC) 1.7% PPC05 signal (nC)

  18. Characterization at PTC czechs.r.o. Beam profiles (protons) Lynx: a scintillator based detector for PBS mode machine QA

  19. Ongoing activities/Future work • Long term stability measurements with 60Co • Further test with PBS proton beams; • Paper about the characterization of the 1D IC array prototype; • Extend the technology to a 2D detector - components drawing already started, first prototype assembling foreseen for the end of the year.

  20. Trainings, conferences, secondments… Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014

  21. Secondments UCSF RadOnc dep., S.Francisco (28.07 - 08.09.2014) Klinikum rechts der Isar, Munich (22.04 - 03.05.2014) • Upcoming: • B&A secondment at STMicroelectronics (Catania, July 2015 tentative) • UCSF (early 2015 tentative)

  22. Conferences, trainings… • AAPM, Austin (TX), 20.07 - 25.07.2014 • Poster presentation: “An Ionization Chamber Array with High Spatial Resolution for External Beam Radiotherapy”, M.Togno, D.Menichelli, J.J.Wilkens. • Upcoming: AAPM 2015, … • Trainings: • Clinical trainings at MRI & UCSF (already mentioned); • Radiation protection training (IBA Dosimetry); • Training on SBRT radiotherapy technique (Klinikum rechts der Isar); • ARDENT B&A training (CERN): Technology Transfer, Management, Entrepreneurship. • PhD activities:already spent more than 2 weeks at the University to teach students at physics labs.

  23. Thank you! Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014