An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Surface measurements with ATLAS12A Matthew Domingo, Hartmut F.-W. Sadrozinski, VitaliyFadeyevZachary Galloway,Zhijun Liang SCIPP, UCSC

  2. Electrical tests of ATLAS12A mini sensorsUCSC status • Samples • ATLAS12A samples irradiated with protons at Birmingham by 27MeV protons • Samples have not been annealed yet • 8 fluences : 1e12, 5e12,1e13,1e14,1e15,2e15,5e15 Neq/cm2 • Zones : BZ3C and BZ3F • This talk focus on the following measurements • R(interstrip) • C(interstrip) • PTP • Implant sheet resistance

  3. Inter-strip capacitance • C(interstrip) Measurements setups • BZ3C sensors , Measured in -10 °C , before annealed • 5 probe methods: • Middle strip connected to to LCR meters low side • First neighbor strips connected to LCR meters high side. • The Second neighbor strips grounded • 3 probe methods: • Similar to 5 probe methods • Except that the 2nd neighbor strips are not connected

  4. C(interstrip) before irradiation • ATLAS07 and ATLAS12 have similar C(interstrip) • 3 probe measurement are ~15% higher wrt 5 probe methods • due to second neighbor contribution C(interstrip) [pF/cm] Frequency = 1MHz Voltage(V)

  5. Inter-strip capacitance VS frequency • Measured BZ3C sensors by 3 probe method at -10 °C • We test Frequency range from 50kHz to 2MHz. • Low frequency region has a higher variance. • Frequency dependence is reduced when Frequency >500kHz Inter-strip capacitance ( pF/cm) Inter-strip capacitance ( pF/cm) Frequency [kHz] Frequency [kHz]

  6. Bias Voltage dependence • Inter-strip capacitance shows a stronger voltage dependence at low frequency (i.e. 100kHz) Inter-strip capacitance ( pF/cm) Inter-strip capacitance ( pF/cm) 1MHz 100kHz Bias voltage(V) Bias voltage(V)

  7. Inter-strip capacitance VS Fluence • Main result: Less impact by radiation in high frequency region • Inter-strip capacitance varied by 15% In high frequency region (>500kHz) for most of samples • One outliner varyied by 50% at Fluence =1e14. • Inter-strip capacitance In low frequency region (eg. 100kHz) is not stable • Measured by 3 probe method with 300 V reserve bias at -10 °C Inter-strip capacitance ( pF/cm) Before irradiation 100kHz 1Mhz • Fluence [Neq/cm2]

  8. Inter-strip resistance • The inter-strip resistance setup as follow • BZ3C sensors • Measured at -10 °C Measure nA current for GΩlevel resistance Relative low S/N Do a careful job in grouding to reduce the noise

  9. Inter-strip resistance VS Fluence (Φ) • R(interstrip) drops significantly after Φ>1e14 • Reduce from GΩlevel to MΩ level • R(interstrip) /R(bias) <10, when Φ>2e15 • ATLAS Specification R(interstrip) >10*R(bias) Before irradiation R(interstrip) /R(bias) • R(interstrip) (Ohm) • Fluence [Neq/cm2] • Fluence [Neq/cm2]

  10. PTP measurements • Measure resistance of PTP structure vs. voltage on implant (DC pad) • Measure punch-through voltage after irradiation • Compare the different PTP structure • Mainly the BZ3C and BZ3F

  11. BZ3C VS BZ3F • BZ3C has a smaller punch-through voltage before radiation. • After radiation, both sensors have similar punch-through voltage • BZ3F have a higher asymptotic resistance before radiation Φ=5e15 Resistance(Ohm) Resistance(Ohm) Voltage (V) Voltage (V)

  12. resistance of PTP structure vs. voltage • As fluenceincrease • The resistance of PTP responses slower to voltage • punch-through voltage increased BZ3F Resistance(Ohm) Resistance(Ohm) BZ3C Voltage (V) Voltage (V)

  13. punch-through voltage VS Fluence • Punch-through voltage (V) BZ3C BZ3F • Fluence [Neq/cm2]

  14. Implant sheet resistance • Measurement of sheet resistance of n-implant • ATLAS specification: Implant resistance < 20 KΩ/cm • Measured resistance per strip ~ 19KΩ/cm • Uniform in all the strips • No dependence of bias voltage Implant resistance (KΩ/cm) Implant resistance (KΩ/cm) Bias voltage(V) Strip number

  15. ATLAS12A VS ATLAS12A endcap • The results of ATLAS12A compared to “ATLAS12” and “ATLAS12A endcap” • See strong drop of R(interstip) with fluence increase as others sensors • R(interstip) Drop to MΩ level at Φ>2e15 in this result and Prague result (ATLAS12A EC) • R(interstip) Drop to MΩ level at Φ>1e15 in Lancaster measurement(ATLAS12A) • See need to see higher frequencies for C(interstip) measurements • At 100kHz, see larger Variation (>100%) • Larger than Lancaster result (ATLAS12A) (20%) and Prague result (ATLAS12A EC)“ (~50%) • At 1MHz, See similar variation (20%) with fluence as other measurements • “ATLAS12A endcap” results by Prague group: • https://indico.cern.ch/event/312262/contribution/6/material/slides/0.pdf • ATLAS12 results by Lancaster: • https://indico.cern.ch/event/292247/contribution/12/material/slides/1.pdf

  16. Summary • ATLAS12A mini sensors irradiated by protons are measured. • Inter-strip capacitance in irradiated samples • Increased by one order of magnitude at low frequency (100k) • Increased by 10~20 % at high frequency • ATLAS Specification < 0.8 pF/cm at 100kHz • does not meet the specification at 100kHz When fluences >1e13 • Very close to meet the Specification (0.8 pF/cm) at 1MHz • May revise the specification • if the S/N is still large once we know ABCN130 performance • Inter-strip resistance • ATLAS Specification R_INT>10*R(bias) • Accorinding to Specification interstrip resistance is not sufficient after fluences >2e15 Neq/cm2 • After Φ>5e15, R_INT is about 0.3MOhm • Still larger than the amplifier's input impedance (kOhm) • PTP • There is no defined acceptance criterion in ATLAS Specification • BZ3C and BZ3F samples are measured. • PTP performances are acceptable at all fluences • Implant resistance • 19KΩ/cm, meet ATLAS specification (< 20 KΩ/cm)

  17. Next steps • Test BNL gamma radiated sensors • Compare protons radiation VS gamma radiation damage • Starting on laser-based dynamic PTP study • Aim to report the results in next meeting • Will work on sensors performance after annealing,

  18. Backup R(bias) • Fluence [Neq/cm2]

  19. resistance of PTP structure vs. voltage • As fluenceincrease • The resistance of PTP responses slower to voltage • punch-through voltage increased

  20. Backup: Bias Voltage dependence • Inter-strip capacitance shows a stronger voltage dependence at low frequency (i.e. 100kHz) Inter-strip capacitance ( pF/cm) Inter-strip capacitance ( pF/cm) 1MHz 100kHz Bias voltage(V) Bias voltage(V)