Setup for large area low-fluence irradiations with quasi-monoenergetic 0.1−5 MeV light ions

Setup for large area low-fluence irradiations with quasi-monoenergetic 0.1−5 MeV light ions. M. Laitinen 1 , T. Sajavaara 1 , M. Santala 2 and Harry J. Whitlow 1. 1 Department of Physics, P.O.B 35, FIN-40014 University of Jyväskylä, Finland

wing-burns
Télécharger la présentation

Setup for large area low-fluence irradiations with quasi-monoenergetic 0.1−5 MeV light ions

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Setup for large arealow-fluence irradiations withquasi-monoenergetic 0.1−5 MeVlight ions M. Laitinen1, T. Sajavaara1, M. Santala2 and Harry J. Whitlow1 1 Department of Physics, P.O.B 35, FIN-40014 University of Jyväskylä, Finland 2 Laboratory of Advanced Energy Systems, P.O.B 4100, FIN-02015 HUT, Finland email: mikko.i.laitinen@jyu.fi

  2. Motivation for the setup • Particle response of prototype detectors, requirements for testing facility: • Large area for testingmultiple detectors at thesame time • Up to 10 cm x 10 cm • Low fluence • Well known uniform distribution needed • Flux lower than 1 particle s-1 mm-2 • Monoenergetic beam • Energy distribution of the flux below 10 keV (FWHM), from 100 keV to 5 MeV • Quick change of beam energy From JET collaboration web-page

  3. Methods for low fluence large area irradiations • How to irradiate ? • Radioactive sources • Limited energy (~3-6 MeV) range • Limited ion range (only 4He ions) • Implanters/ion sources • High flux • Energy range limited to < 1 MeV(normally < 100 keV)

  4. Methods for low fluence large area irradiations • RADEF (JYFL cyclotron, with direct beam) • Minimum energy ~5 MeV • Small accelerators (with direct beam) • Scatterers/wobblers needed → energy spread • Small fluxes difficult • Large homogenous areas difficult

  5. The idea of the setup • Not using the primary beam but instead the secondary beam of the small linear accelerator in JYFL • Secondary beam from particles that have undergone backscattering is used in detector testing with well known properties • Target is made of thin self-supporting carbon foil (10 mg cm-2 ~50nm) where thin layer (2-25 nm) of single isotope element (Au, Rh, Nb, Co, Al eg.) has been deposited • Most of the primary beam goes through the target and hits the targetholder’s backwall from where it cannot backscatter to the detectors → virtually no background Sample holder backwall Sample holder Removable samplesupports Thin Au layer on top of carbon foil Backscattering from Au (+ C) to detectors IncomingPrimary beam0.2-5.2 MeV(for He)

  6. Setup at the Pelletron accelerator

  7. True valuefor backscattered beam closer to8 keV FWHM~16 keV Reference detector data • Flux and energy calibrated from reference detector and multiple targets • Implanted Si surface barrier detector ~14 keV FWHM energy resolution

  8. Reference spectra and target • Sample turret can be modified to take up to 20 targets • In current system 5 targets can be loaded same time to the chamber • Logarithmic scale shows that there is a minor background below Au peak • Backscattering starts at the components of steel • Also for part of the background the origin is due to multiple scattering especially for low energies

  9. Flux homogeneity • Can be easily calculated for both energy and intensity Figures of merit for the setup • Energy range • 150 – 5000 keV for He • 150 – 3000 keV for H • Flux • Up to 500 particles per secondper cm2 (100 – 5 s-1 cm-2) • For lowest energies maximumfluxes get lower • Cycle time • During last 2 day test period:He 1st day, H 2nd day, 5-7different accelerator energiesfor 3 different detector setsper day (up to 20 cycles per day). • Multiple detectors at once • 4 minute down-pumping time

  10. NPA – results from the setup • Neutral particle analyzers for Joint European Torus JET • 1st tests showed strange double peak behaviour and bad resolution • 2nd set of detectors performs much better: Improved resolution and effiency, no double peaks but small tail A novel silicon detector for neutral particle analysis in JET fusion research Kalliopuska J, Garcia F, Santala M, et al.NIM A, vol. 591, 1, p. 92-97 (2008)

  11. Future improvements • H - beam currents limited now by ion source • New ion source coming before summer • Order of magnitude increase to H currents • Heavier ions including Li available from new ion source • Beam blanker for the prototype detectors • Total fluxes and energies calibrated from reference detector before letting the beam to the test detector • Full understanding of reference spectrum through Monte Carlo simulations

  12. Thank you for your attention ! Accelerator based materials physics goup in JYFL

  13. NPA schematics

  14. NPA linearity for Hydrogen