1 / 21

T82 Single Gas Monitor

T82 Single Gas Monitor. 1. On. T82 Start-Up. To turn T82 ON / OFF. Press Function Switch. 8. STEL TWA LO HI ALARM PEAK BEEP BAT CALGAS. 188.8. T82 Start-Up. Display Test. Check the display for proper operation. 9. CO. T82 Start-Up. Sensor ID.

mcgoldrick
Télécharger la présentation

T82 Single Gas Monitor

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. T82 Single Gas Monitor 1

  2. On T82 Start-Up To turn T82 ON / OFF Press Function Switch 8

  3. STEL TWA LO HI ALARM PEAK BEEP BAT CALGAS 188.8 T82 Start-Up Display Test Check the display for proper operation 9

  4. CO T82 Start-Up Sensor ID Displays the currently installed sensor 10

  5. Bat T82 Start-Up Battery Status & Alarm Check (optional Vibrating alarm check) 11

  6. 5 T82 Start-Up 5 Sec. Countdown To access instrument settings press the Function Switch 12

  7. 000 T82 Start-Up Normal Operating Mode (15 Second Start-Up Time) 13

  8. 000 T82 Viewing Modes Backlight Screen (3 Seconds) Press & release function switch to view modes 14

  9. co T82 Viewing Modes Sensor ID Press & release function switch to view modes 15

  10. -0- T82 Viewing Modes Double Click Function Switch to Zero “BEEP” Press & release function switch to view modes 16

  11. PEAK 23 T82 Viewing Modes Double Click Function Switch to Clear Peak Hold 17

  12. STEL 000 T82 Viewing Modes Note: STEL & TWA averages are cleared each time T82 is turned off. STEL 18

  13. TWA 000 T82 Viewing Modes Note: STEL & TWA averages are cleared each time T82 is turned off. TWA 19

  14. Bat T82 Viewing Modes Battery Status 3 rows = up to 4400 hrs (lithium) 9-volt 2600 hrs (Alkaline) 9-volt 2 rows = less than 75 hrs 20

  15. T82 Viewing Modes Datalog Usage L12 Show user percentage of Datalog memory used. (L01 – L99 or FUL for 100%) 20

  16. T82 Viewing Modes Datalog Start New Session H03 Displays current session from H01 – H99 Double clicking function switch will start a new session. 20

  17. LO ALARM 010 HI ALARM 015 T82 Settings Low Alarm To access T82 settings, press FS during 5 second countdown. To view T82 settings, press & release FS High Alarm To changeT82 settings, use internal + or - keys 21

  18. STEL ALARM 015 TWA ALARM 010 T82 Settings STEL Alarm To access T82 settings, press FS during 5 second countdown. To view T82 settings, press & release FS TWA Alarm To changeT82 settings, use internal + or - keys 22

  19. CALGAS BEEP OFF T82 Settings Calibration Gas To access T82 settings, press FS during 5 second countdown. 025 To view T82 settings, press & release FS Confidence Beep To changeT82 settings, use internal + or - keys 23

  20. LCL T82 Settings Clear Datalog To access T82 settings, press FS during 5 second countdown. To view T82 settings, press & release FS To Clear T82 Datalog Double click the function switch 23

  21. 000 CAL P T82 Calibration From the Normal Operating Mode... Attach calibration cup and press & hold FS until “CAL” appears on display Apply gas, wait for 2-5 minutes for toxic gases, 30 sec. for O2. A “P” for pass or “F” for fail will appear along with a “Beep”. Turn off gas and remove cup. 24

More Related