Calibration of Volumetric Glassware By Gravimetric Method Nihal Gunasekara Sri Lanka

Bangladesh BEST Programme. Calibration of Volumetric Glassware By Gravimetric Method Nihal Gunasekara Sri Lanka. Bangladesh BEST Programme. Laboratory used Glassware. Burette Single-volume pipette Graduated pipettes One-mark volumetric flasks Graduated measuring cylinders.

zyta
Télécharger la présentation

Calibration of Volumetric Glassware By Gravimetric Method Nihal Gunasekara Sri Lanka

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Bangladesh BEST Programme Calibration of Volumetric GlasswareBy Gravimetric MethodNihalGunasekaraSri Lanka

  2. Bangladesh BEST Programme Laboratory used Glassware Burette Single-volume pipette Graduated pipettes One-mark volumetric flasks Graduated measuring cylinders

  3. Bangladesh BEST Programme DEFINITIONS Unit of volumeCubic centimeter : ml Ref. temperature 20 °C To deliver Apparatusiscalibrated to accurately transfer the stated volume to another container. To contain Apparatus is calibrated to an accurate measure of the volume contained

  4. Bangladesh BEST Programme Apparatus Needed Balance Thermometer Hygrometer Barometer Water Distilled water /De-ironized Pipette Sucker

  5. Bangladesh BEST Programme Recommended Apparatus - Balance

  6. Bangladesh BEST Programme Recommended Apparatus/Liquids

  7. Bangladesh BEST Programme Test Room Conditions M Temperature Maintained at 20 ± 1 °C and during calibration ± 0. 5 °C stability Humidity Maintained 35% to 85% Prefer to maintain at 60 ± 10% Stabilization Volumetric instruments and test water 1 to 2 hours

  8. Preparation Bangladesh BEST Programme Cleaning of volumetric glassware • Internal surface of volumetric vessels shall be cleaned sufficiently before the calibration • Oil or grease is removed by suitable solvent • The vessel should be nearly filled with an aqueous solution of a soap less detergent and shaken vigorously. It should then be repeatedly rinsed with distilled water • Be sure that vessel is cleaned before commence calibration

  9. Calibration Procedure Bangladesh BEST Programme • Prior to any calibration test make sure that all instruments reached the ambient temperature in the laboratory • During the calibration maximum temperature variation in the laboratory should not exceed ±0.5oC. • Set or read a meniscus against a reference line or scale • Be sure that the vessel or weighing bottle and the distilled water are at room temperature.

  10. Calibration Procedure Bangladesh BEST Programme • Fill distilled water up to the required volume • Weigh the testing container with distilled water. • Measure the temperature of distilled water using the thermometer. • Repeat the above step three times and record in the data recording form. • Repeat the above steps for cylinders at least at four different volumes

  11. Calibration Procedure –Flask & Cylinder (contain) Bangladesh BEST Programme • Weigh the empty testing container and record the value • Fill distilled water up to the required volume • Weigh the testing container with distilled water . • Measure the temperature of distilled water using the thermometer. • Repeat the above step 2 through 4 three times and record in the data recording form. • Repeat the above step 2 through 5 for cylinders at least at three different volumes

  12. Calibration Procedure – Pipette (deliver) Bangladesh BEST Programme • Weigh a empty container and record the value • Fill pipette with distilled water up to the index mark • Any water on the outside of the tip is removed with filter paper after the filling is completed • Transfer the distilled water to empty vessel, weighted in step (1) • Weigh the vessel with distilled water . • Measure temperature of distilled water using the thermometer • Repeat the above steps three times and record in the data recording form.

  13. Calibration Procedure – Burette (deliver) Bangladesh BEST Programme • Weigh empty container and record the value • Fill burettes with distilled water up to the index mark • Any water on the outside of the tip is removed with filter paper after the filling is completed • Transfer the distilled water to empty vessel, weighed in step (1) • Weigh the vessel with distilled water • Measure temperature of distilled water using the thermometer. • Repeat the above steps three times and record in the data recording form. • Repeat the above steps 1 through 7 for Burettes at least three different levels

  14. Bangladesh BEST Programme Principle of calibration - volumetric glass ware Bouncy effect Temperature effect Z values are given in Tables B6, B7 and B8 in ISO 4787 : 2010 for different types of glass at common air pressure Vs temperature

  15. Bangladesh BEST Programme Notations R L- is the balance reading of vessel with water reference liquid in grams RE -is the balance reading of empty vessel in grams Ρa - is the density of air, in grams per milliliter ρb - when weighing in air as though the density of the weight was 8.0 g/ml. ρw - is the density of water at toC, in grams per milliliter. ϒ - is the coefficient of cubical thermal expansion of the material of which the item of glassware, tested is made in reciprocal degrees Celsius, t - is the temperature of the water used in testing in degrees Celsius.

  16. Bangladesh BEST Programme Density of Air- Table B3 ISO 4787 Temperature Air Pressure hPa 930 940 950 960 970 980 990 1000 1010 19 1.104 1.116 1.1128 1.140 1.152 1.164 1.176 1.188 1.200 20 1.100 1.112 1.124 1.136 1.148 1.160 1.172 1.183 1.195 21 1.096 1.108 1.120 1.132 1.144 1.155 1.167 1.179 1.191 Density values in 10 grams per milliliter -3

  17. Bangladesh BEST Programme Density of water – Table B4 ISO 4787 Temperature Density(g/ml) °C ρw 19 0.99840 20 0.99820 21 0. 99799

  18. Coefficient of Cubical Thermal Expansion Bangladesh BEST Programme Material Coefficient of cubical thermal expansion ϒ-1 °C 10-6 Borosilicate glass 3.3 9.9 Borosilicate glass 5.0 15 Soda-lime glass 27

  19. Bangladesh BEST Programme Data Recording Use spread sheet for data recording Record Temperature, Humidity, Barometric pressure Record final data in a data recording sheet Review from a third party Prepare the final calibration certificate

  20. Bangladesh BEST Programme Thank you