Step 1. Configure Hyper-terminal to communicate with SC serial interface through Com 1 serial port. Connection speed at 115200. Flow control to “None”. Step 2A. Configure Hyper-terminal to communicate with SC serial interface through Com 1 serial port. 132 column mode. Emulation VT100.
E N D
Step 1. Configure Hyper-terminal to communicate with SC serial interface through Com 1 serial port Connection speed at 115200 Flow control to “None”
Step 2A. Configure Hyper-terminal to communicate with SC serial interface through Com 1 serial port 132 column mode Emulation VT100 ASCII
Step 3. Push Reset button or Turn Abacus50 Off then ON to restart system
Step 4. Before count down to zero, press any key to stop auto booting procedure
Step 5. The HyperTerminal window will display four options, change the network configuration and save settings
Step 6. After updating, Abacus will reboot and you can check the new configurations during boot procedure