Sensivate Post-Column Reactor

Sensivate Post-Column Reactor. Applications  Carbamates by OPA  Glyphosates by OPA  Aflatoxin by Iodine  Amino Acids by OPA  Amino Acids by Ninhydrin. System Features.  Add compact post-column capability to any LC system  Independent reagent pump control

wayde
Télécharger la présentation

Sensivate Post-Column Reactor

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SensivatePost-Column Reactor Applications Carbamates by OPA Glyphosates by OPA Aflatoxin by Iodine Amino Acids by OPA Amino Acids by Ninhydrin

  2. System Features Add compact post-column capability to any LC system Independent reagent pump control User-replaceable, disposable reactor coils Fast & easy setup: One inlet connection (from column) One outlet connection (to detector) All connections accessible from front panel Modular: 1 or 2 Reaction Coils (1 heated) 1 or 2 pumps Easy upgrade from single to dual reagent All PEEK fluid path

  3. Modular Pumps Slide out drawer for easy access and maintenance 2,500 p.s.i. upper pressure limit Automatic Piston Wash  Interactive front keypad: Flow rate set points Pressure read-out Set upper/lower pressure limit RS-232 Interface Pulsation: < 0.5% Integrated Prime-Purge Valve / Pulse Damper / Outlet Filter, with transducer Back Pressure Coil (for proper PD operation)

  4. “All-In-One” Integrated Fluidics Combination Pulse Damper / Prime Purge Valve / Outlet Filter / Transducer Eliminates Tubing and Fittings between Components Fewer Leak Points Very Low Dead Volume All PEEK Fluid Path

  5. Modular Reactors PEEK loop, sealed Operating Range: 10° C above ambient to 150° C Temp. Accuracy: ± 2° C Temp. Repeatability: ± 1° C. Front keypad control

  6. Easy to Replace Reactors Remove fittings Slide out drawer Lift out reactor Replace with an alternate size reactor or replacement unit

  7. All Fittings On The Outside Easy Set-Up Faster Maintenance Visible Leaks

  8. P/N Description Reactor Specification PC10AH000 Single Reagent (single pump) system for AAA, OPA, Polyamines & Paraquat/Diquat 0.15 ml reaction volume, heated PC10BH000 Single Reagent (single pump) system for AAA, Ninhydrin and Formaldehyde 0.50 ml reaction volume, heated PC10CH000 Single Reagent (single pump) system for Alfatoxins 1.40 ml reaction volume, heated PC10BH10A Dual Reagent (two pump) system for Carbamates and Glyphosate 0.50 ml reaction volume, heated 0.15 ml reaction volume, ambient PC10CH10C Dual Reagent (two pump) system 1.40 ml reaction volume, heated 1.40 ml reaction volume, ambient Available Configurations

  9. Application Reactor 1 Temp. Reactor 2 Carbamate NaOH 100o C OPA Glyphosate NaOCl 40o C OPA Aflatoxins Iodine 40o C (not used) Carbohydrates NaOH 40o C (not used) Lipids Phenacyl 100o C (not used) Amino Acids -Ninhydrin Ninhydrin 140o C (not used) Amino Acids - OPA NaOCl 40o C OPA Amino Acids - Primary OPA 30o C (not used) PCR System Applications

  10. Additional Features High pressure cut-off switch for plugged coil Low pressure cut-off system (with auto reset) for interface with HPLC