ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED) FOR NEONATES

ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED) FOR NEONATES. FP-50002 EWMA/GNEAUPP 2014. M. M. Sánchez-Lorente 1 ; P. García-Molina 2-3 ; C. Vergara-Hernández 2 ; E. Sanchis-Sánchez 2 ; J. Martos-Torres 2 ; A. Álvarez-Ordiales 4 ; & E. Balaguer-López 2-3.

makaio
Télécharger la présentation

ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED) FOR NEONATES

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED)FOR NEONATES FP-50002 EWMA/GNEAUPP 2014 M. M. Sánchez-Lorente1; P. García-Molina2-3; C. Vergara-Hernández2; E. Sanchis-Sánchez2; J. Martos-Torres2; A. Álvarez-Ordiales4; & E. Balaguer-López2-3 Conselleria de Sanitat, Valencia (SPA). University of Valencia, Valencia (SPA). Hospital Clínico Universitario, Valencia (SPA). Hospital Universitari i Politècnic “La Fe”, Valencia (SPA). Presenting author: Mr. Pablo García-Molina

  2. FP-50002 EWMA/GNEAUPP 2014 Background Existence of several neonatal supportsurfaces however… Lack of evaluation regarding… • Interface pressure & pressureredistribution • Efficacy in thepreventionortreatment of PU

  3. FP-50002 EWMA/GNEAUPP 2014 Aim 5 2 1 • To assess the interface pressure and pressure redistribution of five devices of constant low pressure (static and mixed support surfaces). 3 4

  4. FP-50002 EWMA/GNEAUPP 2014 Methods Study design • Laboratory study with repeated measures design. Sample • Eight neonatal simulation dummies with different characteristics regarding head circumference, weight and height.

  5. FP-50002 EWMA/GNEAUPP 2014 Methods Supportdevices

  6. FP-50002 EWMA/GNEAUPP 2014 Methods Variables • Mean pressure (mmHg). • Maximum pressure (mmHg). • Mean pressure in head region (mmHg). • Area of contact (cm2). Equipment • XSENSOR X3 PX100:40.40.002 device. Procedure • See: E-poster no. 50006.

  7. FP-50002 EWMA/GNEAUPP 2014 Methods Statistical analysis • Descriptive statistics. • Due to a small sample size, a non-parametric test (Friedman test) was carried out to assess differences among groups. If a significant difference was found, post hoc comparisons among groups were performed using the Wilcoxon-Nemenyi-McDonald-Thompson test. • The analysis was carried out by means of the R environment (The R Foundation for Statistical Computing). • Data set and code employed for the analysis can be found at www.figshare.com (*). *http://dx.doi.org/10.6084/m9.figshare.977866

  8. FP-50002 EWMA/GNEAUPP 2014 Results Lowest values! Highest values!

  9. FP-50002 EWMA/GNEAUPP 2014 Results

  10. FP-50002 EWMA/GNEAUPP 2014 Results Post hoc analysis: mean pressure (p-value)

  11. Results

  12. FP-50002 EWMA/GNEAUPP 2014 Results Post hoc analysis: maximum pressure (p-value)

  13. Results

  14. FP-50002 EWMA/GNEAUPP 2014 Results Post hoc analysis: pressure area (p-value)

  15. Results

  16. Conclusion • The “T” device presents the best values of pressure relief. • The polyester fiber device presents the best values of pressure redistribution. • Adding an upper layer of polyester fiber to the “T” device could result in a better support surface. • Next step: studying the properties of the new support surface in healthy subjects under clinical conditions.