An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Dose Issues in Pediatric Radiology George DavidAssociate Professor of Radiology Medical College of Georgia

  2. Computed Radiography (CR) • Re-usable metal imaging plates replace film & cassette • Uses conventional bucky & x-ray equipment

  3. CR Exposure & Readout

  4. CR Readout

  5. CR Latitude • Much greater latitude than screen/film • Plate responds to many decades of input exposure • under / overexposures unlikely • Computer scales image to provide viewable densities • Unlike film, receptor separate from viewer

  6. Film Screen vs. CR Latitude CR Latitude: .01 – 100 mR 100

  7. Digital Radiography (DR) • Digital bucky • Incorporated into x-ray equipment

  8. Digital Radiography (DR) • Potentially lower patient dose than CR • High latitude as for CR • Digital bucky fragile • First DR portables comingto market

  9. Raw Data Image • Unprocessed image as read from receptor • CR • Intensity data from PMT’s as a result of scanning plate with laser • DR • Raw Data read directly from TFT array • Not a readable diagnostic image • Requires computer post-processing • Specific software algorithms applied to image prior to presenting it as finished radiograph

  10. Enhancing Raw Image (Image Segmentation) * • Identify collimated image border • Separate raw radiation from anatomy • Apply appropriate tone-scale to anatomy • Done with look-up table (LUT) This process is specific to a particular body part and projection

  11. Look Up Table (LUT) • Converts a raw data pixel value to a processed pixel value • “Original” raw data pixel value indicates amount of radiation falling on pixel

  12. Image Segmentation • Computer must establish location of collimated border of image • Computer then defines anatomic region • Finished image produced by tone scaling • Requires histogram analysis of anatomic region

  13. Histogram • Graph showing how much of image is exposed at various levels

  14. Film/Screen Limited Latitude • Film use has little ambiguity about proper radiation exposure

  15. Should I Worry? • In CR & DR, image density is no longer a reliable indicator of exposure factor control.

  16. CR / DR Latitude DANGER Will Robinson!!! • Almost impossible to under or overexpose CR / DR • Underexposures look noisy • Overexposures look GOOD!!!

  17. CR / DR Latitude More DANGER • If adult technique used on peds patients, images look GOOD! • If grid removed for ped patient & grid technique used, image looks GOOD • NO ONE COMPLAINS

  18. Exposure Creep:Tendency of radiographs toward higher-then-necessary exposures • High doses have no detrimental effect on image quality • Desire to see less noise on radiographs • Increased exposure latitude

  19. So how do I judge the exposure if I can’t tell by looking at the image?

  20. Exposure Index

  21. Exposure Index • Each manufacturer provides exposure feedback to technologist • Displayed on CR reader monitor • Displayed on workstations

  22. Exposure Index • Measure of radiation received by receptor below anatomy • Not a direct measure of patient exposure • If exposure index higher than necessary, patient overexposed

  23. Displayed Exposure Index Affected by • X-Ray technique selection • Improper centering of image on cassette • Improper selection of study or projection • Placing two or more views on same cassette • Can cause image to appear dark

  24. Exposure Indication Varies between Manufacturers Fuji • “S” number goes down as exposure goes up! • S is half when exposure doubled • Kodak • Logarithmic scale • EI goes up 300 when exposure doubled

  25. Doubling Exposure Index is Logrithmic • EI = 2000 +1000 * log(exposure)

  26. Initial Set-up for Exposure Index • Kodak recommendation for exposure index: 1800 – 2200 • Manual technique: • Technologist should strive to keep exposure index consistent • Phototiming: • Set-up by service & physics according to manufacturer’s instructions

  27. Imaging is NOT a Beauty Contest • CR/DR exposure should be selected to provide “Maximum tolerable noise” • Keeps dose as low as possible (ALARA) • Noise tolerance depends on study & objective. • Technologist requires noise feedback from radiologists

  28. Phototimed Phantom Image • 75 kVp • 88 mAs • 2460 EI

  29. Let’s Approximately Double mAs • 75 kVp • 88 mAs • 2460 EI • 75 kVp • 160 mAs • 2680 EI

  30. Let’s Go Crazy • 75 kVp • 88 mAs • 2460 EI • 75 kVp • 640 mAs • 3300 EI

  31. How Low Can You Go? Cut mAs in Half! • 75 kVp • 88 mAs • 2460 EI • 75 kVp • 40 mAs • 2060 EI

  32. Let’s Go Crazy Low • 75 kVp • 8 mAs • 1380 EI • 75 kVp • 1 mAs • 550 EI

  33. Fluoroscopy Doses • Beam time • Geometry • Application of features

  34. Last Image Hold • Computer displays last fluoro image before radiation shut off. • Image noisier than for digital spot • Image made at fluoroscopic technique / intensity • Less radiation than digital spot • Allows operator to review static processes while beam off • ideal for teaching • ideal for orthopedic applications such as hip pinning

  35. Fluoro Frame Averaging • Computer averages current with user-selectable number of previous frames • Averages current frame & history • Conventional fluoro only displays current frame

  36. Fluoro Frame Averaging Tradeoff • Advantage: • Reduces quantum noise • Disadvantage • Because history frames are averaged with current frame, any motion can result in lag

  37. Fluoro Frame Rates • Many systems allow option of using lower frame rates • 15, 7.5, 3.75 fps rather than 30 • computer displays last frame until next one • reduces flicker • Most implementations lower patient & scatter exposure • Exposure proportional to frame rate • dynamic studies may be jumpy

  38. Patient CT Patient Dose • Tube rotates around patient during study • Dose distribution different than radiography

  39. USA Today 6/19/2001 “Each year, about 1.6 million children in the USA get CT scans to the head and abdomen — and about 1,500 of those will die later in life of radiation-induced cancer, according to research out today.”

  40. Biggest Question about CT Doses? Appropriateness of Study

  41. Pediatric Color Coding • Broselow-Luten Pediatric System • Based on child’s size & weight • Used in ER’s for • Resuscitation & support apparatus • Medications • IV fluids

  42. Pediatric Color Coding In CT • Slightly modified Broselow-Luten Pediatric System • Color code is used to determine complex CT protocols • Contrast options • Scanner protocols • Multi-slice has even greater variety of options • Detector configuration • Gantry rotation speed • Table speed • Color coding significantly reduces protocol errors

  43. CT Dose depends on X-Ray Beam Image Quality & Processing Selections • noise • detector efficiency • matrix resolution • reconstruction algorithm • kVp • mA • time • pitch • filtration

  44. CT Contrast Resolution Depends on NoiseNoise is function of mA

  45. CT Dose Measurement • Lucite phantom • 5 holes • One center • Four 90o apart in periphery • Chamber placed in one hole • Lucite plugs placed in remaining 4 holes • Slice centered on phantom Chamber Plugs

  46. Measuring CT Dose IndexCTDI • “Pencil” ion chamber used • Pencil oriented in “Z” direction Dose Phantom Chamber Hole for Chamber Active Chamber Area (exposed area of chamber) Z Slice Z

  47. CTDI100 • Measurement made with 100 mm chamber • Includes scatter tails Dose Phantom Chamber 100 mm Slice Z

  48. CTDIW • Weighted average of center & peripheral measurements • Represents “average” dose in scan plane CTDIW = 1/3 CTDI100-center + 2/3 CTDI100-periphery CTDIC CTDiP