NA62 Gigatracker cooling requirements

NA62 Gigatracker cooling requirements. Gigatracker (GTK) modules will operate in vacuum and under high radiation Module has to be replaced on a regular basis C ooling system required to avoid performance loss

caelan
Télécharger la présentation

NA62 Gigatracker cooling requirements

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. NA62 Gigatrackercoolingrequirements • Gigatracker (GTK) modules will operate in vacuum and under highradiation • Module hastobereplaced on a regularbasis • Cooling system required to avoid performance loss • The operation temperature for thefrontendelectronics will be 5°C orlower • Low material budgetforthecoolingsystem

  2. Gigatracker Module supportandalignementstructure Coolingplate Readoutchip (12.5 x 20 mm), heatproduction ca. 3.2W per chip (2 W/cm2) beam direction Sensor, siliconpixels (27 x 60 mm) 3D schematicdrawing of the GTK module

  3. Materials in thesensorarea The total material budget (material in the beam) allowedfor the GTK module is 0.5% X0(radiation length).

  4. Coolingsystemsunderinvestigation • carbonplate, conductivecooling • convectivecooling in a vessel • microchannels

  5. Ourproposal: Microchannelcooling • Goals ofdevelopment: • Integration ofmicrochannelsintofrontendelectronics • Cooling via an separate coolingplatewithmicrochannels • 150mm thickness • 300mm thickness • Benefits: • Uniform temperaturedistribution in theareatobecooled • Small DT betweencoolantandreadoutchip => reduced thermal stress • Single-phaseandtwo-phasecoolingpossible • Technology studiedat EPFL with strong supportofindustrialpartners • Mutual understandingtoshareknowledge (EPFL <> CERN) • Specifityofourapplication: • Verylow material budget • Low heatflux

  6. Tentative layoutofmicrochannelsfortheGigatracker • areatobecooled ~40 x 60mm • channellength ~40mm • channelcrosssection 50mm x 50mm • separationwalls 25mm thick • heatflux in thecoolingregion 2W/cm2 • Support and Connection ofservicesoutside the sensor area and on one side • Single-phasecooling • Open pointsforthe prototype: • U-shapedchannelsorsinglelinechannels (pressureloss <> massflowneeded) • Thermal connectionofthereadoutchiptothecoolingplate

  7. C6F14 cooling liquid ofchoice • radiationhard • thermallyandchemicallystable • nonflammable, nontoxic, nonconducting • knownandusedat CERN (CMS and Atlas Tracker) • used in liquid phase

  8. More peopleforthe GTK cooling Paolo Petagna, CERN PH-DT-PO, DT cooling Project Leader Alessandro Mapelli, EPFL Mircrosystems Laboratory, Microfabricationtechnologies Piet Wertelaers, CERN PH-DT-PO, Consultant for design andlayoutofcoolingcircuit JerômeDaguin, PH-DT, Experience in C6F14plantsat CERN

  9. Plans forthenextmonths August/September: • firstlayoutforthemicrochannels • productionof a prototype atthe EPFL clean room • preparation teststand (coolingunit, instrumention …) • design ofconnectingcomponent • developmentof a CFD-Model Oktober/November: • commissioning teststand • firsttestatroomtemperature • preparationoftestatcoldtemperature, cryostat • assembly prototype andmockupchips Thermal interfacetobedesignedaccordingtochip design andfabrication!!!