410 likes | 634 Vues
E N D
1. Detection and Identification of Chemical Warfare Agents in Environmental Samples Casper de Boer, Christof Mnch, Caroline Mnsterer
Analytical strategies, 18.10.2011
2. Outline
3. Definition Chemical Weapons
4. Chemical agents as a weapon can be spread by: bomb/explosion
crop-dusting plane
aerosol device
...
? tequniques of storing and filling are important (mainly used as deterrent weapons)
5. Dispersal and Fate of Chemical Warfare Agents
6. Example for chemical warfare agents
7. Schedules of chemicals
9. Historical overview Ancient world: salinisation of cropland, poisoned arrows, ...
WW I: tear gas, chlorine gas (lung damage), mustard gas (blistering)
1925: Geneva protocol
WW II: extensive research on CWs but little use. New agents: tabun, sarin, soman (nerve agents)
Vietnam: herbicidal warfare ? use of chemicals to destroy vegetation, food crops (Agent Orange)
From late 60s on: towards chemical disarmament
1997: CWC (chemical weapon convention) enters into force ? arms control agreement between (today 188) states which prohibits production, stockpiling and use of CWs
10. Chemical Weapons Convention CWC
11. Organization for the Prohibition of Chemical Weapons (OPCW)
12. Which analytical equipment would you choose for your OPCW designated laboratory with respect to a limited budget of 2 Mio ?
13. - we must fulfill internationsal standard General requirements for the competence of testing and calibration laboratories (ISO/IEC 17025) - national accreditation by an internationally recognized accreditation body - we must regularly participate and perform successfully in inter- laboratory OPCW Proficiency Tests - respect a limited budget
14. - a huge number of chemicals with different characteristics and unknown concentrations - different and difficults matrices - analysis of each sample by at least two different analytical techniques (one of them has to be a spectrometric technique) - identification must be confirmed by the comparison with reference data.
28. ->Neutral and weakly basic pesticides, such as organophosphates, triazines, and arylureas, ionized well using either method of ionization, positive APCI generally being the more sensitive. Cationic pesticides (paraquat and diquat) and those that exist predominantly as anions in solution (e.g. sulfonic acids) gave much better responses using ESI (positive or negative).
Another factor to consider in the choice of ionization method is the formation of adducts with sodium or other metal ions, in addition to the protonated molecule. Although these may help con?rm the molecular mass, they tend to lower the signal-tonoise ratio of the protonated molecule, resulting in higher detection limits in trace analysis and causing problems in quantitative analysis. Sodium adducts tend to form with positive ESI but not with APCI
Coupled to capillary
or micro LC, it is more sensitive than particle
beam and, like the latter, provides searchable spectra
similar to those produced by electron ionization (EI)
in gas chromatography/mass spectrometry (GC/MS).
It can also operate in CI mode.
It is reported to be more widely applicable than
APCI, extending the useful range of LC/MS to
lower polarity molecules, for example, polyaromatic
hydrocarbons, and may be more sensitive than APCI
for some neutral analyte
Quadrupole mass spectrometers coupled with
LC make up a large part of the market.
Single
quadrupole instruments are limited for unequivocal
identi?cation of unknown analytes in that CID
is limited to in source. As already stated, this
produces additional chemical noise and does not
resolve individual components when coelution from
the LC column occurs. For analysis of complex
matrices, in which resolution of analytes is poor,
these instruments can lead to false positives
(10)
.
Far more ?exible are triple quadrupole systems, in
which ions are selectively transmitted by the ?rst
quadrupole, CID occurs in the second quadrupole
(acting only as a collision cell), and the product ions
are separated in the third quadrupole. These systems
allow a number of different scanning modes (product
ions of selected precursor ions, precursor ions of
selected product ions, and neutral loss scans) and
can provide very low limits of detection in trace
analysis. Unfortunately, triple quadrupole systems
are still relatively expensive. Ion-trap systems offer
a signi?cantly cheaper and more compact MS/MS
capability and they are ?nding increasing use in
analytical laboratories. They allow trapping and
subsequent CID of selected ions within the ion trap,
although scanning is restricted to product ions. Ion
traps are generally more sensitive in full scanning
mode than quadrupole systems but may be less
sensitive in selected ion mode. Although most of the
analytical methods used in the authors laboratory
were developed using a triple quadrupole instrument,
most have transferred to an ion-trap LC/MS system
without major problems. Expensive high-resolution
magnetic sector instruments linked to LC have
declined considerably in use other than for research
purposes. A->Neutral and weakly basic pesticides, such as organophosphates, triazines, and arylureas, ionized well using either method of ionization, positive APCI generally being the more sensitive. Cationic pesticides (paraquat and diquat) and those that exist predominantly as anions in solution (e.g. sulfonic acids) gave much better responses using ESI (positive or negative).
Another factor to consider in the choice of ionization method is the formation of adducts with sodium or other metal ions, in addition to the protonated molecule. Although these may help con?rm the molecular mass, they tend to lower the signal-tonoise ratio of the protonated molecule, resulting in higher detection limits in trace analysis and causing problems in quantitative analysis. Sodium adducts tend to form with positive ESI but not with APCI
Coupled to capillary
or micro LC, it is more sensitive than particle
beam and, like the latter, provides searchable spectra
similar to those produced by electron ionization (EI)
in gas chromatography/mass spectrometry (GC/MS).
It can also operate in CI mode.
It is reported to be more widely applicable than
APCI, extending the useful range of LC/MS to
lower polarity molecules, for example, polyaromatic
hydrocarbons, and may be more sensitive than APCI
for some neutral analyte
Quadrupole mass spectrometers coupled with
LC make up a large part of the market.
Single
quadrupole instruments are limited for unequivocal
identi?cation of unknown analytes in that CID
is limited to in source. As already stated, this
produces additional chemical noise and does not
resolve individual components when coelution from
the LC column occurs. For analysis of complex
matrices, in which resolution of analytes is poor,
these instruments can lead to false positives
(10)
.
Far more ?exible are triple quadrupole systems, in
which ions are selectively transmitted by the ?rst
quadrupole, CID occurs in the second quadrupole
(acting only as a collision cell), and the product ions
are separated in the third quadrupole. These systems
allow a number of different scanning modes (product
ions of selected precursor ions, precursor ions of
selected product ions, and neutral loss scans) and
can provide very low limits of detection in trace
analysis. Unfortunately, triple quadrupole systems
are still relatively expensive. Ion-trap systems offer
a signi?cantly cheaper and more compact MS/MS
capability and they are ?nding increasing use in
analytical laboratories. They allow trapping and
subsequent CID of selected ions within the ion trap,
although scanning is restricted to product ions. Ion
traps are generally more sensitive in full scanning
mode than quadrupole systems but may be less
sensitive in selected ion mode. Although most of the
analytical methods used in the authors laboratory
were developed using a triple quadrupole instrument,
most have transferred to an ion-trap LC/MS system
without major problems. Expensive high-resolution
magnetic sector instruments linked to LC have
declined considerably in use other than for research
purposes. A
29. ->Neutral and weakly basic pesticides, such as organophosphates, triazines, and arylureas, ionized well using either method of ionization, positive APCI generally being the more sensitive. Cationic pesticides (paraquat and diquat) and those that exist predominantly as anions in solution (e.g. sulfonic acids) gave much better responses using ESI (positive or negative).
Another factor to consider in the choice of ionization method is the formation of adducts with sodium or other metal ions, in addition to the protonated molecule. Although these may help con?rm the molecular mass, they tend to lower the signal-tonoise ratio of the protonated molecule, resulting in higher detection limits in trace analysis and causing problems in quantitative analysis. Sodium adducts tend to form with positive ESI but not with APCI
Coupled to capillary
or micro LC, it is more sensitive than particle
beam and, like the latter, provides searchable spectra
similar to those produced by electron ionization (EI)
in gas chromatography/mass spectrometry (GC/MS).
It can also operate in CI mode.
It is reported to be more widely applicable than
APCI, extending the useful range of LC/MS to
lower polarity molecules, for example, polyaromatic
hydrocarbons, and may be more sensitive than APCI
for some neutral analyte
Quadrupole mass spectrometers coupled with
LC make up a large part of the market.
Single
quadrupole instruments are limited for unequivocal
identi?cation of unknown analytes in that CID
is limited to in source. As already stated, this
produces additional chemical noise and does not
resolve individual components when coelution from
the LC column occurs. For analysis of complex
matrices, in which resolution of analytes is poor,
these instruments can lead to false positives
(10)
.
Far more ?exible are triple quadrupole systems, in
which ions are selectively transmitted by the ?rst
quadrupole, CID occurs in the second quadrupole
(acting only as a collision cell), and the product ions
are separated in the third quadrupole. These systems
allow a number of different scanning modes (product
ions of selected precursor ions, precursor ions of
selected product ions, and neutral loss scans) and
can provide very low limits of detection in trace
analysis. Unfortunately, triple quadrupole systems
are still relatively expensive. Ion-trap systems offer
a signi?cantly cheaper and more compact MS/MS
capability and they are ?nding increasing use in
analytical laboratories. They allow trapping and
subsequent CID of selected ions within the ion trap,
although scanning is restricted to product ions. Ion
traps are generally more sensitive in full scanning
mode than quadrupole systems but may be less
sensitive in selected ion mode. Although most of the
analytical methods used in the authors laboratory
were developed using a triple quadrupole instrument,
most have transferred to an ion-trap LC/MS system
without major problems. Expensive high-resolution
magnetic sector instruments linked to LC have
declined considerably in use other than for research
purposes. A->Neutral and weakly basic pesticides, such as organophosphates, triazines, and arylureas, ionized well using either method of ionization, positive APCI generally being the more sensitive. Cationic pesticides (paraquat and diquat) and those that exist predominantly as anions in solution (e.g. sulfonic acids) gave much better responses using ESI (positive or negative).
Another factor to consider in the choice of ionization method is the formation of adducts with sodium or other metal ions, in addition to the protonated molecule. Although these may help con?rm the molecular mass, they tend to lower the signal-tonoise ratio of the protonated molecule, resulting in higher detection limits in trace analysis and causing problems in quantitative analysis. Sodium adducts tend to form with positive ESI but not with APCI
Coupled to capillary
or micro LC, it is more sensitive than particle
beam and, like the latter, provides searchable spectra
similar to those produced by electron ionization (EI)
in gas chromatography/mass spectrometry (GC/MS).
It can also operate in CI mode.
It is reported to be more widely applicable than
APCI, extending the useful range of LC/MS to
lower polarity molecules, for example, polyaromatic
hydrocarbons, and may be more sensitive than APCI
for some neutral analyte
Quadrupole mass spectrometers coupled with
LC make up a large part of the market.
Single
quadrupole instruments are limited for unequivocal
identi?cation of unknown analytes in that CID
is limited to in source. As already stated, this
produces additional chemical noise and does not
resolve individual components when coelution from
the LC column occurs. For analysis of complex
matrices, in which resolution of analytes is poor,
these instruments can lead to false positives
(10)
.
Far more ?exible are triple quadrupole systems, in
which ions are selectively transmitted by the ?rst
quadrupole, CID occurs in the second quadrupole
(acting only as a collision cell), and the product ions
are separated in the third quadrupole. These systems
allow a number of different scanning modes (product
ions of selected precursor ions, precursor ions of
selected product ions, and neutral loss scans) and
can provide very low limits of detection in trace
analysis. Unfortunately, triple quadrupole systems
are still relatively expensive. Ion-trap systems offer
a signi?cantly cheaper and more compact MS/MS
capability and they are ?nding increasing use in
analytical laboratories. They allow trapping and
subsequent CID of selected ions within the ion trap,
although scanning is restricted to product ions. Ion
traps are generally more sensitive in full scanning
mode than quadrupole systems but may be less
sensitive in selected ion mode. Although most of the
analytical methods used in the authors laboratory
were developed using a triple quadrupole instrument,
most have transferred to an ion-trap LC/MS system
without major problems. Expensive high-resolution
magnetic sector instruments linked to LC have
declined considerably in use other than for research
purposes. A
30. Question 3: You can hire employees for your lab. What skills, knowledge and educational background would you look for?
31. education:
university degree in chemical engineering or chemistry from recognised university
32. Question 4: What is a mandatory and fundamental prerequisite not yet covered? How can you fulfill this prerequisite and what are the problems you will encounter? How can you solve them?