230 likes | 556 Vues
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING. REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING One meter is 3.3 feet Body parts should not rest on tube housing. Control panel should indicate. Condition of exposure When x-ray tube is being energized
E N D
TUBE HOUSING • REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING • One meter is 3.3 feet • Body parts should not rest on tube housing
Control panel should indicate • Condition of exposure • When x-ray tube is being energized • kVp, mA or mAs • Visible or audible signal of exposure
SID • Tape measure or laser lights indicate the distance • Must be accurate with 2% of the indicated SID
COLLIMATIONPBLBEAM ALIGNMENT • X-ray beam and light should be within 2% of SID • PBL not required anymore • Beam should line up with image receptor • Proper alignment of beam to film (indicator light)
FILTRATION • 2.5 mm @70 kVp • 1.5 mm between 50-70 kVp • .5 mm below 50 kVp (mammo) • See question on page 569 (refer to chart 31-3 on page 461)
Reproducibility • Linearity • Operator shield
MOBILE RADIOGRAPHY • Lead apron assigned to portable • Exposure switch should allow operator to be 2 meter from tube (6+)feet
FLUOROSCOPY • Source to skin distance – 38 cm • Mobile SSD – 30 cm • When intensifier is in parked position—no fluoro • Intensifier serves as a primary protective barrier and must be 2 mm Pb equivalent. • Filtration should be at least 2.5 mm Al equivalent—Tabletop, patient cradle or other material factored in for total filtration • Collimation—unexposed border should be visible on TV monitor
FLUOROSCOPY • Dead man type exposure switch • Bucky opening covered automatically by .25 mm lead • Protective curtain -- .25 mm Pb equivalent • Timer (audible) when fluoro time has exceeded 5 minutes
Intensity (R ) should not exceed 2.1 R per minute for each mA at 80 kVp DAP DOSE RESPONSE PRODUCT DOSE AND VOLUME OF TISSUE IRRADIATED DAP INCREASES WITH INCREASING FIELD SIZE FLUOROSCOPY
DESIGN CRITERIA • Location of x-ray table • Where is the primary beam directed? • Surrounding environment (controlled area vs. uncontrolled area) • RF room • Dedicated room • Use factor • # of exams in a room
Primary Protective Barrier • Anywhere the primary beam is directed ( dedicated chest rooms) • Lead bonded to sheet rock of wood paneling • Concrete, concrete block, brick • 4 inches of masonry = 1/16 inch of lead • Image intensifier considered a primary protective barrier
SECONDARY BARRIERS • Secondary radiation (scatter, leakage) • Patient is source of scatter • Barrier does not have to be leaded • gypsum board 4 thicknesses of 5/8th inch drywall • glass ½ to 1 inch thickness • lead acrylic • Control booth • Lead aprons (5mm of lead attenuates____%_at _____kVp
Factors that affect thickness of barrier • Distance • Occupancy-levels • Control vs uncontrolled • workload • Use factor
USE FACTOR • Amount of time x-ray beam is directed at wall/floor • Wall given a use factor of ¼ • Floor given a factor of 1 • Secondary barrier use factor of 1 • Dedicated chest room-use factor of 1
FINALLY • Barriers are designed with 75-100 kVp usage in mind so most barriers are thicker than needed • Exposure to outside of room is calculated to result in a DL of 100mrem per week but do not factor in patient and image receptor interception. DL is actually 1/10th of the recommended DL
Exposure switch • Mounted of fixed to control panel • No long cords