GaAs detectors for medical imaging

GaAs detectors for medical imaging. G.I.Ayzenshtat a* , S.M. Gushin a , O.B. Koretskaya b , O.P. Tolbanov b , A.V. Tyazhev b , E.A. Babichev c , V.R. Groshev c , G.A. Savinov c , A.P. Vorobiev d a Research Inst. of Semicond. Dev., Tomsk , Russia

estevan
Télécharger la présentation

GaAs detectors for medical imaging

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. GaAs detectors for medical imaging G.I.Ayzenshtata*, S.M. Gushina,O.B. Koretskayab, O.P. Tolbanovb, A.V. Tyazhevb, E.A. Babichevc, V.R. Groshevc, G.A. Savinovc ,A.P. Vorobievd a Research Inst. of Semicond. Dev., Tomsk, Russia b Siberian Physical Technical Institute, Tomsk, Russia c Budker Institute of Nuclear Physics, Novosibirsk, Russia d Institute of High Energy Physics, Protvino, Russia * ayzen@mail.tomsknet.ru

  2. Low dose digital radiographic device with GaAs - detectors Low dose digital radiographic device • The aim of this work was to test GaAs detectors in a real radiographic device • Gas detector was replaced by a GaAs one in a Low Dose Digital Radiographic Device “LDRD SIBERIA-N” Detector + Electronics

  3. E  const The main properties of compensated SI-GaAs(Cr) detectors • 1. SI-GaAs(Cr)detectors have no Schottky contacts, they have only ohmic ones. I-V characteristics are linear. • 2.The electric field (E) occupies the whole volume of detector. E(kV/cm) const • 3. Charge collection in GaAs-detectors is determined by electrons. • 4. The detector material resistivity is extremely high (ρ>109 Ω∙cm).

  4. I = -I0 Iz = 0 I = +I0 z X-Rays +U1 - U1 Attachment of GaAs-detectors to electronics The problem: the dark current of detector is about 10 nA. It is necessary to reduce the current by a factor of 10 for the electronics to operate. Solution of the problem: we offer to attach the detectors to electronics as shown in the figure.

  5. Symmetrical detector structures with the compensation of dark currents The real structure of the mammo- graphy detector (100 mkm pitch) The hypothetical detector structure I = 0 + U - U

  6. The results ofcurrent compensationmeasurements in the symmetrical device. • The output current of the detector decreases by a factor 30 X- rays screen +U0 -U0

  7. A fragment of hybrid circuit of electronics A fragment of detectors mounting Detector 1 Detector 2 64-channel chip Plane view of detector (h = 200mm)

  8. The magnitude of the output signalas a function of the Roentgen tube current The magnitude of the output signal as a function of dosage rate for various quantum energy

  9. Waveshape of the signal in vertical scanning of the detector SNR2 as a function of dosage rate

  10. «Channel form» obtained by vertical scanning To electronics X-rays U Down

  11. «Channel form» obtained by horizontal scanning The collimator moves along the horizontal axis The normalized amounts of signals for various channels during horizontal scanning To electronics X-rays Detector Collimator

  12. Image of the test pattern obtained by the GaAs detector with a 400 mkm pitch (The limiting image resolution is 1.4 pairs of lines per mm)

  13. Comparision of I-V characteristics for two detectors. 1. The GaAs liquid-phase epitaxy detector 2. The SI-GaAs(Cr) detector Plane view of the detector (h = 200mm) fabricated from GaAs liquid-phase epitaxy

  14. ITThe imageof the test pattern obtained by GaAs detector with a 200 mm pitch (The limiting image resolution is 2.8 pairs of lines per mm)

  15. Conclusion 1. It was shown experimentally that the GaAs material compensated with Cr can be used to fabricate efficient X-ray detectors. 2.The object image with the resolution 2.8 pairsof lines per mm has been obtained with GaAs detectors. 3.We consider the GaAs(Cr) liquid-phase epitaxy material to be a more promising material for obtaining the resolution more than 2.8 pairsof lines per mm.