170 likes | 345 Vues
Vital Signs Monitor UConn BME 4900. Vital Signs Monitor. Purpose As the population ages, many people are required by their doctors to take vital signs on a daily basis. Developed for the average person to use at home. Vital signs transmitted to a computer and ultimately to a doctor.
E N D
Vital Signs Monitor • Purpose • As the population ages, many people are required by their doctors to take vital signs on a daily basis. • Developed for the average person to use at home. • Vital signs transmitted to a computer and ultimately to a doctor.
Measurements • What can it measure? • Electrocardiogram (ECG) • Three leads (Right arm, left arm, right leg)
What is an ECG? • Records a time waveform of heart electrical activity. • Used to diagnose heart problems • Arrhythmia • Myocardial infarction • Conduction blocks (bundle branch block)
Measuring Electrical Activity • 3 leads • 2 arm • 1 leg • As cardiac muscle depolarizes it creates a potential on the skin • Potential measured as a vector • Difference between right and left electrode measured with a differential amplifier.
Measuring Electrical Activity • Circuit consists of: • 3 op amp differential amplifier • Low pass filter (Cutoff: 0.05 Hz) • High pass filter (Cutoff: 160 Hz) • Gain Amplifier
Differential Amplifier Feedback Amplifier Buffer Differential Amplifier • What does the buffer do? • Provides a high input impedance which means low current flow • What is the purpose of C2 and C3? • Remove DC offset • Offset would be amplified by 1000x • 10 mV at 1000x gain = ~10V
Differential Amplifier • V1 & V2 are in the microvolt range • We need it to be in the milivolt range so Rf/R1 is set to ~1,000
1st Order LPF • The ECG is known to be a low frequency signal • Usually less than 100Hz • A LPF can be used to remove the high frequency noise in the signal
High Pass Filter • ECG frequency range ~0.05 Hz – 150 Hz • Why do we need to eliminate frequencies <0.05 Hz? • Avoid distortion of the ST wave
Microprocessor • Provides Analog to Digital (A/D) conversion • Waveform y(t) sampled at a fixed rate Δt • Voltage read every Δt and converted to a number • If resolution is 8 bits • Gives us 28 = 256 counts over 5V range • ~19.5 mV per count
Patient Isolation • Bluetooth is wireless technology • Bluetooth provides separation of the patient from the equipment and line source • Very important to protect patient from electrical shock
Labview Requirements • Receive serial data from bluetooth • Display raw and filtered EKG • Calculate heart rate