1 / 9

How to calibrate TOF counters

How to calibrate TOF counters. Slewing procedure Timing resolution of all region(40cm). Setup of beam line. Type : phase1 p=3GeV/c. TOF3. TOF. ref2. Types of calibration. Slewing : TDC.vs.ADC ⇒ To estimate bias on timing measurement Calibration 1 : w/ref2 , done

dawn
Télécharger la présentation

How to calibrate TOF counters

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. How to calibrate TOF counters Slewing procedure Timing resolution of all region(40cm)

  2. Setup of beam line • Type : phase1 • p=3GeV/c TOF3 TOF ref2

  3. Types of calibration • Slewing : TDC.vs.ADC ⇒To estimate bias on timing measurement • Calibration 1 : w/ref2 , done • Difference between PMTA-PMTB • ->Ideal timing resolution canceling slewing • Calibration 2 : w/ref2, not yet (tof3 done) • Center,width 5mm : No position dependence • From TOF-T1,TOF-ref2,T1-ref2 plots, we can calculate intrinsic timing resolution of each counter • Slewing : tdctofa vs adctofa , tdctofb vs adctofb

  4. Types of calibration2 • Calibration 3 : Calibration of all region(40cm) of the counter • ref2 can’t operate on magnetic field • To compare w/o magnetic field to w/ magnetic field …., calibration of all region • This beam test : w/o spectrometer • We can’t determine the hit position on TOF counter in magnetic field without tracker. • TDC : We have to erase the factor of position. • The solution is TDCave=(tofa+tofb)/2 • Slewing : TDCave vs adctofa , TDCave vs adctofb

  5. -26000 tof3ave-t1ave(tdc,ps) This curve is similar to a•sin(x)+bx+c -31000 0.125 0.325 0.125 0.325 1/sqrt(adctof3a) Time Walk • There are two components in these plot

  6. tof3ave-t1ave(tdc,ps) 100 0 -100 0.12 0.28 0.14 0.28 Time Walk (after correction) • We used sin curve correction • There remained a small distortion even after time walk corrections (1/sqrt(adctof3a)+1/sqrt(adctof3b))/2

  7. Timing Resolution in all region • tof3ave-t1ave σ=83.1ps • σ(t1)=41.8ps ⇒σ(tof3)=71.8ps (all region) tof3ave-t1ave tdc(ps)

  8. Timing resolution of all TOF Timing resolution • There are no difference between w/ magnetic field and w/o magnetic field # of TOF

  9. Summary • Ideal resolution is estimated from the difference between PMTA - PMTB〜60ps • We calibrate the slewing effect on TOF counters in the magnetic field.〜70ps • No significant degrade on timing resolution in the magnetic field.

More Related