1 / 87

ECG in Pacemaker Malfunction

ECG in Pacemaker Malfunction. Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai. ECG in “Pacemaker Malfunction” (ECGs in Pacemaker Function Assessment). Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

vyork
Télécharger la présentation

ECG in Pacemaker Malfunction

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai

  2. ECG in “Pacemaker Malfunction”(ECGs in Pacemaker Function Assessment) Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai

  3. Implantable Pacemaker Systems Contain the Following Components: Lead wire(s) Implantable pulse generator (IPG)

  4. Pseudomalfunctions Pseudomalfunctions are defined as: • Unusual • Unexpected • Eccentric ECG findings that appear to result from pacemaker malfunction but that represent normal pacemaker function

  5. ECGs in Patients with Pacemakers Overview : • Basic principles of pacing • Background information required • Systematic approach • Examples with single chamber pacing • Examples with dual chamber pacing

  6. ECGs in Patients with Pacemakers Overview : • Basic principles of pacing • Background information required • Systematic approach • Examples with single chamber pacing • Examples with dual chamber pacing

  7. ECGs in Patients with Pacemakers What do pacemakers do ? • Pace • Sense – ( what ? ) • Respond – Inhibit , Trigger or Dual • Respond to increased metabolic demand by providing rate responsive pacing • Provide diagnostic information stored by the pacemaker

  8. Timing Intervals Are Expressed in Milliseconds • One millisecond = 1 / 1,000 of a second

  9. Rate Conversion • Conversion • Pacing rate in PPM divided by 60,000 = ms • 60,000 / 60 PPM = 1000 ms • Interval in ms divided by 60,000 = PPM • 60,000 / 1000 ms = 60 PPM 60,000 ms BPM

  10. Paced Rhythm Recognition VVI / 60

  11. Paced Rhythm Recognition VVI / 60

  12. Paced Rhythm Recognition AAI / 60

  13. Noncapture is Exhibited By: • No evidence of depolarization after pacing artifact Loss of capture

  14. No Output • Pacemaker artifacts do not appear on the ECG; rate is less than the lower rate Pacing output delivered; no evidence of pacing spike is seen

  15. Fusion Beat • Definition: The combination of an intrinsic beat and a paced beat. • The morphology varies; in other words, a fusion beat doesn’t really look like a paced beat or an intrinsic beat. • The pacemaker and the patient contribute to depolarization in Fusion beats.

  16. Fusion • Ventricular Fusion

  17. Pseudofusion Beat • Definition: A pacing pulse falls on an intrinsic beat. The pacing pulse is ineffective and the intrinsic complex is not altered. • The pacemaker does NOT contribute to depolarization in Pseudofusion beats.

  18. Pseudofusion • Ventricular Pseudofusion

  19. Dual-Chamber Systems Have Two Leads: • One lead implanted in the atrium • One lead implanted in the ventricle

  20. Paced Rhythm Recognition DDD / 60 / 120

  21. Paced Rhythm Recognition DDD / 60 / 120

  22. Paced Rhythm Recognition DDD / 60 / 120

  23. Paced Rhythm Recognition DDD / 60 / 120

  24. Sensing • Sensing is the ability of the pacemaker to “see” when a natural (intrinsic) depolarization is occurring • Pacemakers sense cardiac depolarization by measuring changes in electrical potential of myocardial cells between the anode and cathode

  25. Sensitivity – The Greater the Number, the Less Sensitive the Device to Intracardiac Events

  26. Sensitivity 5.0 2.5 Amplitude (mV) 1.25 Time

  27. Sensitivity 5.0 2.5 Amplitude (mV) 1.25 Time

  28. Sensitivity 5.0 2.5 Amplitude (mV) 1.25 Time

  29. Undersensing . . . • Pacemaker does not “see” the intrinsic beat, and therefore does not respond appropriately Scheduled pace delivered Intrinsic beat not sensed VVI / 60

  30. VVI / 60 Oversensing • An electrical signal other than the intended P or R wave is detected ...though no activity is present Marker channel shows intrinsic activity...

  31. ECGs in Patients with Pacemakers Overview : • Basic principles of pacing • Background information required • Systematic approach • Examples with single chamber pacing • Examples with dual chamber pacing

  32. ECGs in Patients with Pacemakers Basic Data : • Clinical details – age , indication for pacing, time since implant etc • Type of pacemaker • Programmed parameters • Magnet Behaviour • Special features

  33. Magnet Operation • Magnet application causes asynchronous pacing at a designated “magnet” rate

  34. Rate Responsive Pacing • An accelerating or decelerating rate may be perceived as anomalous pacemaker behavior VVIR / 60 / 120

  35. Hysteresis • Allows a lower rate between sensed events to occur; paced rate is higher Hysteresis Rate 50 ppm Lower Rate 70 ppm

  36. ECGs in Patients with Pacemakers Overview : • Basic principles of pacing • Background information required • Systematic approach • Examples with single chamber pacing • Examples with dual chamber pacing

  37. ECGs in Patients with Pacemakers Single chamber pacing : • Identify underlying intrinsic rhythm if any ( in each chamber) • Verify appropriate sensing ( if possible ) • Verify capture • Measure base rate and check if appropriate • Identify any variations in intervals and interpret • Identify causes of inappropriate sensing or pacing

  38. ECGs in Patients with Pacemakers Overview : • Basic principles of pacing • Background information required • Systematic approach • Examples with single chamber pacing • Examples with dual chamber pacing

  39. Single Chamber ECG Analysis Programmed Parameters Mode………………………………………….. VVI Base Rate……………………………………….. 70 ppm Magnet Response…………………….. Battery Test Hysteresis Rate………………………………… Off ppm T Temporary programmed value 1.0 Second 7 Mar 2000 23:20 1

  40. ECG #1 • VVI • Normal Capture and Sensing 1

  41. Single Chamber ECG Analysis Programmed Parameters Mode………………………………………….. VVI Base Rate……………………………………….. 70 ppm Magnet Response…………………….. Battery Test Hysteresis Rate………………………………… Off ppm T Temporary programmed value 1.0 Second Jun 14 1999 2:57 pm 2

  42. ECG #2 • VVI • Normal Sensing • Capture unknown 2

  43. Single Chamber ECG Analysis 3

  44. ECG #3 • VVI • Normal Capture and Sensing with initiation of Hysteresis 3

  45. Single Chamber ECG Analysis 4

  46. ECG #4 • VVI • Loss of Ventricular Sensing • Ventricular Undersensing with Functional loss of capture on the second to last beat 4

  47. Single Chamber ECG Analysis 5

  48. ECG #5 • VVI • Loss of Ventricular capture with functional loss of sensing 5

  49. Single Chamber ECG Analysis 6

  50. ECG #6 • VVI • Normal Capture • Ventricular Oversensing 6

More Related