1 / 20

Pulse Oximetry

Pulse Oximetry. Dr.C.N.Chandra Sekhar M.D. Oxygenation. Oxygen tension Oxygen content Oxygen saturation. Oxygen tension. Clarke electrode Fuel cell Transcutaneous electrodes. Oxygen content. (SpO 2 x Hb x 1.34) + 0.003 x PO 2. Oxygen Saturation.

umed
Télécharger la présentation

Pulse Oximetry

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Pulse Oximetry Dr.C.N.Chandra Sekhar M.D

  2. Oxygenation • Oxygen tension • Oxygen content • Oxygen saturation

  3. Oxygen tension • Clarke electrode • Fuel cell • Transcutaneous electrodes

  4. Oxygen content (SpO2 x Hb x 1.34) + 0.003 x PO2

  5. Oxygen Saturation • ODC Curve: • Plot of degree of Hb saturation against oxygen tension

  6. PULSE OXIMETRY • Uses spectrophotometry based on the Beer-Lambert law • Differentiates oxy- from deoxyhemoglobin by the differences in absorption at 660nm and 940nm • Minimizes tissue interference by separating out the pulsatile signal

  7. Absorption Spectrum

  8. Signals

  9. PULSE OXIMETRY • Estimates heart rate by measuring cyclic changes in light transmission • Measures 4 types of hemoglobin: deoxy, oxy, carboxy, and met • Estimates functional hemoglobin saturation: oxyhemoglobin/deoxy + oxy

  10. Beer-Lambert’s Law • Beer’s Law: • The intensity of transmitted light decreases exponentially with the concentration of the substance • Lambert’s Law: • The intensity of transmitted light decreases exponentially with the distance it travels

  11. Pulse Oximeter probe

  12. ABSORPTION SPECTRA

  13. SOURCES OF ERROR • Sensitive to motion • Standard deviation is certified to 2-3% down to 70% saturation • Sats below 85% increase the importance of error in the reading • Calibration is performed by company on normal patients breathing various gas mixtures, so calibration is certain only down to 80%

  14. SOURCES OF ERROR • Skin Pigmentation • Darker color may make the reading more variable due to optical shunting. • Dark nail polish has same effect: blue, black, and green polishes underestimate saturations, while red and purple have no effect • Hyperbilirubinemia has no effect

  15. SOURCES OF ERROR • Low perfusion state • Ambient Light • Delay in reading of about 12 seconds • Methylene blue and indigo carmine underestimate the saturation

  16. SOURCES OF ERROR • Dysfunctional hemoglobin • Carboxyhgb leads to overestimation of sats because it absorbs at 660nm with an absorption coefficient nearly identical to oxyhgb

  17. SOURCES OF ERROR • Methgb can mask the true saturation by absorbing too much light at both 660nm and 940nm. Saturations are overestimated, but drop no further than 85%, which occurs when methgb reaches 35%.

  18. Summary • HbF No significant clinical change • Meth Hb False low or high readings • COHb False High readings • Sulph Hb Not a clinical problem • Bilirubin Not a clinical problem

  19. Methylene Blue False low reading • Indocyanine green False low reading • Anaemia Contraversial

  20. Thank you

More Related