An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Specific Absorption Rate (SAR)An Overview and Status RABC Montreal 2007 Presented by Stuart Nicol APREL Laboratories Member of the Spectrum Sciences Institute

  2. RF Exposure and Measurement Standards • IEEE (USA & International) • ICNRP, NCRP • IEC (International) • CENELEC (Europe) • HC, IC, (Canada) • Australia/New Zealand Standards(Australia) • MIC (Korea) • ARIB (Japan) • IDA (Singapore) • Other “derivative” country standards (China, Taiwan, Russia, Mexico, Brazil…)

  3. Published Experimental Measurement Methodology Standards • IEEE 1528 “Recommended Practice for Determining the Peak Spatial-Average Specific Absorption Rate (SAR) in the Human Body Due to Wireless Communications Devices: Experimental Techniques” (APREL Laboratories have 4 staff members recoded in the standard as technical contributors) • IEC 62209 “Human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices – Human models, instrumentation, and procedures – Part 1: Procedure to determine the Specific Absorption Rate (SAR) for hand-held devices used in close proximity of the ear (frequency range of 300 MHz to 3 GHz)” Both Standards Harmonized

  4. IEEE 1528b Background (ICES-TC34) Work started in 2005 • Two working groups (membership from Europe, Asia, North America) • WG 1 Douglas (Motorola) Experimental Measurements(> 65 members) • Recommended Practice for Determining the Spatial-Peak Specific Absorption Rate (SAR) in the Human Body Due to Wireless Communications Devices: Experimental Techniques • WG 2 Kainz (FDA) Computational Methods • P 1529b Recommended Practice for Determining the Spatial-Peak Specific Absorption Rate (SAR) in the Human Body Due to Wireless Communications Devices: Computational Techniques • Objectives to define methodologies for determining SAR

  5. IEC 62209 Part 1 and 2 TC 106 WG4 • One working group (membership from Europe, Asia, North America) • TC106 Chairman Matthias Meier (Motorola) Title Human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices – Human models, instrumentation, and procedures – Part 1: Procedure to determine the Specific Absorption Rate (SAR) for hand-held devices used in close proximity of the ear (frequency range of 300 MHz to 3 GHz). Human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices - Human models, instrumentation, and procedures. Part 2: Procedure to determine the Specific Absorption Rate (SAR) in the head and body for 30 MHz to 6 GHz Handheld and Body-Mounted Devices used in close proximity to the Body.

  6. IEC & IEEE Standards Harmonization • Why harmonize? • Both standards groups accept that the head methodologies are sound so IEC decided to focus on extending frequency range and scope so as to apply the standard to other technologies • What are the implications? • Standards are becoming more complex (academic) where correction factors and uncertainty bias are becoming more applicable rather than physical quantification of measurements • How does/will this affect government regulators? • It can potentially effect post audit exercises where measurements will not be repeatable due to post processing factors

  7. Other Considerations for Standards • Intended use is key test for how to proceed • SAR (reactive near field) vs. MPE (20 cm+, “truly”?) • Head, hand, body, bystander voice/data (VoIP) combinations of usage/configuration • New devices/technologies • PORTABLE VS MOBILE • What about a bystander analysis if I utilize the mobile exclusion rule

  8. Implementation/Approval Issues • RF exposure requirements have changed the game for approvals of devices • Duty cycles, accessories • Exclusions, apparent exclusions, and erring on the side of caution • UWB, MiMO, WiMAX, implantable devices, etc?

  9. Government RSS102SC-6 IEEE 1528b IEC 62209 Part 2 IEEE 1528 IEC 62209 Part 1 Government Regulations (IC Perspective) & the use of International Standards I have a product how do I test? SAR systems must test to all standards and follow all methodologies

  10. SAR Test Evaluation Methods • Experimental methods include activities run on automated test systems e.g. ALSAS-10U • SAR systems provide data up to 6GHz • Requirements now for IEC/ISO 17025 accredited facilities • Homogenous assessments ONLY • Numerical methods include FDTD applications • Can be used to support experimental activities • Used for heterogeneous analysis • Used for implantable heterogeneous studies • Used for network planning efficiency and base station/repeaters positioning

  11. Applications • Device types (technologies) • Cellular Telephones, PDA’s, Laptop’s, Implantable, Body Worn, Telemetry • Manufacturers or integrators expect data from third parties (suppliers) ODM’s antenna manufacturers • Module manufacturers occasionally don’t test • Modular applications can be wide and varied • This is safety • High SAR high concern

  12. Example of Experimental Test System Experimental methods are run using automated SAR systems which assess data utilizing a homogeneous medium ALSAS-10U :- APREL Laboratories SAR Assessment System

  13. SAR Test Requirements • Head SAR in left and right, touch and tilt with antenna extended/retracted, device open and closed on the low mid and high channels for each mode • Body SAR to be tested inside or outside holster (belt clip) if no holster belt clip is available test either with 0mm separation or 15mm separation front and back, with accessories for low mid and high channels for each mode • Separation distances with respect to resonate source from tissue are now being excluded for some wireless devices • Area scan consisting of a minimum step resolution of 20mm^2 providing adequate coverage of the device (1.5 x device size or large enough to cover resonant point/antenna) • Zoom scan (volume 1g or 10g) consists of minimum 8mm^2 (X & Y) and 5mm in Z

  14. Experimental SAR Results • Limits defined in Canada are 1.6 W/kg for 1g • Limits are NOT harmonized globally (e.g. Europe & some parts of Asia 2.0 mW/kg for 10g average)

  15. Numerical Methods • Numerical SAR can be derived using FDTD code • Acceptable for some compliance applications • Strong tool for biological research • Used for circuit design • Excellent method for initial product development and final product improvements • Used for standards development (SAR, MPE and HAC) • Draw backs include time needed to produce and assess data • Limits will remain the same Standards are now being developed to allow for an accepted method to be used as a compliance alternative

  16. Numerical Results • SAR, magnetic field and electrical fields can be calculated to asses complex cases Homogeneous field distribution Heterogeneous model