1 / 20

Designing for Radiation Protection

Designing for Radiation Protection. Bushong Ch. 38. Design Standards for Radiation Protection – Leakage Radiation. LEAKAGE RADIATION may not EXCEED. TUBE HOUSING 100mR / hr or 1 mGy/hr @ 1 meter. Control Panel .

margie
Télécharger la présentation

Designing for Radiation Protection

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Designing for Radiation Protection Bushong Ch. 38

  2. Design Standards for Radiation Protection – Leakage Radiation

  3. LEAKAGE RADIATIONmay not EXCEED • TUBE HOUSING100mR / hr or 1 mGy/hr @ 1 meter

  4. Control Panel • Must have an indicator when the x-ray tube is energized. Sometimes it will be a visible light or audible signal or both

  5. SID – Source to Image Distance • Indicator must be accurate to within 2% of the desired SID

  6. Source-to-Skin Distance • SSD is used to describe to distance between the tube and patient during fluoroscopy • SSD must not be less than 38 cm in fluoro rooms and 30 cm for mobile fluoro

  7. Collimation or PBL • Positive-Beam Limitation = PBL • Detects the size of the IR used and automatically collimates to size of the IR at standard SIDs • Used to be required by law between 1974 – 1994 but no longer required. PBL still remains standard in equipment design

  8. Filtration • General-purpose tubes (above 70 kVp) and the fluoroscopic beam must be filtered at least 2.5 mm Al

  9. Reproducibility & Linearity • Reproducibility = consistent exposures • Linearity = any mA station X msec will produce the same mAs

  10. Exposure switch • In diagnostic rooms the switch must be fixed to prevent the ability to make an exposure in the room • Mobile switch must be at least 6 ft long and the dead-man type • Fluoroscopic exposure control should be the dead-man type as well

  11. Radiography Room Design – Factors Affecting Barrier Thickness • Distance – between radiation source and barrier • Occupancy – the use of the area being protected (rarely occupied vs nursing station) • Control area – occupied by radiation workers or patients • Uncontrolled area – occupied by anyone

  12. MONITORING WORK AREAS • CONTROLLED AREA – Used by occupationally exposed personnel (monitored) • 100mrem or 1mSv / WEEK • UNCONTROLLED AREA – PUBLIC • 2 mrem or 0.02 mSv / week*

  13. ROOM SHIELDING • PRIMARY SHIELD = Any wall or portable wall that the primary beam could possibly be directed at • 1/16 Pb or eq. & 7 Feet High • 4 inches of Concrete will do!

  14. ROOM SHIELDING • SECONDARY – NO PRIMARY BEAM • 1/32 LEAD • CONTROL BOOTH – never point the beam at the control panel • BEAM SCATTERS 2X BEFORE HITTING • LEAD WINDOW – 1.5mm LEAD EQ

  15. Bucky Slot Cover & Protective Curtain • What is the minimum Pb eq. ?

  16. Image Intensifier is a primary barrier = 2 mm Pb equivalent CUMLATIVE TIMER FOR FLUORO SIGNALS AFTER 5 MINUTES OF FLUORO TO MAKE SURE THE RADIOLOGIST IS AWARE OF THE BEAM-ON TIME

  17. Exposure Calculations • Exposure = Exposure rate X mAs • Exposure rate varies by kVp selection and distance

  18. Exposure Calculations • At 75 kVp the exposure rate is 2.6 mR/mAs • What is the exposure? • AP Abdomen = 25 mAs • LPO L-spine = 32 mAs • PA Caldwell = 200 mA X 0.10 sec • AP Pelvis = 200 mA X 0.20 sec

  19. Exposure Calculations • 2.6 mR X 25mAs = 65 mR/mAs • LPO L-spine = 32 mAs • PA Caldwell = 200 mA X 0.10 sec • AP Pelvis = 200 mA X 0.20 sec

  20. Be Safe! Image gently!

More Related