1 / 54

Muon Tracking : Tutorial

Muon Tracking : Tutorial. Updated : September 12th 2008. Introduction Log on the muon_trk machine 1) Data Acquisition 2) Detector Control System 3) Mood 4) logbook Expert Tools (lc2gui, ltu) FAQ and frequent problems. Introduction (i) .

Télécharger la présentation

Muon Tracking : Tutorial

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Muon Tracking : Tutorial Updated : September 12th 2008 • Introduction • Log on the muon_trk machine • 1) Data Acquisition • 2) Detector Control System • 3) Mood • 4) logbook • Expert Tools (lc2gui, ltu) • FAQ and frequent problems

  2. Introduction (i) • A preliminary remarks; during december 2007 cosmic the number of people in the ACR was very high. Please do not consider the ACR as a working room. Seats are reserved for shifters (and experts when specific software developments are required). • This (basic) tutorial is divided in 3 main sections : • Take control of the DAQ • Configure the chambers included in the readout (DCS) • Check data quality with mood

  3. Introduction (ii) • First you will have to log onto our dedicated machine in the ACR (Alice Control Room) : • aldaqacr25 : username is muon_trk (you will get the password during your training) • On this machine, one task = a desktop (please keep this organized this way): • 1- DCA (Detector Control Agent) and DAQ • 2- DCS (Detector Control System) • 3- Monitoring and MOOD • 4- Alice Web : logbook, alicesms, etc.. • We’ll start with DAQ and DCA…

  4. 1) Data Acquisition • To start the DAQ interfaces the MTRK machine in ACR (aldaqacr25), go to desktop 1. If nothing is present on the screen , you can start the interfaces by clicking on • this icon • Then 3 windows will pop up…

  5. 1) Data Acquisition • The infoBrowser display all the info and error messages. Make sure it is online and the Stream selected is MUON_TRK. • You can also add some information like the run number

  6. 1) Data Acquisition • The Detector Control Agent from which you will issue the command for data taking : Take control of the DCA with a click on the icon (if the lock is red, another DCA is open or central DAQ has control) Take control of the DAQ_RC (Run Control ) with a click on the icon. If the lock is red, a RC window is open and own the interface… slide14 If DAQ_RC status is not ready (most probably it would be disconnected), see next slide

  7. 1) Data Acquisition • The Detector Control Agent from which you will issue the command for data taking : If DAQ_RC status is disconnected, you can connect by requesting the DAQ resources. Click on GET_DAQ_RESOURCES

  8. 1) Data Acquisition • The Detector Control Agent should now look like this :

  9. 1) Data Acquisition • The run sequence need the execution of several commands; here a few explanations about the commands. The sequence is given in the next slide (assuming the detector configuration via DCS has been properly done ) • STANDALONE_RUN : start a run. Make sure LDC local recording is off (unless you really need it). GDC : No recording (recording in root format require mStream Recording). • PEDESTAL_RUN : start a pedestal run (detector should be in BEAM_TUNING) • ELECTRONICS_CALIBRATION_RUN : start a 10 calibration runs (detector should be in BEAM_TUNING). About 20 minutes. • RESET_FERO • COMPUTE_THRESHOLDS • LOAD_CROCUS_SW : this command can not be executed twice in a row without power cycling the readout (crocus) • RELEASE_RESSOURCES

  10. 1) Data Acquisition • From the STANDY_CONFIGURED state (the detector is completely off). • Commands to be sent are : • GO_BEAM_TUNING • LOAD_CROCUS_SW (this command should be executed only if the detector was previously in STANBY_CONFIGURED) • PEDESTAL_RUN  • CHECK_FERO (if necessary, normally executed a the end of  the pedestal run) • GO_READY • If the detector is in BEAM_TUNING state. The only necessary • operation to be done is • GO_READY • During runs the detector will remain in the READY state. From time to time it is necessary to start a pedestal run; the shifter should  go to the ECS operator and request a pedestal run. Commands are : • GO_BEAM_TUNING • PEDESTAL_RUN • CHECK_FERO (if necessary, normally executed a the end of  the pedestal run) • GO_READY

  11. 1) Data Acquisition • The third window is a “control panel”. The infoBrowser and the ECS button will start the infoBrowser window and the DCA window ; both are already open. • The runControl button will start the DAQ interface. • The select equipment button will allow you to select the part of the detector that are included into the readout. CLICK on it

  12. 1) Data Acquisition • This is the list of the equipment id or DDL for the muon tracker. Here the DDL 2560, 2561 and 2563. Each time you change the selected DDL you must commit and then quit. With the DAQ interface (after clicking on the runControl button), you will need to select the ldc corresponding to the equipId you’ve selected. This will be explained in a few slides…

  13. 1) Data Acquisition • A few words about the numbering scheme of the readout : • 1 LDC receive the data corresponding to 1 full station : • id: 53 ldc-MTRK-S1-0 hostname:aldaqpc088 station1 • id: 54 ldc-MTRK-S2-0 hostname:aldaqpc089 station2 • id: 55 ldc-MTRK-S3-0 hostname:aldaqpc090 station3 • id: 56 ldc-MTRK-S4-0 hostname:aldaqpc091 station4 • id: 57 ldc-MTRK-S5-0 hostname:aldaqpc092 station5 • 1 full station has 4 DDL (or CROCUS) : 1 LDC read 4 CROCUS • ldc-MTRK-S1-0 read DDL or equipment 2560,2561,2562,2563 • ldc-MTRK-S2-0 read DDL or equipment 2564,2565,2566,2567 • ldc-MTRK-S3-0 read DDL or equipment 2568,2569,2570,2571 • ldc-MTRK-S4-0 read DDL or equipment 2572,2573,2574,2575 • ldc-MTRK-S5-0 read DDL or equipment 2576,2577,2578,2579

  14. 1) Data Acquisition • Up to date values are now located here : • https://twiki.cern.ch/twiki/bin/view/ALICE/CablingAndNumberingScheme • 2560 (EquipId): Chamber 1 Right (DCS) [rorc minor = 3; rorc channel = 0] • 2561 : Chamber 1 Left [rorc minor = 0; rorc channel = 0] • 2562 : Chamber 2 Right [rorc minor = 1; rorc channel = 0] • 2563 : Chamber 2 Left [rorc minor = 2; rorc channel = 0] • 2564 : Chamber 3 Right [rorc minor = 3; rorc channel = 0] • 2565 : Chamber 3 Left [rorc minor = 0; rorc channel = 0] • 2566 : Chamber 4 Right [rorc minor = 1; rorc channel = 0] • 2567 : Chamber 4 Left [rorc minor = 2; rorc channel = 0] • 2568 : Chamber 5 & Chamber 6 Top Left [rorc minor = 3; rorc channel = 0] • 2569 : Chamber 5 & Chamber 6 Top Right [rorc minor = 0; rorc channel = 0] • 2570 : Chamber 5 & Chamber 6 Bottom Left [rorc minor = 1; rorc channel = 0] • 2571 : Chamber 5 & Chamber 6 Bottom Right [rorc minor = 2; rorc channel = 0] • 2572 : Chamber 7 Right [rorc minor = ; rorc channel = 0] • 2573 : Chamber 7 Left [rorc minor = ; rorc channel = 0] • 2574 : Chamber 8 Right [rorc minor = ; rorc channel = 0] • 2575 : Chamber 8 Left [rorc minor = ; rorc channel = 0] • 2576 : Chamber 9 Right [rorc minor = ; rorc channel = 0] • 2577 : Chamber 9 Left [rorc minor = ; rorc channel = 0] • 2578 : Chamber 10 Right [rorc minor = ; rorc channel = 0] • 2579 : Chamber 10 Left [rorc minor = ; rorc channel = 0]

  15. 1) Data Acquisition • The third window is a “control panel”. The infoBrowser and the ECS button will start the infoBrowser window and the DCA window ; both are already open. • The runControl button will start the DAQ interface. CLICK on it

  16. 1) Data Acquisition The lock is red because the control of the DAQ interface is taken by the DCA interface. To control this panel, release the control from the DCA window.

  17. 1) Data Acquisition • Since Friday 15th (Feb. 2008), the bridge between the ECS and DCS has been successfully tested. It simply means that the DCS (HV,LV) is controlled by ECS. This box represents the real detector state. The lock allows to release/take the control. From the pull down menu, it is possible to change the state of the detector FSM. More on the FSM next slide…

  18. FSM : Final State Machine • Our detector has 3 main states : • STANDY_CONFIGURED (off) • BEAM_TUNING (LV on, HV @intermediate value == no gain) • used for pedestal and electronic calibration runs • READY (HV@ at full value and LV on) • get ready for physic + 1 additional state which doesn’t change the HV or LV configuration : • READY_LOCKED • state is lock ; data taking can start (the state ready may not be sufficient). • Rising the low and high voltages (like going from standby to ready) can be harmful to the detector and need to be careful prepared and checked : • select the active parts of the detector corresponding to ones included in the data taking • check the high voltages values • This is done via the DCS interface…

  19. 2) Detector Control System • First go on desktop #2 and start the DCS. Most probably the DCS UI interface will be already open, then go directly to slide 22. • If not : • Start a xterm and launch tsclient (connection to windows machines to start the DCS interface) • >tsclient •  Open •  Choose dcs and ok •  Connect Use your personal NICE “username” and “password” to log in on the DCS application gateway

  20. 2) Detector Control System • Open the Remote Desktop Connection from here and connect to “Operator Node” (computer ALIDCSCOM048). • Type “alidcscom048” and press Connect Use your NICE “username” and “password” to log in on the DCS “Operator Node” server. Once on the machine start the program mchdcsui

  21. FSM NODE STATE TITLE BAR DATE-TIME LHC STATUS USER LOGIN DETECTOR GRAPHIC LOCATOR ENVIRONMENT PARAMETERS USER TOOL BAR DETECTOR LOGO FSM NODE CONTROL USER PANEL FOR MONITORING FSM HIERARCHY TREE BROWSER MESSANGER READER AUXILIARY MONITORING ZONE OF THE FSM MAJOR NODES HOSTS STATE & FSM MAIN CONTROL CLOSE WINDOWS 2) Detector Control System The Standard DCS UI provided by the ACC

  22. 2) Detector Control System • After several seconds, when finished to download the FSM Tree, you must type your user name and password to log in to the DCS system. This user name and password are only for DCS and are different from the NICE ones. You must click on the “Enter” after user name and password. If you are member of some group (Operators, Experts ..) will be opened the “User Panel” for Muon DCS. More information about every part of User Panel one can found in: http://alicedcs.web.cern.ch/AliceDCS/Software/Downloads/AliceDcsUi_v3.0.doc

  23. 2) Detector Control System

  24. 2) Detector Control System FSM Gas Temperature See chambers LV and HV : right click and [view panel] Do not Touch Low Voltages Left side Low Voltages Right side HV

  25. 2) Detector Control System • The top panel (previous slide) of the DCS interface gives a summary of most of the variables that we can control and/or monitor : • FSM • gas flow • temperature monitoring • low voltages (Left and right side) • high voltages view • half chamber view (left side) to access the a particular part of the detector (from which low and high voltages values can be set)

  26. 2) Detector Control System • FSM : • which parts of the detector will be included in the run ? It is required to power up on only the DDL that participate in the DATA taking (this has been defined previously with the select of the equipment Id). If not done properly, done the detector will appear in BUSY state to the DAQ. • the mechanism to include/exclude chambers of the muon spectrometer uses the lock icons associated to each ½ chambers. • the hierarchy of the FSM is the following : MCH_DCS MCH_DCS_RUN MchHvLvLeft  MchHvLvRigth • mchChamber01Rigth • mchChamber02Rigth • mchChamber03Rigth • …. • mchChamber01Left • mchChamber02Left • mchChamber03Left • …..

  27. 2) Detector Control System • The correspondence between the naming scheme of the DCS and the • naming scheme of the DAQ is listed on slides 13 and 14 ( e.g. Chamber 4 Right = 2566, etc…). • So how to include/exclude in the DCS the part of the detector declared in the equipment ID list of the DAQ (slide 12)? • First, you need to pop up the FSM window corresponding to the Top Node MCH_DCS of the Dimuon Spectrometer : Click precisely on the border line surrounding the element (the pointer of the mouse will change from an arrow to a hand).

  28. 2) Detector Control System Click on the lock to see the menu of the possible actions A pop-up window appears : you can take the control of the detector ( then the lock close and become green, as displayed on the left). If you click again on the lock, another window appears : you can release the control of the detector (to release the ownership prefer ReleaseAll to Release.) Now you can close the Top Node (MCH_DCS) FSM window ; the main panel should now look like this…

  29. 2) Detector Control System All the chambers are now selected, which certainly doesn’t correspond to reality : only parts of the detector (specific DDL or EquipId) are readout. Thus the others have to be excluded, according to the selection done in the DCA (slides 12,13 and 14). This will done from the level below the Top Node (see slide 26), MchHvLvLeft and MchHvLvRigth

  30. 2) Detector Control System All the left side of the spectrometer is selected. Let’s remove Chamber03 for an example ; simply click on the lock… In the pop-up window, click on Exclude to remove this chamber from the DCS control.

  31. Chamber03 is deselected; one more click on the lock will allow to lock it out. In the pop-up window, click on LockOut to lock out this chamber when DCS control will be given to the DCA. This operation is crucial to have the correct part of the detector powered up. You exclude and lock out the other chambers on the left side. Then you will go to the rigth side.

  32. This was the default configuration during the February/March cosmic run. Full station 1, full station 2 but 1 DDL (2565). Nothing from Station 3, 4 or 5.

  33. 2) Detector Control System FSM Gas Temperature See chambers LV and HV : right click and [view panel] Do not Touch Low Voltages Left side Low Voltages Right side HV

  34. 2) Detector Control System : High voltages settings • It is crucial to check the high voltage setting of the chamber. A right click on a the chamber name (on the left side of the DCS panel) will allow to view its “voltage panel”. Click on the “Channel Set” button to pop up the HV settings panel.

  35. 2) Detector Control System : High voltages settings • This is the table describing the high voltages settings. Click here to set the all values of a column at once Do not modify these values !! Ask an expert if something doesn’t looks normal to you. Up to date guide/values is now located here : https://twiki.cern.ch/twiki/bin/view/ALICE/HighVoltages

  36. 2) Detector Control System : temperatures probes Click here to see the list of sensors and the temperature measured. • Shifter should check : • temperatures values • in a normal operating mode Tmax ~ 30

  37. 2) Detector Control System : temp. probes Up to date information is available at : https://twiki.cern.ch/twiki/bin/view/ALICE/TemperatureSensors

  38. 2) Detector Control System : GAS system Rack 61 provide gas to station 1 and 2. Rack 62 provide gas to station 3, 4 and 5. • Shifter should check : • Status of the rack • flow in and out • pressure If the gas flow (mix of Ar + C02) stops, a hardwired interlock (not wired yet) will shutdown the HV after a short delay (~10 minutes). Without the proper gas , the gain of the chamber is not adapted to particle detection.

  39. 3) Monitoring and MOOD • Data monitoring is done with MOOD. The machine to muon tracking is aldaqdqm07 (accessible from acr machines, user muon_trk same password). • Once logged on, simply type > mood& • (a development version can be selected with : • > source ~/Guillaume/mood_env.sh ) • On the mood window, “Detector”  “MCH (MUON Tracker)”  “Cosmic Test 2007” ( this name may change, in any case do not use “Test Beam” ) • After a few seconds….

  40. 3) Monitoring and MOOD

  41. 3) Monitoring and MOOD • Configuration is accessible from the tabs : “Setup Monitor” and “MUON Tracker Configuration”. Do not change this configuration unless you know what you are doing ! (for non-gdc monitoring, you need to have the correspondance between the monitoring source and the LDC.)

  42. 3) Monitoring and MOOD Monitoring histograms are grouped in the display tab Summary histos have give occupancy and errors distribution Manu Dump will allow you to see the raw data at the MANU level

  43. 3) Monitoring and MOOD • Look at pedestal/noise and gains files Pedestals and calibration runs are local to a LDC, i.e.to a station. After a pedestal or a calibration run, resuls files will be located in the /tmp of the ldc: pedestal file = /tmp/MUONTRKda_ped_ {rn}.ped calib file = /tmp/MUONTRKda_gain_ {rn}.ped (where {rn} stands for the appropriate run number). NB : all files from different LDC have the same name. So these files need to be transferred from the LDCs aldaqpc088,089,090,091,092 to the mood machine with a different name, to do so : On the monitoring machine (aldaqdqm07), > cd ~/CurrentRun/ > rm *.ped On the aldaqacr machine : > cd ~/ > source copyPedRun 12345 (where 12345 is the proper ped. run number ) Pedestal subtraction can then easily be applied to the data

  44. 3) Monitoring and MOOD • Look at pedestal/noise and gains files Pedestals and calibration runs are local to a LDC, i.e.to a station. After a pedestal or a calibration run, resuls files will be located in the /tmp of the ldc: pedestal file = /tmp/MUONTRKda_ped_ {rn}.ped calib file = /tmp/MUONTRKda_gain_ {rn}.ped (where {rn} stands for the appropriate run number). NB : all files from different LDC have the same name. So these files need to be transferred from the LDCs aldaqpc088,089,090,091,092 to the mood machine with a different name, like : It is not yet possible to reset the pedestal/noise and gain files from mood “memory”. Thus before visualizing a new run, one should clean the directory CurrentRun and restart mood.

  45. 4) Expert tool : [LC]2 • This is the tool that will allow you to modify the configuration files (init{DDL.number}.dat) of the detector sent to each crocus. • 1- go to the working directory (> cd ~/LC2work) • 2- start [LC]2 with the command lc2gui You want to use the “Edit” mode.

  46. 4) Expert tool : [LC]2 • Modify and save a configuration file : Edit mode • The default configuration of the CROCUS, generated from the AliRoot mapping is in ~/LC2ConfigFromAliRoot (numbered from 0 to 19, you need to add 2560 to get the real crocus id) : • Copy the original crocus in the working directory • open it with LC2, • modify it to reflect the real state • of the corresponding part • of the detector • A very good idea is to check FIRST with a • diff command the configuration stored in • the ~/LC2work/init{CrocusId}.dat with the • one use by the ECS : • > diff ~/LC2work/init{CrocusId}.dat • $LC2_CONFIG/init{CrocusId}.dat X X

  47. 4) Expert tool : [LC]2 • Modify and save a configuration file : Edit mode • The default configuration of the CROCUS, generated from the AliRoot mapping is in LC2ConfigFromAliRoot (numbered from 0 to 19, you need to add 2560 to get the real crocus id) : • Copy the original crocus in the working directory • open it with LC2, modify it to reflect the real state of the corresponding part of the detector • save it with the appropriate name init{CrocusId}_modified.dat • move it to the directory from which crocus configuration files are loaded by the ECS scripts > mv ~/LC2work/init{CrocusId}_modified.dat $LC2_CONFIG/init{CrocusId}.dat • After any operation with LC2, it is safer to do > rm /tmp/*.scr

  48. 4) Expert tool : [LC]2 • Creation of the original configuration file : • The original configuration file is first taken from aliroot and modified to reflect the active patch buses of the detecteur : • Set the proper rorc minor id • rorc channel number id is always 0 • Set the proper equipment id • Set the last bullet (CRT1 resets FRT board at init). This is NEW and has to be done for every file !

  49. 5) Expert tool : Trigger Emulator • Trigger emulator window is accessible from the DCA window ;  View  Show trigger Control  Configuration

  50. 5) Expert tool : Trigger Emulator • Default configuration of the MCH Local Trigger Unit • You may need to change • the trigger rate : • This is the “bcrate” • put a new value in hertz • Save • Quit the interface Do not modify anything in the trigger setup for the muon tracking… This can take days to be fixed ! If you “think” you have modified something, please contact me asap !!

More Related