Electronic Spectroscopy Ultraviolet and visible

Electronic Spectroscopy Ultraviolet and visible. Where in the spectrum are these transitions?. Why should we learn this stuff? After all, nobody solves structures with UV any longer!. Many organic molecules have chromophores that absorb UV

geigerd
Télécharger la présentation

Electronic Spectroscopy Ultraviolet and visible

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Electronic SpectroscopyUltraviolet and visible

  2. Where in the spectrum are these transitions?

  3. Why should we learn this stuff?After all, nobody solves structures with UV any longer! Many organic molecules have chromophores that absorb UV UV absorbance is about 1000 x easier to detect per mole than NMR Still used in following reactions where the chromophore changes. Useful because timescale is so fast, and sensitivity so high. Kinetics, esp. in biochemistry, enzymology. Most quantitative Analytical chemistry in organic chemistry is conducted using HPLC with UV detectors One wavelength may not be the best for all compound in a mixture. Affects quantitative interpretation of HPLC peak heights

  4. “hypsochromic” shift “bathochromic” shift  Uses for UV, continued Knowing UV can help you know when to be skeptical of quant results. Need to calibrate response factors Assessing purity of a major peak in HPLC is improved by “diode array” data, taking UV spectra at time points across a peak. Any differences could suggest a unresolved component. “Peak Homogeneity” is key for purity analysis. Sensitivity makes HPLC sensitive e.g. validation of cleaning procedure for a production vessel But you would need to know what compounds could and could not be detected by UV detector! (Structure!!!) One of the best ways for identifying the presence of acidic or basic groups, due to big shifts in  for a chromophore containing a phenol, carboxylic acid, etc.

  5. The UV Absorption process •   * and   * transitions: high-energy, accessible in vacuum UV (max <150 nm). Not usually observed in molecular UV-Vis. • n  * and   * transitions: non-bonding electrons (lone pairs), wavelength (max) in the 150-250 nm region. • n  * and   * transitions: most common transitions observed in organic molecular UV-Vis, observed in compounds with lone pairs and multiple bonds with max = 200-600 nm. • Any of these require that incoming photons match in energy the gap corrresponding to a transition from ground to excited state. • Energies correspond to a 1-photon of 300 nm light are ca. 95 kcal/mol

  6. π* π* π* Example for a simple enone π* π* π* π* π* π* n n n π π π π π π -*; max=218 =11,000 n-*; max=320 =100 What are the nature of these absorptions? Example:   * transitions responsible for ethylene UV absorption at ~170 nm calculated with ZINDO semi-empirical excited-states methods (Gaussian 03W): h 170nm photon LUMOg antibonding molecular orbital HOMOu bonding molecular orbital

  7. How Do UV spectrometers work? Rotates, to achieve scan Matched quartz cuvettes Sample in solution at ca. 10-5M. System protects PM tube from stray light D2 lamp-UV Tungsten lamp-Vis Double Beam makes it a difference technique Two photomultiplier inputs, differential voltage drives amplifier.

  8. Diode Array Detectors Diode array alternative puts grating, array of photosens. Semiconductors after the light goes through the sample. Advantage, speed, sensitivity, The Multiplex advantage Disadvantage, resolution is 1 nm, vs 0.1 nm for normal UV Model from Agilent literature. Imagine replacing “cell” with a microflow cell for HPLC!

  9. Experimental details What compounds show UV spectra? Generally think of any unsaturated compounds as good candidates. Conjugated double bonds are strong absorbers Just heteroatoms are not enough but C=O are reliable Most compounds have “end absorbance” at lower frequency. Unfortunately solvent cutoffs preclude observation. You will find molar absorbtivities  in L•cm/mol, tabulated. Transition metal complexes, inorganics Solvent must be UV grade (great sensitivity to impurities with double bonds) The NIST databases have UV spectra for many compounds

  10. 1.0 maxwith certain extinction  UV Visible Absorbance 0.0 400 800 200 Wavelength, , generally in nanometers (nm) An Electronic Spectrum Make solution of concentration low enough that A≤ 1 (Ensures Linear Beer’s law behavior) Even though a dual beam goes through a solvent blank, choose solvents that are UV transparent. Can extract the  value if conc. (M) and b (cm) are known UV bands are much broader than the photonic transition event. This is because vibration levels are superimposed on UV.

  11. Solvents for UV (showing high energy cutoffs) THF 220 CH2Cl2 235 CHCl3 245 CCl4 265 benzene 280 Acetone 300 Various buffers for HPLC, check before using. Water 205 CH3CN 210 C6H12 210 Ether 210 EtOH 210 Hexane 210 MeOH 210 Dioxane 220

  12. Organic compounds (many of them) have UV spectra One thing is clear Uvs can be very non-specific Its hard to interpret except at a cursory level, and to say that the spectrum is consistent with the structure Each band can be a superposition of many transitions Generally we don’t assign the particular transitions. From Skoog and West et al. Ch 14

  13. An Example--Pulegone Frequently plotted as log of molar extinction  So at 240 nm, pulegone has a molar extinction of 7.24 x 103 Antilog of 3.86

  14. Can we calculate UVs? Semi-empirical (MOPAC) at AM1, then ZINDO for config. interaction level 14 Bandwidth set to 3200 cm-1

  15. The orbitals involved Showing atoms whose MO’s contribute most to the bands

  16. P0 (power in) P (power out) The Quantitative Picture • Transmittance: T = P/P0 • Absorbance: A = -log10 T = log10 P0/P B(path through sample) • The Beer-Lambert Law (a.k.a. Beer’s Law): A =ebc Where the absorbance A has no units, since A = log10 P0 / P e is the molar absorbtivity with units of L mol-1 cm-1 b is the path length of the sample in cm c is the concentration of the compound in solution, expressed in mol L-1 (or M, molarity)

  17. Beer-Lambert Law Linear absorbance with increased concentration--directly proportional Makes UV useful for quantitative analysis and in HPLC detectors Above a certain concentration the linearity curves down, loses direct proportionality--Due to molecular associations at higher concentrations. Must demonstrate linearity in validating response in an analytical procedure.

  18. Polyenes, and Unsaturated Carbonyl groups;an Empirical triumph R.B. Woodward, L.F. Fieser and others Predict max for π* in extended conjugation systems to within ca. 2-3 nm. Attached group increment, nm Extend conjugation +30 Addn exocyclic DB +5 Alkyl +5 O-Acyl 0 S-alkyl +30 O-alkyl +6 NR2 +60 Cl, Br +5 Homoannular, base 253 nm Acyclic, base 217 nm heteroannular, base 214 nm

  19. b b O  x Similar for Enones 227 239 202 215 Base Values, add these increments… b g  d,+ X=H 207 X=R 215 X=OH 193 X=OR 193 With solvent correction of….. Water +8 EtOH 0 CHCl3 -1 Dioxane -5 Et2O -7 Hydrcrbn -11

  20. Some Worked Examples Base value 217 2 x alkyl subst. 10 exo DB 5 total 232 Obs. 237 Base value 214 3 x alkyl subst. 30 exo DB 5 total 234 Obs. 235 Base value 215 2 ß alkyl subst. 24 total 239 Obs. 237

  21. Distinguish Isomers! Base value 214 4 x alkyl subst. 20 exo DB 5 total 239 Obs. 238 Base value 253 4 x alkyl subst. 20 total 273 Obs. 273

  22. Generally, extending conjugation leads to red shift “particle in a box” QM theory; bigger box Substituents attached to a chromophore that cause a red shift are called “auxochromes” Strain has an effect… max 253 239 256 248

  23. Interpretation of UV-Visible Spectra • Transition metal complexes; d, f electrons. • Lanthanide complexes – sharp lines caused by “screening” of the f electrons by other orbitals • One advantage of this is the use of holmium oxide filters (sharp lines) for wavelength calibration of UV spectrometers. See Shriver et al. Inorganic Chemistry, 2nd Ed. Ch. 14

  24. Benzenoid aromatics UV of Benzene in heptane From Crewes, Rodriguez, Jaspars, Organic Structure Analysis

  25. Substituent effects don’t really add up Can’t tell any thing about substitution geometry Exception to this is when adjacent substituents can interact, e.g hydrogen bonding. E.g the secondary benzene band at 254 shifts to 303 in salicylic acid In p-hydroxybenzoic acid, it is at the phenol or benzoic acid frequency

  26. Heterocycles Nitrogen heterocycles are pretty similar to the benzenoid anaologs that are isoelectronic. Can study protonation, complex formation (charge transfer bands)

  27. Quantitative analysis Great for non-aqueous titrations Example here gives detn of endpoint for bromcresol green Binding studies Form I to form II Isosbestic points Single clear point, can exclude intermediate state, exclude light scattering and Beer’s law applies Binding of a lanthanide complex to an oligonucleotide

  28. More Complex Electronic Processes • Fluorescence:absorption of radiation to an excited state, followed by emission of radiation to a lower state of the same multiplicity • Phosphorescence:absorption of radiation to an excited state, followed by emission of radiation to a lower state of different multiplicity • Singlet state:spins are paired, no net angular momentum (and no net magnetic field) • Triplet state:spins are unpaired, net angular momentum (and net magnetic field)