1 / 79

INSTRUMENTAL ANALYSIS CHEM 4811

INSTRUMENTAL ANALYSIS CHEM 4811. CHAPTER 4. DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university. CHAPTER 4 INFRARED SPECTROSCOPY (IR). IR SPECTROSCOPY. - Used for qualitative identification of organic and

kyle
Télécharger la présentation

INSTRUMENTAL ANALYSIS CHEM 4811

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER 4 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university

  2. CHAPTER 4 INFRARED SPECTROSCOPY (IR)

  3. IR SPECTROSCOPY - Used for qualitative identification of organic and inorganic compounds - Used for checking the presence of functional groups in molecules - Can also be used for quantitative measurements of compounds - Each compound has its unique IR absorption pattern - Wavenumber with units of cm-1 is commonly used - Wavenumber = number of waves of radiation per centimeter

  4. IR SPECTROSCOPY - The IR region has lower energy than visible radiation and higher energy than microwave The IR region is divided into Near-IR (NIR): 750 nm – 2500 nm Mid-IR: 2500 nm – 20000 nm Far-IR: 20000 nm – 400000 nm

  5. IR ABSORPTION BY MOLECULES - Molecules with covalent bonds may absorb IR radiation - Absorption is quantized - Molecules move to a higher energy state - IR radiation is sufficient enough to cause rotation and vibration - Radiation between 1 and 100 µm will cause excitation to higher vibrational states - Radiation higher than 100 µm will cause excitation to higher rotational states

  6. IR ABSORPTION BY MOLECULES - Absorption spectrum is composed of broad vibrational absorption bands - Molecules absorb radiation when a bond in the molecule vibrates at the same frequency as the incident radiant energy - Molecules vibrate at higher amplitude after absorption - A molecule must have a change in dipole moment during vibration in order to absorb IR radiation

  7. IR ABSORPTION BY MOLECULES Absorption frequency depends on - Masses of atoms in the bonds - Geometry of the molecule - Strength of bond - Other contributing factors

  8. DIPOLE MOMENT (µ) µ = Q x r Q = charge and r = distance between charges - Asymmetrical distribution of electrons in a bond renders the bond polar - A result of electronegativity difference - µ changes upon vibration due to changes in r - Change in µ with time is necessary for a molecule to absorb IR radiation

  9. DIPOLE MOMENT (µ) - The repetitive changes in µ makes it possible for polar molecules to absorb IR radiation - Symmetrical molecules do not absorb IR radiation since they do not have dipole moment (O2, F2, H2, Cl2) - Diatomic molecules with dipole moment are IR-active (HCl, HF, CO, HI) - Molecules with more than two atoms may or may not be IR active depending on whether they have permanent net dipole moment

  10. PRINCIPAL MODES OF VIBRATION Stretching - Change in bond length resulting from change in interatomic distance (r) Two stretching modes - Symmetrical and asymmetrical stretching - Symmetrical stretching is IR-inactive (no change in µ)

  11. PRINCIPAL MODES OF VIBRATION Bending - Change in bond angle or change in the position of a group of atoms with respect to the rest of the molecule Bending Modes - Scissoring and Rocking - In-plane bending modes (atoms remain in the same plane) - Wagging and Twisting Out-of-plane (oop) bending modes (atoms move out of plane)

  12. PRINCIPAL MODES OF VIBRATION 3N-6 possible normal modes of vibration N = number of atoms in a molecule Degrees of freedom = 3N H2O for example - 3 atoms - Degrees of freedom = 3 x 3 = 9 - Normal modes of vibration = 9-6 = 3

  13. PRINCIPAL MODES OF VIBRATION Linear Molecules - Cannot rotate about the bond axis - Only 2 degrees of freedom describe rotation 3N-5 possible normal modes of vibration CO2 for example - 3 atoms - Normal modes of vibration = 9-5 = 4

  14. TRANSITIONS Fundamental - Excitation from the ground state Vo to the first excited state V1 - The most likely transition and have strong absorption bands V3 V2 V1 Vo

  15. FUNDAMENTAL TRANSITIONS Overtone - Excitation from ground state to higher energy states V2, V3, …. - Result in overtone bands that are weaker than fundamental - Frequencies are integral multiples of fundamental absorption - Fewer peaks are seen than predicted on spectra due to IR-inactive vibrations, degenerate vibrations, weak vibrations - Additional peaks may be seen due to overtones

  16. COUPLING - The interaction between vibrational modes - Two vibrational frequencies may couple to produce a new vibrational frequency Combinational Band ν1 + ν2 = ν3 Difference Band ν1 - ν2 = ν3

  17. VIBRATIONAL MOTION - Consider a bond as a spring c = speed of light (cm/s) f = force constant (dyn/cm; proportional to bond strength) f for a double bond = 2f for a single bond f for a triple bond = 3f for a single bond - M1 and M2 are masses of vibrating atoms connecting the bond

  18. INSTRUMENTATION Components - Radiation source - Sample holder - Monochromator - Detector - Computer

  19. RADIATION SOURCES Intensity of radiation should - Be continuous over the λ range and cover a wide λ range - Constant over long periods of time - Have normal operating temperatures between 1100 and 1500 K - Have maximum intensity between 4000 and 400 cm-1 (mid-IR region) - Modern sources include furnace ignitors, diesel engine heaters - Sources are usually enclosed in an insulator to reduce noise

  20. RADIATION SOURCES Mid-IR Sources Nernst Glowers - Heated ceramic rods - Cylindrical bar - Made of zirconium oxide, cerium oxide, thorium oxide - Heated electrically to 1500 K – 2000 K - Resistance decreases as temperature increases - Can overheat and burn out

  21. RADIATION SOURCES Mid-IR Sources Globar - Silicon carbide bar - Heated electrically to emit continuous IR radiation - More sensitive than the Nernst glower

  22. RADIATION SOURCES Mid-IR Sources Heated Wire Coils - Heated electrically to ~ 1100 oC - Similar in shape to incandescent light bulb filament - Nichrome wire is commonly used - Rhodium is also used - May easily fracture and burn out

  23. RADIATION SOURCES Near-IR Sources Quartz halogen lamp - Contains tungsten wire filament - Also contains iodine vapor sealed in a quartz bulb - Evaporated tungsten is redeposited on the filament increasing stability - Output range is between 25000 and 2000 cm-1

  24. RADIATION SOURCES Far-IR Sources High pressure Hg discharge lamp - Made of quartz bulb containing elemental Hg - Also contains inert gas and two electrodes

  25. RADIATION SOURCES IR Laser Sources - Excellent light source for measuring gases Two types Gas-phase (tunable CO2) laser and solid-state laser Laser - Light source that emits monochromatic radiation

  26. MONOCHROMATORS - Salt prisms and metal gratings are used as dispersion devices - Mirrors made of metal with polished front surface - Spectrum is recorded by moving prism or grating such that different radiation frequencies pass through the exit slit to the detector - Spectrum obtained is %T verses wavenumber (or frequency)

  27. FT SPECTROMETERS - Based on Michelson interferometer - Employs constructive and destructive interferences - Destructive interference is a maximum when two beams are 180o out of phase - An FT is used to convert the time-domain spectrum obtained into a frequency-domain spectrum - The system is called FTIR

  28. FT SPECTROMETERS - Has higher signal-to-noise ratio - More accurate and precise than dispersive monochromators (Conne’s advantage) - Much greater radiation intensity falls on the detector due to the absence of slits (throughput or Jacquinot’s advantage)

  29. FT SPECTROMETERS Focusing Mirrors of Interferometer - Made of metal and polished on the front surface - Gold coated to prevent corrosion - Focus source radiation onto the beam splitter - Focus light emitting from sample onto detector

  30. FT SPECTROMETERS Beam Splitter of Interferometer - Made of Ge or Si coated onto a highly polished IR-transparent substrate - Ge coated on KBr is used for mid-IR - Si coated on quartz is used for NIR - Mylar film is used for far-IR - Transmits and reflects 50% each of beam (ideally)

  31. DETECTORS Two classes Thermal detectors and photon-sensitive detectors

  32. DETECTORS Thermal Detectors - Bolometers - Thermocouples - Thermistors - Pyroelectric devices

  33. DETECTORS Thermal Detectors Bolometers - Very sensitive electrical resistance thermometer - Older ones were made of Pt wire (resistance change = 0.4% per oC) - Modern ones are made of silicon placed in a wheatstone bridge - Is a few micrometers in diameter - Fast response time

  34. DETECTORS Thermal Detectors Thermocouples - Made up of two wires welded together at both ends - Two wires are from different metals - One welded joint (hot junction) is exposed to IR radiation - The other joint (cold junction) is kept at constant temperature - Hot junction becomes hotter than cold junction - Potential difference is a function of IR radiation - Slow response time - Cannot be used for FTIR

  35. DETECTORS Thermal Detectors Thermistors - Made of fused mixture of metal oxides - Increasing temperature decreases electrical resistance which is a function of IR radiation - Resistance change is about 5% per oC - Slow response time

  36. DETECTORS Thermal Detectors Pyroelectric Detectors - Made of dielectric materials (insulators), ferroelectric materials, or semiconductors - A thin crystal of the material is placed between two electrodes - Change in temperature changes polarization of material which is a function of IR radiation exposed - The electrodes measure the change in polarization

  37. DETECTORS Photon Detectors - Made of materials such as lead selenide (PbSe), indium gallium arsenide (InGaSa), indium antimonide (InSb) - Materials are semiconductors - S/N increases as temperature decreases (cooling is necessary) - Very sensitive and very fast response time - Good for FTIR and coupled techniques

  38. RESPONSE TIME - The length of time that a detector takes to reach a steady signal when radiation falls on it - Response time for older dispersive IR spectrometers were ~ 15 minutes - Response time is very slow in thermal detectors due to slow change in temperature near equilibrium

  39. RESPONSE TIME Semiconductors - Are insulators when no radiation falls on them but conductors when radiation falls on them - There is no temperature change involved - Measured signal is due to change in resistance upon exposure to IR radiation ( is rapid) - Response time is the time required to change the semiconductor from insulator to conductor (~ 1 ns)

  40. SATURATION - Very high levels of light can saturate detector - Linearity is very important especially in the mid-IR region

  41. TRANSMISSION (ABSORPTION) TECHNIQUE - Can be used for both qualitative and quantitative analysis - Provides very high sensitivity at relatively low cost - Sample cell material must be transparent to IR radiation (NaCl, KBr) - Samples should not contain water since materials are soluble in water

  42. TRANSMISSION (ABSORPTION) TECHNIQUE Solid Samples Three sampling techniques Mulling, pelleting, thin film Mulling - Samples are ground to power with a few drops of viscous liquid like Nujol (size < 2 µm) - 2-4 mg sample is made into a mull (thick slurry) - Mull is pressed between salt plates to form a thin film

  43. TRANSMISSION (ABSORPTION) TECHNIQUE Solid Samples Three sampling techniques Mulling, pelleting, thin film Pelleting - 1 mg ground sample is mixed with 100 mg of dry KBr powder - Mixture is compressed under very high pressure - Small disk with very smooth surfaces forms (looks like glass)

  44. TRANSMISSION (ABSORPTION) TECHNIQUE Solid Samples Three sampling techniques Mulling, pelleting, thin film Thin Film - Molten sample is deposited on the surface of KBr or NaCl plates - Sample is allowed to dry to form a thin film on substrate - Good for qualitative identification of polymers but not good for quantitative analysis

  45. TRANSMISSION (ABSORPTION) TECHNIQUE Liquid Samples - Many liquid samples are analyzed as is - Dilution with the appropriate solvent may be necessary (CCl4, CS2, CH3Cl) - Solvent must be transparent in the region of interest - Salt plates are hygroscopic and water soluble (avoid water) - Special cells are used for water containing samples (BaF2, AgCl)

  46. TRANSMISSION (ABSORPTION) TECHNIQUE Gas Samples - Gas cells have longer pathlengths to compensate for very low concentrations of gas samples - Generally about 10 cm (commercial ones are up to 120 cm) - Temperature and pressure are controlled if quantitative analysis is required

  47. TRANSMISSION (ABSORPTION) TECHNIQUE Background Absorption - Absorption from material other than sample Two sources - Solvent - Air in the light path - Background spectrum is subtracted from sample spectrum - CO2 and H2O absorb IR radiation (sealed desiccants are used to reduce CO2 and H2O absorption)

  48. ATTENUATED TOTAL REFLECTION (ATR) TECHNIQUE - Uses high refractive index optical element - Called internal reflection element (IRE) or ATR crystal (germanium, ZnSe, silicon, diamond) - Sample has lower refractive index than the ATR crystal - Sample is placed on the ATR crystal - When light travelling in the crystal hits the crystal-sample interface, total internal reflection occurs if the angle of incidence is greater than the critical angle

  49. ATTENUATED TOTAL REFLECTION (ATR) TECHNIQUE - The beam of light penetrates the sample a very short distance - This is known as the evanescent wave - Light beam is attenuated (reduced in intensity) if the evanescent wave is absorbed - IR absorption spectrum is obtained from the interactions between the evanescent wave and the sample

  50. SPECULAR REFLECTANCE TECHNIQUE - Occurs when angle of reflected light equals angle of incident light (45o angle of incidence is typically used) - Occurs when light bounces off a smooth surface - Nondestructive method for studying thin films - Beam passes through thin film of sample where absorption occurs - Beam is reflected from a polished surface below the sample - Reflected beam passes through sample again

More Related