An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Brightness of Horizontal (H5) and Vertical (H1) Cold Sources Ken Andersen

  2. H5 H1

  3. ExtractBrightnessfrom Neutron Measurement • Needs end-of-guide position with chopper and TOF analysis • Readilyavailable on H1: Figaro, D17, PF1B • Not soreadilyavailable on H5 (IN15?) • Propagatebackwards • Correct for attenuationthroughwindowmaterials etc. • Scaleusing Au-foil measurement

  4. Use spectrummeasured on Figaro (H172)

  5. Detector spectrumcorrected for chopper opening time Δt Δt distance time

  6. Al Bragg edgesmeasured on D17

  7. Al window transmission T=0.875 ⇒t=20mm

  8. Al window transmission

  9. Detector efficiency Windowthickness = 10mm - half of the total 20mm

  10. Air attenuation Ageronmeasurementat 60% humidity 50cm through air

  11. Attenuationthrough frame-overlapmirror 0.7mm Si at 3° angle

  12. Final correctedspectrum Flux (arb. units)

  13. Propagatebackwardsthrough guide

  14. Gold foil measurement August 2007 after installation of H17 guides at VS position: entrance intolevel C casemate H171 (Figaro) Φcapt=1.07×1010 n/cm2/s H172 (Granit) Φcapt=1.57×1010n/cm2/s

  15. Evaluate capture flux at VS and rescale

  16. Comparison of capture fluxes

  17. Spectral flux at Granit monochromator

  18. Brightnesses of HCS & VCS

  19. Summary • Previousestimate of 8.9Å flux at Granit was about right • Ageron VCS brightnessisslightlylow • needs 50% readjustment in the 1.5-4Å range • needs 20% readjustmentaround 9Å • HCS and VCS brightnesses are probably about the same (within 30%)