An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Detailed Operation: Model 1130 and Model 1135 Rev: 1.11 November 3, 2003

  2. Model 1135 – Particulate Mercury Monitor • Unit stacks on top of Model 1130 • Allows simultaneous deter-mination of Hg0, RGM and HgP • Detection limit < 2 pg/m3

  3. Instrument Characteristics • Fully automated, unattended operation • Denuder exchanged every 1 or 2 weeks • Operates in any environment • Polar (Greenland, USA, Canada, Antarctica) • Sub-tropical (Florida Everglades) • Desert (Nevada) • Marine (ocean cruises) • High altitude (Mauna Loa)

  4. Data Characteristics • High time resolution Hg0 data provided during sampling phase • Typically every 2½ or 5 min • Data is averaged to yield one Hg0 point per sampling period • Single RGM integrated value every sample period • Sample period: 0.5 – 3 hours • Analysis Period: 25 – 60 min • Data values are for the exact same time period

  5. Zero Air Inlet Sample Air Inlet Elbow (not used with 1135) Annulus Coupler Coupler Elutriator Jet Impactor Body & Disk Quartz Denuder Assembly

  6. Tail Quartz Wool Pyrolyzer (quartz chips) Air Gap Quartz membrane filter disk Quartz Wool Inlet (from denuder) RPF • Quartz wool captures bulk of particulate matter • Quartz filter membrane provides sharp cut-off (<0.1 um) • Built in pyrolyzer section converts desorbed species to Hg0 • Carbon based particulate matter on the filter is oxidized to CO2 during desorption cycle • Tail provides cooling

  7. Analytical Sequence • Wait state is entered whenever: • Model 1130 or Model 2537A is first started up • Model 2537A goes into Idle or Calib mode • Wait state is exited and monitoring cycle begins when: • Model 2537A is in Run mode, and • Cycle transition is detected

  8. Typical Flow Rate Settings

  9. Analytical Functions

  10. Event Flag Assignments

  11. Typical Temperature Settings • Temperatures are set directly on the PID controllers • System simply selects between the two set-points on analytical PID controllers • Case temperature controllers independently actuate either: • case heater, or • cooling fan

  12. Method Calculated Values • Values are calculated both by the controller and by XLS spread sheet • Most are for info only • Factors must be entered into 2537A to ensure direct reading in pg/m3

  13. Step 0: Sample

  14. Step 1: Flush

  15. Step 2: Pyro-Ht

  16. Step 3: Part-Ht

  17. Step 4: RGM-Ht

  18. Step 5: RGM-H2

  19. Step 6:Cool

  20. Step 7: Wait

  21. Model 1130/35 Data Output Hg0BlankHgPRGM ng/m3 pg/m3pg/m3pg/m3 Input File: D:\TEKRAN\TRAINING\1135.DAT Output File: D:\TEKRAN\TRAINING\1135.col Number of columns= 4 Column 1 Field: ng/m3 Cart: AB Status: 0 Column 2 Field: ng/m3 Cart: AB Status: 1 Column 3 Field: ng/m3 Cart: AB Status: 2 Column 4 Field: ng/m3 Cart: AB Status: 3 Date Time Typ C Stat AdTim Vol Bl BlDev MaxV Area ng/m3 Column1 Column2Column3Column4 01-10-01 00:55:01 CONT A OK 0 300 6.25 0.107 .036 0.165 203582 5.606 5.606 01-10-01 01:00:01 CONT B OK 0 300 6.25 0.107 .053 0.166 204356 5.622 5.622 01-10-01 01:05:01 CONT A OK 0 300 6.25 0.107 .041 0.165 200855 5.530 5.530 01-10-01 01:10:01 CONT B OK 0 300 6.25 0.107 .044 0.166 202105 5.561 5.561 01-10-01 01:15:01 CONT A OK 0 300 6.25 0.108 .037 0.165 201903 5.560 5.560 01-10-01 01:20:01 CONT B OKF 1 300 6.26 0.108 .024 0.112 13488 2.892 2.892 01-10-01 01:25:01 CONT A NPF 1 300 6.25 0.108 .049 .000 0 .000 .000 01-10-01 01:30:01 CONT B OKF 1 300 6.25 0.108 .029 0.108 1438 0.309 0.309 01-10-01 01:35:01 CONT A OKF 2 300 6.25 0.108 .029 0.109 5205 1.120 1.120 01-10-01 01:40:01 CONT B OKF 2 300 6.25 0.108 .042 0.119 41463 8.914 8.914 01-10-01 01:45:01 CONT A OKF 2 300 6.25 0.107 .040 0.109 5696 1.226 1.226 01-10-01 01:50:01 CONT B OKF 2 300 6.25 0.108 .023 0.109 4865 1.046 1.046 01-10-01 01:55:01 CONT A OKF 3 300 6.25 0.108 .021 0.144 126103 27.124 27.124 01-10-01 02:00:01 CONT B OKF 3 300 6.25 0.108 .053 0.116 26556 5.709 5.709 01-10-01 02:05:01 CONT A OKF 3 300 6.25 0.105 .082 0.108 9549 2.054 2.054 01-10-01 02:10:01 CONT B OKF 1 300 6.25 0.105 .026 0.106 2561 0.550 0.550 01-10-01 02:15:01 CONT A NPF 1 300 6.25 0.107 .037 .000 0 .000 .000 01-10-01 02:20:01 CONT B OK 0 300 6.23 0.108 .050 0.165 197747 5.455 5.455 01-10-01 02:25:01 CONT A OK 0 300 6.25 0.108 .025 0.165 198108 5.456 5.456 01-10-01 02:30:01 CONT B OK 0 300 6.25 0.108 .047 0.165 199661 5.493 5.493 01-10-01 02:35:01 CONT A OK 0 300 6.25 0.108 .032 0.164 198080 5.455 5.455 01-10-01 02:40:01 CONT B OK 0 300 6.25 0.108 .063 0.165 197029 5.419 5.419 01-10-01 02:45:01 CONT A OK 0 300 6.25 0.108 .043 0.164 194434 5.355 5.355 01-10-01 02:50:01 CONT B OK 0 300 6.25 0.108 .032 0.164 196977 5.419 5.419 01-10-01 02:55:01 CONT A OK 0 300 6.25 0.108 .024 0.164 195478 5.384 5.384 01-10-01 03:00:01 CONT B OK 0 300 6.25 0.108 .052 0.165 198595 5.464 5.464 01-10-01 03:05:01 CONT A OK 0 300 6.25 0.108 .041 0.163 193697 5.335 5.335 01-10-01 03:10:01 CONT B OK 0 300 6.25 0.108 .064 0.164 194459 5.350 5.350 01-10-01 03:15:01 CONT A OK 0 300 6.25 0.108 .037 0.164 194536 5.357 5.357

  22. Operation

  23. Installation Requirements • Location • Need outdoor location within 25 ft. of 2537A • Fetch is very important • Vertical height. Should be 3-10 m above grade • Power requirements (when desorbing only) • Approx. 1200 VA required by Model 1130 • Further 1300 VA required by Model 1135 • Two dedicated standard 115 VAC 15A circuits • Pump module can share circuit with 2537A

  24. Sample Installations Figure 1 - Temporary Model 1130/35 Installation (AES, Quebec Canada) Figure 2 - Permanent Arctic Model 1130 Installation (NOAA, Barrow Alaska) Figure 3 - Model 1130-P (prototype) Mounted on roof of trailer (US EPA, NERL) Figure 4 - Temporary Model 1130-P Installation (US EPA & Florida DEP) Figure 5 - Model 1130-P sampling in Antarctica (GKSS & AWI, Germany)

  25. Operational Requirements • Denuder requires changing every week or two • RPF requires exchanging every 8 to 12 weeks • Weekly or bi-weekly site visits recommended • Filter changes, leak checks, etc. as required during site visits

  26. Lab Support • Extra denuders & RPF’s are usually kept on hand • Several units recoated / restuffed at one time • Blanked, capped and shipped to site • Tube furnace required for blanking • Source of DIW is required • Elements may be exchanged in the field using relatively unskilled personnel

  27. Contact fhs@tekran.com Visit www.tekran.com End: Detailed Operation