Reprocessing complex and general surgical instruments: Endowrist

Reprocessing complex and general surgical instruments: Endowrist. Martin Johnson 26 th May 2011. The evolution of the market… ...the growth of our range. Our philosophy. “If it’s not clean, it can’t be sterilised”. Proving that it’s clean: the Radio Nucleide Method (RNM).

Télécharger la présentation

Reprocessing complex and general surgical instruments: Endowrist

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Reprocessing complex and general surgical instruments: Endowrist Martin Johnson 26th May 2011

  2. The evolution of the market… ...the growth of our range

  3. Our philosophy “If it’s not clean, it can’t be sterilised”

  4. Proving that it’s clean: the Radio Nucleide Method (RNM)

  5. Visualisation for Optimisation Test instrument: fan retractor Before cleaning 15 seconds with water pistol and 45 minutes in washer 10 minutes in the Medisafe system

  6. The da Vinci system

  7. The Endowrist instrument

  8. Test Conclusions • Ultrasonics – on the inside and outside of the instrument • High pressure – 2 bar minimum ( 30psi) • Detergent injection • Time specific – for the enzymatic detergent and for the exposure to ultrasonic power

  9. ISO 17664 validated cycle • Pre-wash – internal and external at 3 bar • Detergent injection • 20 minute main wash with sonic irrigation internally and externally at 3 bar • Pre-rinse – internal and external at 3 bar • Thermal disinfection • Empty – each cycle

  10. How the PCF works

  11. Ready to be cleaned

  12. The SI PCF and the Niagara SI PCF from Medisafe – The only validated automated method for cleaning complex and general instruments using high pressure irrigation and combined ultrasonics.

  13. Instrument manufacturer approvals • Major manufacturers • Conformity to official standards • More to come...