Intoxilyzer 5000EN Factory Technical Specifications

Intoxilyzer 5000EN Factory Technical Specifications. Rev 3.09. Intoxilyzer 5000EN Factory Technical Specifications. RANGE: 0.005 to 0.450 BrAC grams/210 liters Majority of BrAC results are between 0.005 BrAC and 0.450 BrAC Results less than 0.005 BrAC = 0.000 BrAC on display

devi
Télécharger la présentation

Intoxilyzer 5000EN Factory Technical Specifications

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Intoxilyzer 5000ENFactory Technical Specifications Rev 3.09

  2. Intoxilyzer 5000ENFactory Technical Specifications • RANGE: 0.005 to 0.450 BrAC grams/210 liters • Majority of BrAC results are between 0.005 BrAC and 0.450 BrAC • Results less than 0.005 BrAC = 0.000 BrAC on display • Results greater than 0.450 Brac = Range Exceeded Error Message 0.005 BrAC 0.450 BrAC <0.005 BrAC = 0.000 BrAC >0.450BrAC = Range Exceeded Error Message Majority of BrAC Results Rev 3.09

  3. Intoxilyzer 5000ENFactory Technical Specifications • Accuracy: +3% or +0.003 BrAC which ever is greater • Same accuracy as similar laboratory equipment Rev 3.09

  4. Intoxilyzer 5000ENFactory Technical Specifications • Precision: Standard Deviation of 0.003 BrAC or better • If 10 tests of a known ethanol solution were run on the instrument, all results would be within 0.003 BrAC of each other Rev 3.09

  5. Intoxilyzer 5000ENFactory Technical Specifications • Test Time: Approximately 7 minutes • For Colorado test sequence data entry time + (ACABAIA2ABACA) Rev 3.09

  6. Intoxilyzer 5000ENFactory Technical Specifications • Operating Temperature: • 68 F to 86 F (20 C to 30 C) room temperature • Keep room temperature at approximately 75 F Rev 3.09

  7. QUESTIONS? Rev 3.09