1 / 15

PREDICTION OF UNCOMFORTABLE LOUDNESS LEVELS IN NORMAL HEARING ADULTS FROM THE AUDITORY STEADY STATE RESPONSES

Universidad de La Laguna. PREDICTION OF UNCOMFORTABLE LOUDNESS LEVELS IN NORMAL HEARING ADULTS FROM THE AUDITORY STEADY STATE RESPONSES. Barajas de Prat, J.J. (1) , Zenker Castro, F. (2) , Mora Espino, R. (3) and Fernández Belda, R. (2).

remy
Télécharger la présentation

PREDICTION OF UNCOMFORTABLE LOUDNESS LEVELS IN NORMAL HEARING ADULTS FROM THE AUDITORY STEADY STATE RESPONSES

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Universidad de La Laguna PREDICTION OF UNCOMFORTABLE LOUDNESS LEVELS IN NORMAL HEARING ADULTS FROM THE AUDITORY STEADY STATE RESPONSES Barajas de Prat, J.J. (1), Zenker Castro, F. (2), Mora Espino, R. (3) and Fernández Belda, R. (2) • Universidad de La Laguna (2) Clínica Barajas (3) Fundación Canaria Dr. Barajas para la Prevención e Investigación de la Sordera • SANTA CRUZ DE TENERIFE - CANARY ISLANDS - SPAIN XXIX International Congress of Audiology, Hong Kong, June 8 - 12, 2008

  2. Uncomfortable Loudness Levels Uncomfortable Loudness Level Amplified Speech Spectrum Threshold

  3. ULL Measurement

  4. ULL Objective estimation Acoustic Reflex Auditory Brainstem Response Thornton, Scand Audiol, 1989 Kiessling, Audiology, 1980

  5. ASSRs Recording Procedure 500 Hz 3 kHz Acoustic Stimuli: AM Tones .5 & 3 kHz dB re ULL from -10 to -60 Amplifier: Low Cut (A): 20 Hz High Cut (B): 300 Hz Gain: 100 000 Notch: On Electrode Arrangement: Non-inverted: Cz Inverted:1M Ground: forehead Impedance < 5k

  6. ULLs Measurement Procedure Electrical Waveform 500 Hz 3 kHz Acoustical Spectrum 500 Hz 3 kHz

  7. Characteristic of the study group Means and SDs audiometric thresholds (N=14)

  8. ASSR Recordings ULL 500 Hz Right ear ULL = 90 dB HL Left ear ULL = 90 dB HL

  9. Subjective ULL (dB HL)

  10. 3 kHz Amplitud-Intensity function(n = 14)

  11. 500 Hz Amplitud-Intensity function(n = 14)

  12. Inter subject average amplitude-intensity function (n = 14)

  13. Objective ULLs predicted from the intensity-amplitude function

  14. Conclusions • High intersubject variabilty in the amplitude-intensity function was found. • Higher correlations was found for the 3 kHz carrier frequency. • In general a stepper amplitude - intensity function were found for the left ear than for the rigth ear. • ULL prediction based solely on individual ASSR amplitude-intensity function proved to be problematic.

  15. WELCOME! EFAS 2009 TENERIFE SPAIN