An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. DNA Digital Level Series

  2. DNA Product Training forLeica Sales & Support Representatives DNA Product Training, V1.0

  3. Content 1. Purpose and target of training 2. Overview 3. Comparison DNA to NA series 4. Operating and functions 5. Application software 6. Data concept 7. Competition 8. User benefits 9. Summary DNA Product Training, V1.0

  4. 1) Purpose and target of training Purpose: • This product training will train you: • To use the DNA Digital Levels. • To demonstrate the DNA Digital Levels. Objectives: • At the end of the training you will: • Be familiar with the functions and operation of the instrument. • Be able to demonstrate confidently the applications provided in the instruments. • Be able to conduct training for end-users and customers. DNA Product Training, V1.0

  5. 2) Overview - General • The 2nd generation of Leica Digital Levels. • Based on over 10 years of experience. • Based on TPS700 software and user concept. • The first Leica instrument with data storage into internal memory and data saving on PC card. DNA03 / DNA10 DNA Product Training, V1.0

  6. 2) Overview – Design (1) • Modern design, fixed handle with integrated sight. • Generous LC-display presents all measured and calculated data at a glance. • Alphanumeric keyboard adapted from TPS700. • Optimal view to the circular level. Separate illumination for circular level. • Slot for PC Card. New ! DNA Product Training, V1.0

  7. 2) Overview – Design (2) • Weight • including GEB111 • battery:2.8kg • Dimensions • Instrument • 168 x 240 x 210mm DNA Product Training, V1.0

  8. 2) Overview – Design (3) • Measure button in central position for minimal impact. • Endless horizontal drives on both sides. • Unchanged bar code for continuous use of existing staffs. • Serial interface RS232. • Inside: Magnetically damped compensator. DNA Product Training, V1.0

  9. 2) Overview - Battery concept • … adapted to TPS • Standard camcorder battery • - GEB111 (1.8Ah, 6V NiMH) • - GEB121 (3.6Ah, 6V NiMH) • - GEB70 (2.2Ah, 12V NiCd) • Alcaline cells • - 6x LR6/AA/AM3, 1.5V in battery adapter GAD 39 DNA Product Training, V1.0

  10. 2) Overview – Common with TPS700 (1) The following components are same or similar to TPS700: - Alphanumeric keyboard - Display - Batteries - Program structure and user guidance - Data storage and data output concept  reduced learning curve for users who are familiar with TPS700. DNA Product Training, V1.0

  11. 2) Overview – Common with TPS700 (2) DNA Product Training, V1.0

  12. 2) Overview - Equipment • Standard equipment: • Instrument without batteries • Rain cover • Tool kit • User manual, Field manual • CD-ROM with Leica Survey Office and more • Container • Optional accessory set: • 2  GEB 111 batteries • 1  GKL 111 charger • 1  GEV 102 cable • 1  MFC16MB-3 PCMCIA-ATA flash card • 1  GSB5 sunshade • Enhanced Accessory Set: • 1*GDF 121 tribrach • 2*GEB 121 batteries • 1*GKL 122 charger • 1*GEV 102 cable DNA Product Training, V1.0

  13. 2) Overview - Data storage concept • Internal memory: • Job number adaptable up to 16 jobs. • 7000 fixed points or 6000measurements or 1650 stations BF. Data saving: • PCMCIA-card for data exchange (max. size  32MB). • No direct recording to the PC card! DNA Product Training, V1.0

  14. 2) Overview - Accuracy The new DNA digital levels are available in two accuracy versions. Choose DNA10 for typical construction levellings, for high-precision surveys choose DNA03. DNA Product Training, V1.0

  15. 3) Comparison DNA to NA series Differences betweenDNA03/DNA10 and NA3003/NA2002 DNA Product Training, V1.0

  16. 3) Comparison – Main differences (1) DNA03/DNA10 NA3003/NA2002 • General • - Modern design of housing. • - Fixed handle. • - Horizontal drives on both sides. • - Horizontal circle with combined 360° and 400g (50g steps) graduation. • - Display: 8 x 24, with heating. • - Alphanumeric keyboard. • - Measure button in central position. • - Circular level: position optimised and with illumination. • - Serial interface RS232 on right side. • - Slot for PC card. • General • - Design 1990. • - Movable handle. • - Horizontal drive on right side only. • - Horizontal circle with 360° or 400g graduation. • - Display: 2 x 8, no heating. • - Numeric keyboard. • - Measure button outside of centre. • - Circular level: no illumination. • - Serial interface RS232 on left side. • - Slot for REC Module. DNA Product Training, V1.0

  17. 3) Comparison – Main differences (2) DNA03/DNA10 NA3003/NA2002 • Battery concept • - GEB111, GEB121 (TPS). • - Alcaline cells supported. • Data storage concept • - Internal memory as permanent storage device for direct recording. • - Every measurement is automatically stored in the internal memory except in “Meas Only – without recording”. • - Optional data saving on PCMCIA card (copying data from internal memory to card). • - Data formats: GSI-8, GSI-16, XML, User (GSI-8 different from GSI-8 of NA3003). • - Recording to RS232 only from “Meas & Rec”. • Battery concept • - GEB79. • - Alcaline cells not supported. • Data storage concept • - REC module as removable storage device for direct recording. • - On/Off setting for recording. • - No other storage device. • - Data format: GSI-8. • - Recording to RS232 from all appli- cations (Line levelling, Meas Only). DNA Product Training, V1.0

  18. 3) Comparison – Main differences (3) DNA03/DNA10 NA3003/NA2002 • Measure modes • - Single (typical 3 sec.), Mean (Average), Median, Mean with predef. standard deviation, --- Repeated single. • On-board applications • - Measure & Record. • - Measure Only. • - Line Levelling BF, aBF, BFFB, aBFFB. • - Check & Adjust. • - Remark (WI71), Free code with 8 informations, Quick Code. • - Stake out: height, height diff., distance. • Measure modes • - Single (typical 4 sec.), Mean (Average), Median, Mean with predef. standard deviation, Integration time (NA3003). --- • On-board applications • - --- • - Measure Only. • - Line Levelling BF, BFFB. • - Check & Adjust. • - Free Code with 4 Informations. • - Stake out: height. DNA Product Training, V1.0

  19. 3) Comparison – Main differences (4) DNA03/DNA10 NA3003/NA2002 • Performance • - Accuracy DNA03: (Stand. dev. for 1km double run) 0.3mm with invar staff. • - Automatic staff end recognition for 4.05m(4m), 3m, 2.7m, 2m, 1.82m staffs. • - Maximum Distance: Staff length  3m 110m for invar staff 60m recommended Staff length = 2.7m 100m Staff length = 1.82m/2m 60m • - Field of view for electr. measurement close far Coarse measurement: 2° … 1.4° Fine measurement: 1.25 … 1.1° • - Coverage on the staff: avoid coverage in centre of field of v. • Performance • - Accuracy NA3003: (Stand. dev. for 1km double run) 0.4mm with invar staff. • - Manual staff end settings: 4.05m/2.7m and 3m/2m. • - Maximum Distance: depending on staff end settings 4.05m/2.7m: 100m 3m/2m: 60m • - Field of view of electr. measurement 2°. • - Coverage permitted anywhere on staff up to 20%. DNA Product Training, V1.0

  20. 4) User interface layout Alphanumeric entry User Defined FNC priority Other Functions Confirmation Execution On / Off Circular level User programs Cancel last entry Second key assignment Shift for Alpha Eyepiece Settings and Data menu Esc to exit programs DNA Product Training, V1.0

  21. 4) User interface • Statussymbols for battery, data sets and further pages. • Choice fieldmarked with symbol. Left and right navigation keys scroll through options. • Navigationkeys available anytime, paging for more displayed data. • Direct-Keysfor measurements, programs and functions including light. • Buttonsare used on screen for branching and execution. These are selected with the navigation keys andactivated withRETURN. Direct keys Choice field Status Navigation keys Button DNA Product Training, V1.0

  22. 4) Keys different from TPS700 • The direct keys are adapted to digital level functionality. • DATA Separate key for immediate access to the data manager. • ESC Separate key for quick step-by-step escape from programs and menus. DNA Product Training, V1.0

  23. 4) Key: SET OUT and INT SET OUT (Shift + INT) Switch to Setting-out sight for the setting-out of … - Heights - Height differences - Distances INT Switch to Intermediate sight. In general a special display is required for switching to INT and SET OUT. For example in “Measure & Record” the Foresight display is required. DNA Product Training, V1.0

  24. 4) Key: INV and MODE INV (Shift + MODE) Switch between inverted and normal staff position. Staff readings will have negative sign for inverted staff. MODE Select the Measure mode: - Single - Mean - Mean s - Median - Repeated single DNA Product Training, V1.0

  25. 4) Key: FNC and USER FNC (Shift + USER) Select a function: - Test measurement - View measurement - Code - PtID & increment - Manual input USER Opens the FNC function assigned to the USER key. Assign the function in:MENU – Quick Settings. DNA Product Training, V1.0

  26. 4) Key: MENU and PROG MENU (Shift + PROG) Set instrument settings: - Quick Settings - All Settings - System Info - Check with collimator (DNA03) PROG Select an application program: - Measure Only - Line Levelling - Check & Adjust DNA Product Training, V1.0

  27. 4) Key: Light and DATA Light (Shift + DATA) Set display and circular level illumination: - Economic - Permanent - Circular level only DATA Access to Data manager: - View / edit data - Initialize memory - Memory information - Data export - Data import <CARD> PC card functions DNA Product Training, V1.0

  28. 4) Key: ESC ESC Direct key for quick step-by-step escape of programs and menus. DNA Product Training, V1.0

  29. 4) Measure display Easy control of multiple readings during the measuring process. The count The staff reading value(according to Mode) The standard deviation of ... - a single measurement - the mean The spread Observe the update of values ... DNA Product Training, V1.0

  30. 4) Quick coding Proven in the TPS700 • Quick Code allows the user to automatically attach a code to the measurement. The measurement is launched by pressing two numeric keys (00 - 99)! • Pressing two numeric keys automatically initiates: • A full measurement. • Recording of the measurement. • Recording of the pre-assigned code to the measurement. • The code is stored before or after the measurement accor-ding to the setting in MENU – Quick Settings. • Quick Code is available in the ”Measure Only” application. DNA Product Training, V1.0

  31. 5) Levelling basics - Determining a height difference Intermediate Sight (=Int) Backsight (=B) Foresight (=F) Stake Out (=SO) DH = B – F DHInt= B – Int DHSO= B– SO DNA Product Training, V1.0

  32. 5) Software components Application Programs PROG INT DATA Intermediate Set Out Data Manager SET OUT Startup program ”Measure & Record” FNC MODE Functions Measure Modes MENU Config. / Settings USER DNA Product Training, V1.0

  33. 5) Overview system software Measure & Record (ready after system startup) MENU PROGRAMS FUNCTIONS DATA MODE Test Measurement View Measurement Code PtID & Increment Manual Input Measure Only Line levelling Check & Adjust Quick Settings All Settings System Info *Check with Collimator *only DNA03 View/Edit Data Initialize Memory Memory Information Data Import Data Export <Card> Single Rep. Single Mean Mean s Median Coding “Quick Coding” DNA Product Training, V1.0

  34. 5) Branching from and to programs PROG Measure & Record INT DATA Programs  Measure Only  Line levelling  Check & Adjust SET OUT Recording to RS232 FNC INV MODE 1 Set Job 2 Set Line 3 Start 1 Set Job 2 Set Line 3 Set Tolerances 4 Start MENU 1 Set Job 2 Set Method 3 Start USER DNA Product Training, V1.0

  35. 5) Measure & Record Instantly ready and easy! • Switch on the instrument… • The instrument shows the “Meas & Rec” display. • The instrument is ready for (Backsight) measurement(s). • Optional: • Enter the starting point height . • Enter a remark code. • Measure… • The instrument displays staff reading, distance and height of collimation. • Do more measurements or press <CONT> to continue to the Foresight display. DNA Product Training, V1.0

  36. 5) Application programs <Quick Code> DNA Product Training, V1.0

  37. 5) Program ”Measure Only” (1) Program “Measure Only” is available in a recording and non-recording version. Simply set Job, set Line and START !! DNA Product Training, V1.0

  38. 5) Program ”Measure Only” (2) Quick Coding is implemented only in this Program. It’s really convenient. Just enter the 2 digits to start the measurement and to store both code and measurement at the same time • Code Lists for Quick Coding • can be … • entered in the field via FNC • defined and uploaded with Leica Survey Office DNA Product Training, V1.0

  39. 5) Line levelling methods BF, aBF B2 B2 B1 F2 F2 F3 F3 B1 • BF • aBF odd station even station DNA Product Training, V1.0

  40. 5) Line levelling methods BFFB, aBFFB B22 B22 F23 F23 B21 B21 F22 F22 B12 B12 F13 F13 F12 F12 B11 B11 • BFFB • aBFFB odd station even station DNA Product Training, V1.0

  41. 5) Program line levelling Ready Steady GO !!! DNA Product Training, V1.0

  42. 5) Check & adjust methods Method A x x B Method A x Bx ... + manual entry of collimation error. DNA Product Training, V1.0

  43. 5) Stake out methods Stake out by height Enter design height of a stake out point. Start the measurement and receive shifting values as CUT / FILL values Stake out by height difference Enter design height difference between last backsight and stake out point. Start the measurement and receive shifting values as CUT / FILL values Stake out by distance Enter design distance between the Instrument and a stake out point. Start the measurement and receive shifting values as IN / OUT values DNA Product Training, V1.0

  44. 6) Data organisation DATA Job Measure & Record Measurement Data Line n Line 2 Line 1 Line 2 Line n Line 1 Line n Line 1 Line 2 Job 1 Job 2 Job n Fixpoint Data Fixpoints Job 2 Fixpoints Job n Fixpoints Job 2 DNA Product Training, V1.0

  45. 6) Job Management  Up to 16 jobs available for measurements and fixpoints. A job contains measurements or fixpoints only or measurements and fixpoints. Capacity of internal memory: Approx. 6000 measurements or 1650 stations BF.  Export of job data through serial interface or via PCMCIA card. DNA Product Training, V1.0

  46. 6) Data storage 32...8+03061009 332.08+00151306 52....+0003+001 110005+00000001 573..8-00184215 574..8+05937802 83..08+00009770 110006+00000001 32...8+01349377 331.08+00155976 52....+0003+000 110007+00000002 32...8+01157432 332.08+00157026 52....+0003+001 110008+00000002 573..8+00007729 574..8+08444611 83..08+00008719 110009+00000002 32...8+01889342 331.08+00164827 52....+0003+001 110010+00000003 32...8+01875907 332.08+00159179 52....+0003+002 110011+00000003 573..8+00021164 574..8+12209860 83..08+00014368 110012+00000003 32...8+01953231 331.08+00162498 52....+0003+002 110013+00000004 32...8+02020837 332.08+00159628 52....+0003+002 110014+00000004 573..8-00046442 574..8+16183929 83..08+00017237 110015+00000004 32...8+02080802 331.08+00163303 52....+0003+001 110016+00000005 32...8+02146420 332.08+00164573 52....+0003+001 110017+00000005 573..8-00112059 574..8+20411151 83..08+00015966 110018+00000005 32...8+02136300 331.08+00164488 52....+0003+001 110019+00000006 32...8+02131652 332.08+00148424 52....+0003+000 110020+00000006 573..8-00107411 574..8+24679103 83..08+00032030 110021+00000006 32...8+02266377 331.08+00152865 52....+0003+002 110022+00000007 32...8+02183851 332.08+00154855 52....+0003+000 110023+00000007 573..8-00024885 574..8+29129332 83..08+00030040 110024+00000007 32...8+02066983 331.08+00159862 52....+0003+001 110025+00000008 32...8+02131336 332.08+00153770 52....+0003+001 110026+00000008 573..8-00089238 574..8+33327650 83..08+00036133 110027+00000008 32...8+02223770 331.08+00159367 52....+0003+002 110028+00000009 32...8+02160936 332.08+00150116 52....+0003+001 110029+00000009 573..8-00026404 574..8+37712357 83..08+00045384 110030+00000009 32...8+02219720 331.08+00149857 52....+0003+003 110031+00000010 32...8+02229407 332.08+00150815 52....+0003+000 110032+00000010 573..8-00036090 574..8+42161483 83..08+00044426 110033+00000010 32...8+02361630 331.08+00164413 52....+0003+000 110034+00000011 32...8+02098768 332.08+00154958 52....+0003+000 110035+00000011 573..8+00226771 574..8+46621881 83..08+00053881 110036+00000011 32...8+02125435 331.08+00162216 52....+0003+002 110037+00000012 32...8+02249602 332.08+00136302 52....+0003+000 110038+00000012 573..8+00102603 574..8+50996918 83..08+00079795 410039+?......1 110040+00000011 83..08+00053881 110041+00000011 32...8+02276099 331.08+00136003 52....+0003+002 110042+00000011 32...8+02095376 332.08+00161919 52....+0003+003 110043+00000011 573..8+00180723 574..8+04371475 83..08+00027965 110044+00000011 32...8+01992489 331.08+00153933 52....+0003+001 110045+00000010 32...8+02470745 332.08+00163388 52....+0003+003 110046+00000010 573..8-00297533 574..8+08834708 83..08+00018511 110047+00000010 32...8+02381207 331.08+00151976 52....+0003+000 110048+00000009 32...8+02062949 332.08+00151010 52....+0003+000 110049+00000009 573..8+00020726 574..8+13278864 83..08+00019477 110050+00000009 32...8+02189513 331.08+00151333 52....+0003+001 110051+00000008 32...8+02202097 332.08+00160586 52....+0003+001 110052+00000008 573..8+00008142 574..8+17670474 83..08+00010223 110053+00000008 32...8+02165932 331.08+00149973 52....+0003+001 110054+00000007 32...8+02040257 332.08+00156060 52....+0003+001 110055+00000007 573..8+00133817 574..8+21876663 83..08+00004137 110056+00000007 32...8+02187454 331.08+00158342 52....+0003+003 110057+00000006 32...8+02250208 332.08+00156363 52....+0003+001 110058+00000006 573..8+00071064 574..8+26314325 83..08+00006116 110059+00000006 32...8+02136287 331.08+00149445 52....+0003+004 110060+00000005 32...8+02124423 332.08+00165518 52....+0003+000 110061+00000005 573..8+00082928 574..8+30575035 83..08-00009957 110062+00000005 32...8+02110351 331.08+00163310 52....+0003+001 110063+00000004 32...8+02136694 332.08+00162031 52....+0003+001 110064+00000004 573..8+00056585 574..8+34822079 83..08-00008678 110065+00000004 32...8+01942128 331.08+00159527 52....+0003+001 110066+00000003 32...8+02024913 332.08+00162395 52....+0003+001 110067+00000003 573..8-00026200 574..8+38789121 83..08-00011546 110068+00000003 32...8+01913745 331.08+00160075 52....+0003+001 110069+00000002 32...8+01862889 332.08+00165723 52....+0003+001 110070+00000002 573..8+00024655 574..8+42565755 83..08-00017195 110071+00000002 32...8+01253228 331.08+00159216 52....+0003+002 110072+00000001 32...8+01236217 332.08+00158164 52....+0003+002 110073+00000001 573..8+00041665 574..8+45055200 83..08-00016143 110074+00000001 32...8+03028479 331.08+00152728 52....+0003+002 110075+00000000 32...8+02906374 332.08+00162501 52....+0003+001 110076+00000000 573..8+00163771 574..8+50990053 83..08-00025916 410077+?......1 110078+00000012 83..08+00079795 110079+00000012 32...8+03199192 331.08+00139365 52....+0003+005 110080+00000012 32...8+03285258 332.08+00139244 52....+0003+001 110081+00000012 573..8-00086066 574..8+06484450 83..08+00079916 110082+00000012 32...8+03180560 331.08+00148633 52....+0003+004 110083+00000013 32...8+03092979 332.08+00154215 52....+0003+001 110084+00000013 573..8+00001514 574..8+12757988 Operator : HEF DNA03 : Instr.Nr.: 13 Remarks : RAIN / OVERCAST Date/Time : 05.12.2001 / 9:32:27 ------------------------------------------------------------------------------------ Line Levelling ------------------------------------------------------------------------------------ Line: 1 / BF ======================================================== Date: 05.12.2001 / Time: 10:03:22 ======================================================== ¦ PointID ¦Backsight ¦ Intermdt.¦Foresight ¦ ¦ Distance ¦Remarks ¦ ¦(Station Result)¦ ¦ ¦ ¦(delta H) ¦ (D.bal.) ¦(Pt.Hgt.) ¦ ------------------------------------------------------------------------------------ ¦Start PtID ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦1 ¦ ¦ ¦ ¦ ¦ ¦( 0.00000)¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦1 ¦ 1.61076¦ ¦ ¦ ¦ 28.77¦ --------¦ ¦1 ¦ ¦ ¦ 1.51306¦ ¦ 30.61¦ --------¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦( 1) ¦ ¦ ¦ ¦(+0.09770)¦( -1.842)¦( 0.09770)¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦1 ¦ 1.55976¦ ¦ ¦ ¦ 13.49¦ --------¦ ¦2 ¦ ¦ ¦ 1.57026¦ ¦ 11.57¦ --------¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦( 2) ¦ ¦ ¦ ¦(-0.01050)¦( +0.077)¦( 0.08719)¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦2 ¦ 1.64827¦ ¦ ¦ ¦ 18.89¦ --------¦ ¦3 ¦ ¦ ¦ 1.59179¦ ¦ 18.76¦ --------¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦( 3) ¦ ¦ ¦ ¦(+0.05648)¦( +0.212)¦( 0.14368)¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦3 ¦ 1.62498¦ ¦ ¦ ¦ 19.53¦ --------¦ ¦4 ¦ ¦ ¦ 1.59628¦ ¦ 20.21¦ --------¦ • Job, line, meas mode, correction, start point, measurements, targets, results, correlation coefficient,… • Standard conversions include XML and GSI, as well as 3 user-definable output formats • Fieldbook style output can be created on-board DNA Product Training, V1.0

  47. 6) Data formats GSI-8(slightly different from NA3003 GSI) 410001+?......1 110002+0000BM12 83...8+41269800 110004+0000BM12 32...8+03275491 331.108+00162953 71....+CONCRETE 110005+00000001 32...8+03470938 332.108+00146285 New ! GSI-16 *410001+00000000?......1 *110002+00C000000000BM12 83...8+0000000041269800 *110004+000000000000BM12 32...8+0000000003275491 331.108+0000000000162953 71....+00000000CONCRETE *110005+0000000000000001 32...8+0000000003470938 332.108+0000000000146285 New ! XML(used by LevelPak-Pro) <Measurement PointId = "100" Active = "Yes" Type = "B1" StaffReading = "1.86622491255000000000" Distance = "4.82324538899000000000" Source = "Measured" User formats(configurable) ¦1 ¦ Back¦ 1.61076¦ 28.77¦ ¦ - ¦1 ¦ Fore¦ 1.51306¦ 30.61¦ 0.098¦ - ¦1 ¦ Back¦ 1.55976¦ 13.49¦ ¦ - ¦2 ¦ Fore¦ 1.57026¦ 11.57¦ 0.087¦ - New ! DNA Product Training, V1.0

  48. 6) Configurable data output formats Line: 1 / BF ================================================================= Date: 05.12.2001 / Time: 10:03:22 Staff1: -------- / Staff2: -------- ================================================================= ¦ PointID ¦ Type ¦ Staff ¦ Distance ¦Pt.Height ¦Rem ------------------------------------------------------------------ ¦1 ¦ Back¦ 1.61076¦ 28.77¦ ¦ - ¦1 ¦ Fore¦ 1.51306¦ 30.61¦ 0.098¦ - ¦1 ¦ Back¦ 1.55976¦ 13.49¦ ¦ - ¦2 ¦ Fore¦ 1.57026¦ 11.57¦ 0.087¦ - ¦2 ¦ Back¦ 1.64827¦ 18.89¦ ¦ - ¦3 ¦ Fore¦ 1.59179¦ 18.76¦ 0.144¦ - ¦3 ¦ Back¦ 1.62498¦ 19.53¦ ¦ -- Customizable formats for any style of data outputs. (Sample format files are stored on the CD-ROM) List ==================================================================================== ¦ PointID ¦Backsight ¦ Intmdt ¦Foresight ¦ delta H ¦ Distance ¦Point Hgt.¦ ------------------------------------------------------------------------------------ ¦Start PtID ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦1 ¦ ¦ ¦ ¦ ¦ ¦ 0.000¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦1 ¦ 1.61076¦ ¦ ¦ ¦ 28.77¦ ¦ ¦1 ¦ ¦ ¦ 1.51306¦ +0.09770¦ 30.61¦ 0.098¦ ¦1 ¦ 1.55976¦ ¦ ¦ ¦ 13.49¦ ¦ ¦2 ¦ ¦ ¦ 1.57026¦ -0.01050¦ 11.57¦ 0.087¦ ¦2 ¦ 1.64827¦ ¦ ¦ ¦ 18.89¦ ¦ ¦3 ¦ ¦ ¦ 1.59179¦ +0.05648¦ 18.76¦ 0.144¦ Basic Staff1: -------- / Staff2: -------- ==================================================================================== ¦ PointID ¦Backsight ¦Fore/Intm.¦ Rise ¦ Fall ¦ Distance ¦Pt.Height ¦ ------------------------------------------------------------------------------------ ¦Start Point ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦1 ¦ ¦ ¦ ¦ ¦ ¦ 0.000¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦1 ¦ 1.61076¦ ¦ ¦ ¦ 28.77¦ ¦ ¦1 ¦ ¦ 1.51306¦ +0.09770¦ +¦ 30.61¦ 0.098¦ ¦1 ¦ 1.55976¦ ¦ ¦ ¦ 13.49¦ ¦ ¦2 ¦ ¦ 1.57026¦ +¦ -0.01050¦ 11.57¦ 0.087¦ ¦2 ¦ 1.64827¦ ¦ ¦ ¦ 18.89¦ ¦ ¦3 ¦ ¦ 1.59179¦ +0.05648¦ +¦ 18.76¦ 0.144¦ Rise and Fall DNA Product Training, V1.0

  49. 6) Leica Survey Office (LSO) DNA Product Training, V1.0

  50. 6) LSO Components Leica Survey Office Data Exchange Manager Codelist Manager Format Manager Coordinate Editor Configuration Manager Up- and downloading of: - Measurement Data - Code lists - Formats - Coordinate lists “Creator Tool” for codelists containing: Layers Codes Attributes Configurable output formats Formatting to back office requirements while downloading Facility to provide reports directly from the download file Simplifies data flow from field to finish Tool for creation and modification of coordinate lists with: Point ID, Easting, Northing, Elevation. Configures all TPS and DNA Series instruments GSI Converter converts GSI datafrom DNA toGSI-8 of NA instr. Coding “Quick Coding” Software Upload DNA Product Training, V1.0