210 likes | 324 Vues
Join our informative X-Ray webinar covering essential topics like online preventative maintenance, calibration backup, and the latest SUI updates for X-Ray units. Learn how to perform yearly maintenance checks, including electrical and mechanical inspections, and the importance of calibration phantoms. Discover how to access yearly maintenance procedures online and ensure your unit's optimal performance. Stay updated with the latest Schick updates and utilize the correct manuals for your X-Ray units, ensuring compliance and functionality.
E N D
X-Ray Webinar 3/27/2013 X-Ray update
Topics of the day • Online preventative maintenance. • Calibration backup • SUI updates • Schick Update
Yearly PM now live on-line also for X-Ray units • Required for 3D units covered by service contracts - Same log-in as for CAD/CAM units - Just choose “Imaging Products” on the left - Look for the correct manual for your unit. - Print out the checklist if you wish
Yearly maintenance (1) • Electrical/Mechanical check: (Follow the manual) • Check serial number of unit (bottom right of unit) • Check all buttons and lights and audible tone(s).(Dead-man switch release button).
Yearly maintenance (2) • Electrical check: (Follow the manual) • Measure Power outlet adequacy (voltage drop). • Tip: Can be measured later while you do the dose check in calibration.
Yearly maintenance (3) - Depending on machine there can be one (Galileos) to four (XG3D Ceph) lasers. • Attention: The Galileos laser can be obstructed • by headrest (if present). • (Take it out now, it must be removed for calibration • anyway). • Check lasers
Yearly maintenance (4) All good so far ? Relax… … a minute
Yearly maintenance (5) • Before you start: • Make sure you have all calibration phantoms !Pin Phantom (XG3D) • Geometry Phantom (XG3D, Galileos) Distortion Phantom (Galileos)
Yearly maintenance (6) • Ceph Phantom (XG3D Ceph) • Constancy test (XG3D, Galileos) 2D !
Yearly maintenance (7) • Go to sironasupport.comVideo page:
Yearly maintenance (8) • Make a backup of the calibration data !!! • If you don’t and the calibration fails, the unit will not work anymore.
Yearly maintenance (9) • Go through ALL steps start to finish, regardless if the symbols are green, yellow or red!(Except radiation field). Galileos: Tool can be downloaded from web-site: Sironasupport.com
Yearly maintenance (10) • Perform the Acceptance or Constancy test: • Sidexis -> Utilities -> Acceptance/Constancy test
Coming up: SUI updates (Software update image) • One Moment………... Ready?
SUI updates (1) • “SUI update” refers to the installation/upgrade of the “firmware” of any XG/Galileos X-Ray system. Sidexis Manager -> SiXabcon • Attributes -> software update. • Select SW package you want to add. 4.07…4.12
SUI updates (2) Do not pay too much attention to “red” and “green” BUT: Boards that show V00.00as their actual version might have an issue. There is a third way to update units: Modul-update which can be advantageous. ALWAYS IGNORE: Timout “vermutet” (expected). Wait until update is finished.
SUI updates (2) • When do you need a SUI update? 3D machines: With (almost) every (PC) software update. Spareparts!- come with “random” software loaded, not necessarily matching the unit. Unit old – sparepart new.Unit brand new – sparepart older. Attention: Sometimes SUIs have to be upgraded in steps…
SUI updates (3) • What to do? • Update/downgrade the spare part to unit Software? • Update the unit to (newer) spare part Software? CALL SIRONA Technical service!! This can both be dangerous!!!Older SW might not support newer units/parts, upgrading to newest SUI might lead to malfunctions with (older) PC software.