420 likes | 518 Vues
Explore indoor calibration methods, hierarchy of traceability, uncertainties, & industry requirements for enhanced solar energy assessments.
E N D
State of the art of indoor calibration of pyranometers and pyrheliometers
Main points • Most field pyranometers are calibrated indoors • Many procedures for indoor calibration • Not all well connected to ISO 98-3 GUM • Industry requires straightforward approach
Industry • Meteorology - Solar renewable energy • Site assessment • Installation performance • Professionalisation / IEC
Conclusion • Points for discussion • Normal incidence calibration is preferred (diffuse dome not) • Uncertainty & accuracy of reference can be optimised • Pyrheliometer indoor calibration must be added
Myself • Kees VAN DEN BOS • Engineer Physics • Director / owner Hukseflux Thermal Sensors • Last 20 years sensor design
Founded 1993 Thermal sensors 15 employees 5 radiometry compensation pyrheliometer Hukseflux DR01 pyrheliometer
Background • Most pyranometers and pyrheliometers have indoor calibration • Exception: highest accuracy (BSRN, outdoor) • Exception: Japan, China (outdoor) • Cost, time, weather; outdoor calibration is unacceptable to industry
Present status (excerpt) • Eppley, US Weather Bureau: indoor integrating diffuse source • Kipp, Hukseflux: indoor normal incidence • EKO, JMA: outdoor tracker with collimation tube (changes) • China: outdoor
Background • Measurement uncertianty is a function of: • Characterisation / class • Calibration • Measurement & maintenance conditions • Environmental conditions
Background • Indoor calibration covered by ISO 9847 • Present ASME: “Indoor Transfer of Calibration from Reference to Field Pyranometers”
Hierarchy of Traceability • A: Reference calibration (uncertainty) • B: Correction of reference to indoor conditions (uncertainty) • C: Indoor calibration of field instrument (uncertainty) • Indoor calibration uncertainty estimate (A+B+C) • Field measurement uncertainty estimate
Hierarchy of Traceability • A: Reference calibration (uncertainty) • B: Correction of reference to indoor conditions (uncertainty) • C: Indoor calibration of field instrument (uncertainty) • Indoor calibration uncertainty estimate (A+B+C)
Hierarchy of Traceability • KNMI TR 235 "uncertainty in pyranometer and pyrheliometer measurements at KNMI in De Bilt".
Hierarchy of Traceability • A: Reference calibration (uncertainty) (conditions and class) • B: Correction of reference to indoor conditions (uncertainty) • C: Indoor calibration of field instrument (uncertainty) • Indoor calibration uncertainty estimate (A+B+C) • Field measurement uncertainty estimate (conditions & class)
Strange… • Errors in reference calibration re-appear in measurement errors • Counted double • At least systematic errors (Zero offset A and directional errors) can be avoided.
One step back • Present approach works well if calibrated instruments are used: • Outdoor / unventilated • Without applying GUM analysis • At same latitude
One step back • Present approach does not work well calibrated if instruments are used: • As indoor reference • Applying GUM analysis • At other latitudes • Ventilated
Typical calibration • Irradiance 800 W/m2 • 40 to 60 degrees angle of incidence, + / - 30 degrees azimuth • Zero offset A: -9 +/- 3 W/m2 (larger than ISO9060) • Directional: +/- 10 W/m2 @ 1000 W/m2 , now estimated +/- 5 W/m2
Typical calibration • PMOD specified uncertainty +/- 1.3% • Systematic error -1%? Type B.
Improved approach • Zero offset A: -9 +/- 3 W/m2 (larger than ISO9060) • Directional: +/- 10 W/m2 • Solution 1: ventilation • Solution 2: single angle of incidence
For consideration • Japanese collimated tube with tilt correction and ventilation
Diffuse sphere source • Uniformity of sphere top-edge (experimental -13%) • Weighing for non uniform source requires weighing of reference with source • Diffuse sphere: weighing requires weiging of field instrument with source. Complicated! • Normal incidence: weighing of field instrument is not necessary
Conclusion • Indoor calibration offers only acceptable solution for manufacturers and solar industry • “Normal incidence” calibration fits within ISO 98-3 GUM
Conclusion • Indoor normal incidence calibration is preferred (diffuse dome not) • Accuracy and precision of reference can be optimised (single angle, ventilated) • Pyrheliometer indoor calibration must be added
P.S. • Memo available via server